Medical Devices
Multi-Med, Inc. Huber (Non-coring) Needles and Finished Devices Containing Huber Needles Labeled as "All-Med Medical Products, Inc." and "Marquette Medical, Inc." Brands and Provided to Some Consignees in Bulk Without Labeling
Recall Class: Class I
Date Recall Initiated: August 30, 2010
Products:
Huber needles manufactured by Multi-Med, Inc and sold under the brand names “All-Med Medical Products, Inc.” and “Marquette Medical, Inc.” These products were also provided to consignees in bulk without labeling.
All Huber needles and products containing Huber needles, such as administration sets, distributed from July 12, 2005 to May 28, 2010 are affected by the recall.
To find the complete list of products affected by this recall, see the attachments below.
Samples of All-Med Medical Products, Inc. and Marquette Medical, Inc. product labeling highlighting identifying information are provided in Attachment C.
Use:
Huber needles are used on vascular access ports to access ports implanted under the skin of patients in need of repeated intravenous therapy. These needles are intended to penetrate the port septum without cutting or dislodging any silicone cores (or slivers) from the ports into which they are inserted.
Recalling Firm:
Multi-Med, Inc.
26 Victoria Ct.
Keene NH 03431
Reason for Recall:
Multi-Med has expanded its previous recall ucm224068 to include all Huber needles and finished products containing Huber needles distributed from July 12, 2005 to May 28, 2010.
A "coring" Huber needle could damage the implanted port by removing silicone slivers from the access membrane. The defect in the port as a result of coring can cause the ports to leak. The core can also potentially enter a patient’s body when the port is initially accessed if it is not flushed correctly. These issues may lead to the decreased effectiveness of the port, replacement of the port, infection, damage or death of tissue, swelling, or other serious adverse health consequences occurring as a result of the core travelling through blood vessels into the patient’s lungs or other organs. These issues may potentially cause death.
Public Contact:
Multi-Med Inc. representative Susan Starkey can be contacted by phone at 1-603-357-8733, Monday through Friday, from 8:00 to 5:00 PM (Eastern Standard Time) for instructions for disposition of the needles.
FDA District: New England
FDA Comments:
The company has advised distributors to:
- Immediately examine your inventory and quarantine product subject to the recall.
- Identify customers that have received these needles and notify them at once of this product recall.
FDA recommends clinicians and other users:
- Discontinue the use of the recalled product.
- If you do not have an alternate and unaffected non-coring needle, weigh the risk and benefits of continuing to use the affected product.
- Consider following FDA recommendations when using any Huber needle.
Class 1 recalls are the most serious type of recall and involve situations in which there is a reasonable probability that use of these products will cause serious adverse health consequences or death.
Health care professionals and consumers may report adverse reactions or quality problems they experienced using these products to the FDA’s MedWatch Adverse Event Reporting program either online, by regular mail, or by FAX.
Useful Links:
- Information About Huber Needles
- MedWatch: The FDA Safety Information and Adverse Event Reporting Program
Attachment A: All-Med Medical Products, Inc. Huber Needles and Products Containing Huber Needles Affected by the Recall
| PRODUCT CODE | PRODUCT DESCRIPTION | REORDER NUMBER | LOTS |
|---|---|---|---|
| M191 | 19ga x 1" Winged Administration Set | 50-1931 | 7H007A 9F027A |
| M191H | 19ga x 1" RA Huber Needle | 55-1930 | 5J008A 9F039A 9F042A |
| M191Y | 19ga x 1" Winged Administration Set w/Y Site | 50-1932 | 6I048A 7G048A 8A005A 9F021A |
| M195H | 19ga x .5" RA Huber Needle | 55-1910 | 7D070A |
| M195Y | 19ga x .5" Winged Administration Set w/Y Site | 50-1912 | 8F064A |
| M1915H | 19ga x 1.5" RA Huber Needle | 55-1940 | 6E055A |
| M1915Y | 19ga x 1.5" Winged Administration Set w/Y Site | 50-1942 | 5G011A 5H018A |
| M1975 | 19ga x .75" Winged Administration Set | 50-1921 | 7EO31A 7G045A 8L040A 9E013A 9F025A . |
| M1975H | 19ga x .75" RA Huber | 55-1920 | 5I062A 8L038A |
| M1975S | 19ga x .75" Standard Bore Winged Administration Set | 50-1923 | 7G055A 9H008A |
| M1975Y | 19ga x .75" Winged Administration Set w/Y Site | 50-1922 | 9B026M |
| M201 | 20ga x 1" Winged Administration Set | 50-2031 | 5G059A 5H042A 5K013A 5K022A 6K034A 7A002A 7I036A 7J007A 9F036A 9F044A |
| M201H | 20ga x 1" RA Huber Needle | 55-2030 | 5H050A 6H037A 7A041A 7I012A 7J031A 9J014A 9K014A |
| M201HS | 20ga x 1" Straight Huber Needle | 55-2035 | 7D049A 8C053A 8K017A |
| M201S | 20ga x 1" Standard Bore Winged Administration Set | 50-2033 | 6K031A |
| M201Y | 20ga x 1" Winged Administration Set w/Y Site | 50-2032 | 5G055A 5J034A 5J036A 6H027A 6I010A 6J053A 6K027A 7J008A 8D025A 8F029A 8L043A 9A024A |
| M2015 | 20ga x 1.5" Winged Administration Set | 50-2041 | 7F048A 8B011A 8G006A 9C031A 9D020A |
| M2015H | 20ga x 1.5" 90°RA Huber Needle | 55-2040 | 5H058A 5J007A 6H041A 7J029A 8D029A 8K055A 8K058A 8L034A |
| M2015HS | 20ga x 1.5" Straight Huber Needle | 55-2045 | 6A061A 6E030A 6E056A 6J014A 6K035A 7D006A 7E047A 7H022A 8J030A 8K005A 9B014A 9B044A 9H003A |
| M2015Y | 20ga x 1.5" Winged Administration Set w/Y Site | 50-2042 | 5G007A 6I038A 7F049A 7H014A 9F024A |
| M205 | 20ga x .5" Winged Administration Set | 50-2011 | 6A001A 9G036A |
| M205H | 20ga x .5" 90°RA Huber Needle | 55-2010 | 6J022A 7D069A |
| M205Y | 20ga x .5" Winged Administration Set w/Y Site | 50-2012 | 6A028A 7E044A 9G039A |
| M2075 | 20ga x .75" Winged Administration Set | 50-2021 | 5J012A 6A003A 6B056A 6E004A 6E011A 6H052A 6K026A 6L005A 7A016A 7B060A 7C027A 7D001A 7F008A 7I010A 7L026A 8J065A 9E024A 9F020A 9G033A 9H018A |
| M2075H | 20ga x .75" 90°RA Huber Needle | 55-2020 | 5H053A 7I014A 8B027A |
| M2075S | 20ga x .75" Standard Bore Winged Administration Set | 50-2023 | 7F024A |
| M2075Y | 20ga x .75" Winged Administration Set w/Y Site | 50-2022 | 0D013A 5J017A 6B060A 6D040A 6L052A 7B059A 7C025A 7C034A 7E076A 7K011A 8C051A 8F028A 8F063A 8H016A 8J019A 9D022A 9D034A 9E018A 9K026A |
| M221 | 22ga x 1" Winged Administration Set | 50-2231 | 5I030A 7F036A 7F046A 7G023A 8K011A 9K024A |
| M221H | 22ga x 1" RA Huber Needle | 55-2230 | 5K044A 7D052A 9A009A 9L019A |
| M221HS-NS | 22ga x 1" Straight Huber Needle - Bulk | 55-2230 | 5J031M 6B042A 6C016M 6K008A 6L040A 7C061M |
| M221S | 22ga x 1" Standard Bore Winged Administration Set | 50-2233 | 7I005A |
| M221Y | 22ga x 1" Winged Administration Set w/Y Site | 50-2232 | 6G023A 6H014A 6J046A 7A001A 8K013A 9I010A |
| M2215 | 20ga x 1.5" Winged Administration Set | 50-2241 | 6A030A 6F031A 6H050A 6L044A 7A040A 7D048A 7G012A 7J055A 7K035A 7L019A 7L024A 8H044A 8I014A 8I044A 8L022A 8L033A 9B034A 9C034A 9D021A 9F037A 9F049A 9J032A 9K029A |
| M2215H | 22ga x 1.5" 90 °RA Huber Needle | 55-2240 | 5K045A 7D058A 7K038A 9E046A |
| M2215HS | 22ga x 1” Straight Huberer Needle | 55-2235 | 7D017A 9K021A |
| M2215HS | 22ga x 1.5" Straight Huber Needle | 55-2245 | 0C006A 6D013A 6I017A 6J015A 6K037A 7E054A 7F042A 8A069A 8I061A 9C035A 9G030A 9I026A 9I027A 9J039A 9K020A |
| M2215S | 22ga x 1.5" Standard Bore Winged Administration Set | 50-2243 | 7D061A 7G037A |
| M2215Y | 22ga x 1.5" Winged Administration Set w/Y Site | 50-2242 | 7B020A 8I045A 9K022A |
| M225 | 22ga x .5" Winged Administration Set | 50-2211 | 6E057A 7C013A 7G024A 8D013A 8K025A 9F051A |
| M225H | 22ga x .5” RA Huber Needle | 55-2210 | 6J025A 7B013A 7C024A |
| M225Y | 22ga x .5” Winged Administration Set w/Y Site | 50-2212 | 8F024A |
| M2275 | 22ga x .75" Winged Administration Set | 50-2221 | 5I024A 6G037A 6H010A 6J006A 6L056A 7A018A 7A022A 7H012A 7J050A 8B016A 8K035A 9K031A 9K040A |
| M2275H | 22ga x .75” 90°RA Huber Needle | 55-2220 | 5H057A 7B031A 8L006A 8L036A |
| M2275HS | 22ga x .75" Straight Huber Needle | 55-2225 | 6J021A 7D050A 7H038A |
| M2275HS-NS | 22ga x .75" Straight Huber Needle-Bulk | 55-2235 | 5I038A 6B014M |
| M2275S | 22ga x .75" Standard Bore Winged Administration Set | 50-2223 | 8F002A |
| M2275Y | 22ga x .75" Winged Administration Set w/Y Site | 50-2222 | 5G001A 6H009M 7D064A 7H015A 7L014A 8L021A 9G056A 9J051A |
Attachment B: Marquette Medical Products Containing Huber Needles Affected by the Recall
| CATALOG NUMBER | PRODUCT DESCRIPTION | LOT NUMBERS |
|---|---|---|
| MW191 | Trap-A Safety Huber Needle 19ga x 1" Winged Administration Set | 6A051M 6C059M 7J002M |
| MW191-HT | Trap-A Safety Huber Needle 19ga x 1" Winged Administration Set | 5G040M 5H030M 9E026M |
| MW191Y | Trap-A Safety Huber Needle 19ga x 1" Winged Administration Set w/Y Site | 5G009M 5K015M 5K032M 6K029M 6K049M 8B008M |
| MW191Y-HT | Trap-A Safety Huber Needle 19ga x 1" Huber Trap with Needleless Y | 5F060M 5H059M |
| MW193 | Trap-A Safety Huber Needle 19ga x .75" Winged Administration Set | 5H027M |
| MW193-HT | Trap-A Safety Huber Needle 19ga x .75" Winged Administration Set | 5G003M 5K031M |
| MW193Y | Trap-A Safety Huber Needle 19ga x .75" Winged Administration Set w/Y Site | 5H028M 5I051M 6K050M 7G033M 7G047M 7H023M |
| MW201 | Trap-A Safety Huber Needle 20ga x 1" Winged Administration Set | 5K033M 6A059M 6G007M 6I040M |
| MW201Y | Trap-A Safety Huber Needle 20ga x 1" Winged Administration Set w/Y Site | 5I054M 6A039M 6E023M 7H024M 8F031M 9C032M |
| MW201Y-HT | Trap-A Safety Huber Needle 20ga x 1" Huber Trap with Needleless Y | 5G014M 5I002M 5I067M 7G026M 7H051M 8L032M |
| MW203 | Trap-A Safety Huber Needle 20ga x .75" Winged Administration Set | 5I032M 5K034M 6B044M 6C035 6C048M 6C065M 6H004M 7B022M |
| MW203-HT | Trap-A Safety Huber Needle 20ga x .75" Huber Trap | 5G010M 5H034M 5H037M 7D054M 8A033M 8K061M 9C013M 9C029M |
| MW203Y | Trap-A Safety Huber Needle 20ga x .75" Winged Administration Set w/Y Site | 5I049M 6B050M 6J040M 7D023M 7H063M |
| MW203Y-HT | Trap-A Safety Huber Needle 20ga x .75" Huber Trap with Needleless Y | 5H041M 5H071M 7E065M 8C072M 8H042M 8I046M 9A033M 9E028M |
| MW221 | Trap-A Safety Huber Needle 22ga x 1" Winged Administration Set | 6E001M 6J013M |
| MW221Y | Trap-A Safety Huber Needle 22ga x 1 Winged Administration Set w/Y Site | 5I059M 5I060M 5K057M 6A002M |
| MW223 | Trap-A Safety Huber Needle 22ga x .75" Winged Administration Set | 5H036M 5J020M 6E029M 6H003M 6K028M 7C032M |
| MW223-HT | Trap-A Safety Huber Needle 22ga x .75" Huber Trap | 5H049M 5I007M 5I061M 8J034M 9E034M |
| MW223Y | Trap-A Safety Huber Needle 22ga x .75" Winged Administration Set w/Y Site | 5G033M 5H007M 5I050M 6F029M 7J003M |
| MW223Y-HT | Trap-A Safety Huber Needle 22ga x .75" Huber Trap with Needleless Y | 5F069M 5H088M 5K041M |
| MWS191 | Trap-A Safety Huber Needle 19ga x 1" Standard Bore Winged Administration Set | 6B070M 6D030M 8D040M 9I012M |
| MWS193 | Trap-A Safety Huber Needle 19ga x .75" Standard Bore Winged Administration Set | 5G022M 6B021M 6E039M 7C030M |
| MWS201 | Trap-A Safety Huber Needle 20ga x 1" Standard Bore Winged Administration Set | 0D003M 5l060M 6A060M 6B052M 6E028M 6H056M 6I039M 7B016M 7D008M 7E056M 7F038M 7F039M 7G052M 7K053M 8F055M 9E048M |
| MWS203 | Trap-A Safety Huber Needle 20ga x .75" Winged Administration Set w/Y Site | 0B053M 0C050M 5G020M 5G039M 5I031M 6A052M 6B029M 6F048M 6J012M 6L014M 7B021M 7C068M 7H062M 7K051M 7L022M 8B053M 8G009M 8L019M 9A012M 9C033M 9K048M |
| MWS221 | Trap-A Safety Huber Needle 22ga x 1" Standard Bore Winged Administration Set | 5H038M 6B036M 6D031M |
| MWS223 | Trap-A Safety Huber Needle 22ga x .75" Standard Bore Winged Administration Set | 5L041M 6A062M 6F011M 6F013M 6K060M 8K054M |
Attachment C: Samples of All-Med Medical Products, Inc. and Marquette Medical, Inc. Product Labeling