• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Radiation-Emitting Products

  • Print
  • Share
  • E-mail
-

When performing a physics survey or mammography equipment evaluation on a unit with multiple target/filter combinations, what tests or measurements must be performed for each combination?

Policy Guidance Help System logo

For a unit with multiple target/filter combinations, the following tests must be performed for each clinically used target/filter combination:

  • Focal spot condition (for different target materials (tracks) only) (21 C.F.R. 900.12(e)(5)(iii)(A)(3)

  • X-ray field/light field/image receptor/compression paddle alignment (for different target materials (tracks) only) (21 C.F.R. 900.12(e)(5)(vii))

  • Beam quality and half-value layer (21 C.F.R. 900.12(e)(5)(iv) and (vi))

  • System artifacts (21 C.F.R. 900.12(e)(5)(ix)) Also see Approved Alternative Standards

Back to Annual Physics Survey

-
-