Baxter Meridian® Hemodialyis (HD) Instrument
Class 1 Recall
| Date Recall Initiated: |
September 28, 2005 |
| Product: | Baxter Meridian® Hemodialyis (HD) Instrument (Product Codes 5M5576 and 5M5576R) |
| Use: | The Baxter Meridian® is a single patient hemodialysis instrument that prepares dialysis solution, circulates blood through an external tubing system and hemodialyzer, and monitors the system for operating conditions. |
| Recalling Firm: | Baxter Healthcare Corporation 1620 Waukegan Rd. , Bldg. R McGaw Park , IL 60085-6730 |
| Reason for Recall: | There have been reports of hemolysis (broken blood cells) related to kinks in the blood tubing sets used with the Meridian®. To date, there have been reports of at least one death and one serious injury, associated with kinking of blood tubing sets routed through both channels of the two channel clips mounted on the front of the Meridian®. The original labeling for the device is inadequate because it instructs users to route the blood tubing sets through both channels of the two-channel clips. |
| Public Contact: | Center for One Baxter Baxter Healthcare Corporation 1 Baxter Parkway Deerfield , IL 60015 800-422-9837 |
| FDA District: | Chicago |
| FDA Comments: |
|
|
|
Consumers with questions about the recall may contact the company at (800) 422-9837. Those who have technical questions may contact the company at (800) 553-6898. |
-
-