Citation:

900.12(e)(5)(iv): Beam quality and half-value layer (HVL). The HVL shall meet the specifications of 1020.30(m)(1) of this chapter for the minimum HVL. These values, extrapolated to the mammographic range, are shown in Table 2. Values not shown in Table 2 may be determined by linear interpolation or extrapolation.

Table 2

X-ray Tube Voltage (kilovolt peak) and Minimum HVL
Designed Operating Range (kV)

Measured Operating Voltage (kV)

Minimum HVL

(millimeters of aluminum) 

Below 50

20

25

30 

0.20

0.25

0.30 

 

Questions:
  1. Must the compression paddle be placed in the x-ray beam during half value layer (HVL) measurements?

 