2010 Scorecard Statistics
- December 2010
- November 2010
- October 2010
- September 2010
- August 2010
- July 2010
- June 2010
- May 2010
- April 2010
- March 2010
- February 2010
- January 2010
December 2010
| Certified facilities, as of October 1, 2010 | 8,652 | ||
| Certification statistics, as of December 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,619 / 12,412 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 6,174 / 8,957 | ||
| FY 2011 inspection statistics, as of December 1, 2010 | |||
| Facilities inspected | 1,205 | ||
| Total units at inspected facilities | 1,672 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.7% | ||
| Level 2 violation | 13.5% | ||
| Level 3 violation | 3.1% | ||
| Percent of inspections with no violation | 82.7% | ||
| Total annual mammography procedures reported, as of December 1, 20101 | 38,965,863 | ||
November 2010
| Certified facilities, as of October 1, 2010 | 8,652 | ||
| Certification statistics, as of November 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,643 / 12,427 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 6,106 / 8,874 | ||
| FY 2011 inspection statistics, as of November 1, 2010 | |||
| Facilities inspected | 538 | ||
| Total units at inspected facilities | 723 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.4% | ||
| Level 2 violation | 13.6% | ||
| Level 3 violation | 2.4% | ||
| Percent of inspections with no violation | 83.6% | ||
| Total annual mammography procedures reported, as of November 1, 20101 | 38,945,731 | ||
October 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of October 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,652 / 12,445 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 6,004 / 8,748 | ||
| FY 2010 inspection statistics, as of October 1, 2010 | |||
| Facilities inspected | 8,371 | ||
| Total units at inspected facilities | 11,686 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.8% | ||
| Level 2 violation | 14.5% | ||
| Level 3 violation | 3.4% | ||
| Percent of inspections with no violation | 81.3% | ||
| Total annual mammography procedures reported, as of October 1, 20101 | 38,924,837 | ||
September 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of September 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,646 / 12,412 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,903 / 8,614 | ||
| FY 2010 inspection statistics, as of September 1, 2010 | |||
| Facilities inspected | 7,605 | ||
| Total units at inspected facilities | 10,625 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.8% | ||
| Level 2 violation | 14.5% | ||
| Level 3 violation | 3.4% | ||
| Percent of inspections with no violation | 81.2% | ||
| Total annual mammography procedures reported, as of September 1, 20101 | 38,808,689 | ||
August 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of August 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,651 / 12,432 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,826 / 8,514 | ||
| FY 2010 inspection statistics, as of August 1, 2010 | |||
| Facilities inspected | 6,844 | ||
| Total units at inspected facilities | 9,602 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.9% | ||
| Level 2 violation | 14.5% | ||
| Level 3 violation | 3.5% | ||
| Percent of inspections with no violation | 81.1% | ||
| Total annual mammography procedures reported, as of August 1, 20101 | 38,823,849 | ||
July 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of July 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,649 / 12,437 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,748 / 8,412 | ||
| FY 2010 inspection statistics, as of July 1, 2010 | |||
| Facilities inspected | 6,164 | ||
| Total units at inspected facilities | 8,671 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.8% | ||
| Level 2 violation | 14.6% | ||
| Level 3 violation | 3.6% | ||
| Percent of inspections with no violation | 81.1% | ||
| Total annual mammography procedures reported, as of July 1, 20101 | 38,822,632 | ||
June 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of June 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,655 / 12,492 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,663 / 8,312 | ||
| FY 2010 inspection statistics, as of June 1, 2010 | |||
| Facilities inspected | 5,351 | ||
| Total units at inspected facilities | 7,536 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.9% | ||
| Level 2 violation | 14.6% | ||
| Level 3 violation | 3.6% | ||
| Percent of inspections with no violation | 81.0% | ||
| Total annual mammography procedures reported, as of June 1, 20101 | 38,384,929 | ||
May 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of May 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,664 / 12,519 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,582 / 8,193 | ||
| FY 2010 inspection statistics, as of May 1, 2010 | |||
| Facilities inspected | 4,626 | ||
| Total units at inspected facilities | 6,549 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.8% | ||
| Level 2 violation | 14.7% | ||
| Level 3 violation | 3.7% | ||
| Percent of inspections with no violation | 80.8% | ||
| Total annual mammography procedures reported, as of May 1, 20101 | 38,353,211 | ||
April 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of April 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,669 / 12,562 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,508 / 8,105 | ||
| FY 2010 inspection statistics, as of April 1, 2010 | |||
| Facilities inspected | 3,934 | ||
| Total units at inspected facilities | 5,592 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.9% | ||
| Level 2 violation | 14.6% | ||
| Level 3 violation | 3.7% | ||
| Percent of inspections with no violation | 80.8% | ||
| Total annual mammography procedures reported, as of April 1, 20101 | 38,135,035 | ||
March 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of March 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,681 / 12,623 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,402 / 7,963 | ||
| FY 2010 inspection statistics, as of March 1, 2010 | |||
| Facilities inspected | 3,162 | ||
| Total units at inspected facilities | 4,493 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.9% | ||
| Level 2 violation | 15.2% | ||
| Level 3 violation | 3.6% | ||
| Percent of inspections with no violation | 80.2% | ||
| Total annual mammography procedures reported, as of March 1, 20101 | 38,057,377 | ||
February 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of February 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,681 / 12,630 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,327 / 7,864 | ||
| FY 2010 inspection statistics, as of February 1, 2010 | |||
| Facilities inspected | 2,444 | ||
| Total units at inspected facilities | 3,494 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 0.9% | ||
| Level 2 violation | 15.9% | ||
| Level 3 violation | 3.8% | ||
| Percent of inspections with no violation | 79.4% | ||
| Total annual mammography procedures reported, as of February 1, 20101 | 37,975,545 | ||
January 2010
| Certified facilities, as of October 1, 2009 | 8,713 | ||
| Certification statistics, as of January 1, 2010 | |||
| Total certified facilities / Total accredited units | 8,692 / 12,705 | ||
| Certified facilities with FFDM2 units / Accredited FFDM units | 5,246 / 7,755 | ||
| FY 2010 inspection statistics, as of January 1, 2010 | |||
| Facilities inspected | 1,888 | ||
| Total units at inspected facilities | 2,668 | ||
| Percent of inspections where the highest noncompliance was a: | |||
| Level 1 violation | 1.0% | ||
| Level 2 violation | 15.7% | ||
| Level 3 violation | 3.7% | ||
| Percent of inspections with no violation | 79.6% | ||
| Total annual mammography procedures reported, as of January 1, 20101 | 37,477,657 | ||
| 1 This number is an aggregate of the total number of procedures performed annually as reported by facilities to their accreditation bodies. Facilities are asked to disclose this information at their initial accreditation, and then at the time of their re-accreditation, which takes place once every three years. FDA began collecting these data in 1998. The aggregate does not reflect the current number of procedures performed at these facilities, but only the numbers reported by them during the three-year period prior to the current date. We have aggregated only the numbers reported by certified, non-Veterans Administration facilities. 2 FFDM - Full Field Digital Mammography unit. |
-
-