Radiation-Emitting Products
-
Focal Spot Dimension QC Test

Citation:
900.12(e)(5)(iii)(B): Focal spot dimensions. Measured values of the focal spot length (dimension parallel to the anode cathode axis) and width (dimension perpendicular to the anode cathode axis) shall be within the tolerance limits specified in Table 1.
Table 1
|
Focal Spot Tolerance Limit |
||
|
Nominal Focal Spot Size (mm) |
Maximum Measured Dimensions |
|
|
Width (mm) |
||
|
0.10 0.15 0.20 0.30 0.40 0.60 |
0.15 0.23 0.30 0.45 0.60 0.90 |
0.15 0.23 0.30 0.65 0.85 1.30 |
-
-
-
-