Radiation-Emitting Products

What are the required quality control tests for new mammographic modalities?

Policy Guidance Help System logo

Mammography systems with image receptor modalities other than screen-film must undergo periodic quality control tests following procedures that are recommended by the manufacturer of the mammographic modality. As is the case for screen-film QC, the facility must keep records of these tests and all applicable corrective actions for the longer of: last 12 months or since the last annual inspection which verifies compliance, or until the test has been done two additional times at the required frequency. 21 C.F.R. 900.12(d)(2). Regardless of the mammographic modality, the mean glandular dose delivered during a single cranio-caudal view of an FDA-accepted phantom must not exceed 3.0 milligray (mGy) (0.3 rad) per exposure. 21 C.F.R. 900.12(e)(5)(vi). If the test results exceed the action level, corrective action must be taken within the required time frames. 21 C.F.R. 900.12(e)(8).

Back to Other Modalities Quality Control Tests

Page Last Updated: 08/12/2014
Note: If you need help accessing information in different file formats, see Instructions for Downloading Viewers and Players.