St. Jude Medical 6 French Engage™ Introducer Devices

Recall Class: Class I

Date Recall Initiated: June 28, 2010
 
Product(s): Certain batches of the 6 Fr. Engage™ Introducer produced between April 27 and June 03, 2010. Affects approximately 5,120 devices.
 
Affected products and batch numbers:
  • Engage TR Introducer, 6 Fr. - ACT (2.25 mm), 7 cm length, .025" Max Guide wire, Batch: 3109782
  • Engage TR Introducer, 6 Fr. - ACT (2.25 mm), 12 cm length, .025" Max Guide wire, Batch: 3105838
  • Engage TR Introducer, 6 Fr. - ACT (2.25 mm), 25 cm length, .035" Max Guide wire, Batch: 3107645
  • Engage TR Introducer, 6 Fr. - ACT (2.25 mm), 12 cm length, .035" Max Guide wire, Batches: 3103891, 3110889, 3110892, 3118794, 3123051
  • Engage TR Introducer, 6 Fr. - ACT (2.25 mm), 12 cm length, .038" Max Guide wire, Batches: 3107789, 3107790
 
St. Jude Medical sent customers an "Urgent Medical Device Recall Notice" letter dated June 24, 2010. The letter was addressed to Cath Lab Manager/ Risk Manager and described the problem and the product involved.
 
Sales Representatives were also sent a St. Jude Medical Memorandum "Field Action Notice" dated June 24, 2010 and a reconciliation form. The memorandum described the problem, product, and advised Sales Representatives to visit the affected accounts and perform the assigned tasks.
 
Reason for Recall: The Company identified that some devices in the affected batches have the potential for separation of the shaft (sheath) from the hub or for a break in the hub assembly. If either of these were to occur during use, a potentially life threatening episode of bleeding could occur.
 
Use: The Engage™ Introducer is used to implant catheters and electrodes into blood vessels during surgery and helps to prevent blood loss.
 
Recalling Firm:          St. Jude Medical, Inc. Cardiovascular Division
                              14901 Deveau Place
                              Minnetonka, Minnesota 55345
 
Public Contact:         
St. Jude Medical Cardiology Customer Service at minnetonkacustomerservice@sjm.com or call 1-952-933-4700, or 1-888-864-7444
 
FDA District: Minneapolis
 
FDA Comments:
The company is advising customers to discontinue use of the product.