Citation:
900.12(e)(4)(i),(ii),(iii)(A)(B): Semiannual quality control tests. Facilities with screen-film systems shall perform the following quality control tests at least semiannually:
(i) Darkroom fog. The optical density attributable to darkroom fog shall not exceed 0.05 when a mammography film of the type used in the facility, which has a mid-density of no less than 1.2 OD, is exposed to typical darkroom conditions for 2 minutes while such film is placed on the counter top emulsion side up. If the darkroom has a safelight used for mammography film, it shall be on during this test.