-
Medical Devices
-
Respironics PLV Continuum Ventilator (PLVC I)
Class 1 Recall
| Date Recall Initiated: |
March 20, 2006 |
| Product: | Respironics PLV Continuum Ventilator (PLVC I) |
| Use: | The PLVC I is a mechanical ventilator used to control or assist breathing. The ventilator is intended for home, institutional and portable settings and may be used for invasive as well as noninvasive ventilation. It is not intended for use as an emergency transport ventilator. |
| Recalling Firm: | Respironics California, Inc. 2271 Cosmos Ct. Carlsbad, CA 92009-7517 |
| Reason for Recall: | A design flaw can cause lead wires in the air flow valve to break during use. When this happens the ventilator stops providing mechanical ventilation. |
| Public Contact: | Mary Funk Regulatory Affairs Manager Respironics, California, Inc. 2271 Cosmos Ct. Carlsbad, CA 92011 760-918-7328 |
| FDA District: | Los Angeles |
| FDA Comments: |
|
|
|
Class I recalls are the most serious type of recall and involve situations in which there is a reasonable probability that use of the product will cause serious injury or death. |
-
-

