Medical Devices
Letter of Authorization: Emergency Use of Diatherix 2009 H1N1 Test
October 9, 2009
Dennis L. Grimaud
Chairman and Chief Executive Officer
DIATHERIX Laboratories, Inc.
601 Genome Way, Suite 4208
Huntsville, AL 35806
Dear Mr. Grimaud:
This letter is in response to your request that the Food and Drug Administration (FDA) issue an Emergency Use Authorization (EUA) for emergency use of the DIATHERIX H1N1-09 Influenza Test for the diagnosis of 2009 H1N1 influenza virus infection, pursuant to section 564 of the Federal Food, Drug, and Cosmetic Act (the Act) (21 U.S.C. §360bbb-3), by DIATHERIX Laboratories, Inc., a CLIA High Complexity Laboratory certified under the Clinical Laboratory Improvement Amendments of 1988, 42 U.S.C. §263a, to perform high complexity tests.
On April 26, 2009, pursuant to section 564(b)(1)(C) of the Act (21 U.S.C. §360bbb-3(b)(1)(C)), the Secretary of the Department of Health and Human Services (HHS) determined that there is a public health emergency under 42 U.S.C. §247d that affects, or has a significant potential to affect, national security, and that involves a specified biological, chemical, radiological, or nuclear agent or agents, or a specified disease or condition that may be attributable to such an agent or agents - in this case, 2009 H1N1 influenza virus.1 Pursuant to section 564(b) of the Act (21 U.S.C. §360bbb-3(b)), and on the basis of such determination, the Secretary of HHS then declared an emergency justifying the authorization of the emergency use of certain in vitro diagnostics for the detection of 2009 H1N1 influenza virus, subject to the terms of any authorization issued under 21 U.S.C. §360bbb-3(a).
Having concluded that the criteria for issuance of this authorization under section 564(c) of the Act (21 U.S.C. §360bbb-3(c)) are met, I am authorizing the emergency use of the DIATHERIX H1N1-09 Influenza Test (as described in the scope section of this letter (Section II)) for the diagnosis of 2009 H1N1 influenza virus infection in individuals with signs and symptoms of respiratory infection, subject to the terms of this authorization.
I. Criteria for Issuance of Authorization
I have concluded that the emergency use of the DIATHERIX H1N1-09 Influenza Test for the diagnosis of 2009 H1N1 influenza virus infection in individuals with signs and symptoms of respiratory infection meets the criteria for issuance of an authorization under section 564(c) of the Act, because I have concluded that:
- The recently isolated 2009 H1N1 influenza virus can cause influenza, a serious or life threatening disease or condition to humans infected by this virus.
- Based on the totality of scientific evidence available to FDA, it is reasonable to believe that the DIATHERIX H1N1-09 Influenza Test may be effective for the diagnosis of 2009 H1N1 influenza virus infection, and that the known and potential benefits of the DIATHERIX H1N1-09 Influenza Test, when used in the diagnosis of 2009 H1N1 influenza virus infection, outweigh the known and potential risks of such product; and
- There is no adequate, approved, and available alternative to the emergency use of the DIATHERIX H1N1-09 Influenza Test for the diagnosis of 2009 H1N1 influenza virus infection.2
Therefore, I have concluded that the emergency use of the DIATHERIX H1N1-09 Influenza Test for the diagnosis of 2009 H1N1 influenza virus infection in individuals with signs and symptoms of respiratory infection meets the above criteria for issuance of an authorization.
II. Scope of Authorization
I have concluded, pursuant to section 564(d)(1) of the Act, that the scope of this authorization is limited to the use of the authorized DIATHERIX H1N1-09 Influenza Test for the diagnosis of 2009 H1N1 influenza virus infection in individuals with signs and symptoms of respiratory infection.
The Authorized DIATHERIX H1N1-09 Influenza Test:
The DIATHERIX H1N1-09 Influenza Test is a multiplexed molecular diagnostic assay that performs target enriched multiplex PCR nucleic acid amplification on the ABI 9700 thermocycler followed by probe hybridization and subsequent detection on the Qiagen Luminex LiquiChip 100 platform for the in vitro qualitative detection of 2009 H1N1 influenza viral RNA in nasopharyngeal swabs, nasal swabs, throat swabs, nasal aspirates, and nasopharyngeal aspirates specimens from patients with signs and symptoms of respiratory infection.
The DIATHERIX H1N1-09 Influenza Test includes the following primer and probe sets:
- H1 (H109C): a total of 4 primers designed for nested PCR to detect the presence of the hemagglutinin gene specifically found in the 2009 H1N1 influenza A virus.
- N1 (N109B): a total of 4 primers designed for nested PCR to detect the presence of the neuraminidase gene specifically found in the 2009 H1N1 influenza A virus.
- Probes: each amplicon is hybridized to complementary capture probes (H109C De and N109B De, respectively), which are covalently coupled to color coded beads detectable by the Luminex technology.
The DIATHERIX H1N1-09 Influenza Test also includes the following control materials:
- ABM, extraction positive control: Acinetobacter baumannii, ATCC strain 19606, will be used as a culture stock diluted and subjected to extraction as an additional sample during each batch of patient specimen extractions to demonstrate the effectiveness of the extraction method (rule out false negatives due to extraction failure and false positives due to carryover contamination).
- PCR Blank, negative control: A water blank will be run as an additional PCR sample during each batch of patient specimen testing to demonstrate that no carryover contamination has occurred during the PCR process (rule out false positives).
- PCR positive control: Nucleic acid from Haemophilus influenzae, ATCC strain 10211, will be run as a separate PCR sample with each batch run of patient specimens. The PCR positive control demonstrates the effectiveness of the PCR reaction to amplify targets in the assay (rule out PCR false negatives).
- PCR positive internal control: DNA from Acinetobacter baumannii ATCC strain 19606 will be added to each PCR to act as an internal amplification control. For each sample, if either the ABM target or any other target is positive, the PCR passes. If both ABM and all other targets fail to produce a positive signal, the PCR has failed and must be repeated.
The DIATHERIX H1N1-09 Influenza Test requires the following hardware with corresponding software:
- Thermo Fisher Kingfisher 96, software version 2.6.2: nucleic acid extraction instrument.
- ABI 9700 Thermocycler, software version 3.09: PCR amplification instrument.
- Qiagen Luminex LiquiChip 100, software version 2.3.182: bead detection instrumentation.
The DIATHERIX H1N1-09 Influenza Test requires the use of the following additional reagent kits:
- Starplex Collection Kit (tube, swab, medium, biohazard bag and shipping box) (Catalog number: SP132-FL75).
- Scigenix MagnetX Extraction Kit (Catalog number: R2-2400-DTX-I0).
- Qiagen OneStep RT-PCR Kit (Catalog number: 210212).
- BioRad Luminex beads (Catalog number: 171506xx).
The above described DIATHERIX H1N1-09 Influenza Test, when labeled consistently with the labeling authorized by FDA, entitled Diatherix Laboratories H1N1-09 Influenza Test Package Insert, as may be revised with written permission of FDA, is authorized to be used by DIATHERIX Laboratories, Inc.,3 under this EUA, despite the fact that it does not meet certain requirements otherwise required by federal law.
The above described DIATHERIX H1N1-09 Influenza Test is authorized to be accompanied by the following information pertaining to the emergency use, which are authorized to be made available to healthcare providers and patients:
- Fact Sheet for Healthcare Providers: Interpretation of the DIATHERIX H1N1-09 Influenza Virus Test Results
- Fact Sheet for Patients: Understanding the DIATHERIX H1N1-09 Influenza Virus Test Results
As described in section IV below, DIATHERIX Laboratories, Inc., is also authorized to make available additional information relating to the emergency use of the authorized DIATHERIX H1N1-09 Influenza Test that is consistent with, and does not exceed, the terms of this letter of authorization.
I have concluded, pursuant to section 564(d)(2) of the Act, that it is reasonable to believe that the known and potential benefits of the authorized DIATHERIX H1N1-09 Influenza Test in the specified population, when used for diagnosis of 2009 H1N1 influenza virus infection, outweigh the known and potential risks of such a product.
I have concluded, pursuant to section 564(d)(3) of the Act, based on the totality of scientific evidence available to FDA, that it is reasonable to believe that the authorized DIATHERIX H1N1-09 Influenza Test may be effective in the diagnosis of 2009 H1N1 influenza virus infection pursuant to section 564(c)(2)(A) of the Act. The FDA has reviewed the scientific information available including the information supporting the conclusions described in Section I above, and concludes that the authorized DIATHERIX H1N1-09 Influenza Test, when used to diagnose 2009 H1N1 influenza virus infection in the specified population, meets the criteria set forth in section 564(c) of the Act concerning safety and potential effectiveness.
The emergency use of the authorized DIATHERIX H1N1-09 Influenza Test under this EUA must be consistent with, and may not exceed, the terms of this letter, including the scope and the conditions of authorization set forth below. Subject to the terms of this EUA and under the circumstances set forth in the Secretary of HHS' determination under section 564(b)(1)(C) described above and the Secretary of HHS' corresponding declaration under section 564(b)(1), the DIATHERIX H1N1-09 Influenza Test described above is authorized to diagnose 2009 H1N1 influenza virus infection in individuals with signs and symptoms of respiratory infection.
This EUA will cease to be effective when the declaration of emergency is terminated under section 564(b)(2) of the Act or when the EUA is revoked under section 564(g) of the Act.
III. Waiver of Certain Requirements
I am waiving the following requirements for the DIATHERIX H1N1-09 Influenza Test during the duration of this emergency use authorization:
- Current good manufacturing practice requirements, including the quality system requirements under 21 CFR Part 820 with respect to the design, manufacture, packaging, labeling, and storage of the DIATHERIX H1N1-09 Influenza Test.
- Labeling requirements for cleared, approved, or investigational devices, including labeling requirements under 21 CFR 809.10 and 809.30, except for the intended use statement (21 CFR 809.10(a)(2), (b)(2)), adequate directions for use (21 U.S.C. 352(f)), (21 CFR 809.10(b)(5) and (8)), any appropriate limitations on the use of the device including information required under 21 CFR 809.10(a)(4), and any available information regarding performance of the device, including requirements under 21 CFR 809.10(b)(12).
IV. Conditions of Authorization
Pursuant to section 564 of the Act, I am establishing the following conditions on this authorization:
DIATHERIX Laboratories, Inc.
- DIATHERIX Laboratories, Inc., will not sell or distribute the DIATHERIX H1N1-09 Influenza Test to other laboratories.
- DIATHERIX Laboratories, Inc., will include with reports of the results of the DIATHERIX H1N1-09 Influenza Test the authorized DIATHERIX H1N1-09 Influenza Test Fact Sheets for Healthcare Providers and the authorized DIATHERIX H1N1-09 Influenza Test Fact Sheets for Patients.
- DIATHERIX Laboratories, Inc., will make available on its Web site the authorized DIATHERIX H1N1-09 Influenza Test Fact Sheets for Healthcare Providers and the authorized DIATHERIX H1N1-09 Influenza Test Fact Sheets for Patients.
- DIATHERIX Laboratories, Inc., will clearly and conspicuously state on reports of the results of the DIATHERIX H1N1-09 Influenza Test that this test is only authorized for the diagnosis of 2009 H1N1 influenza virus and not for seasonal influenza A, B, or any other pathogen.
- DIATHERIX Laboratories, Inc., will inform state and/or local public health authority(ies) of this EUA, including the terms and conditions herein.
- All advertising and promotional descriptive printed matter relating to the use of the DIATHERIX H1N1-09 Influenza Test shall be consistent with the Fact Sheets and authorized labeling, as well as the terms set forth in this EUA and other requirements set forth in the Act and FDA regulations.
- All advertising and promotional descriptive printed matter relating to the use of the DIATHERIX H1N1-09 Influenza Test shall clearly and conspicuously state that:
- This test has not been FDA cleared or approved;
- This test has been authorized by FDA under an Emergency Use Authorization;
- This test has been authorized only for the detection of 2009 H1N1 influenza virus and not for any other viruses or pathogens;
- This test is only authorized for the duration of the declaration of emergency under section 564(b)(1) of the Act, 21 U.S.C. §360bbb-3(b)(1); and
- The declaration of emergency will expire on April 26, 2010, unless it is terminated or revoked sooner or renewed.
- No advertising or promotional descriptive printed matter relating to the use of the DIATHERIX H1N1-09 Influenza Test assay may represent or suggest that this test is safe or effective for the diagnosis of 2009 H1N1 influenza virus.
- DIATHERIX Laboratories, Inc., will have a process in place for reporting test results to health care providers and federal, state and/or local public health authorities, as appropriate.
- DIATHERIX Laboratories, Inc., will track adverse events and report to FDA as required under 21 CFR part 803.
- Through a process of inventory control, DIATHERIX Laboratories, Inc., will maintain records of device usage.
- DIATHERIX Laboratories, Inc., will collect information on the performance of the assay, and report to FDA any suspected occurrence of false positive or negative results of which DIATHERIX Laboratories, Inc., becomes aware.
- DIATHERIX Laboratories, Inc., is authorized to make available additional information relating to the emergency use of the authorized DIATHERIX H1N1-09 Influenza Test that is consistent with, and does not exceed, the terms of this letter of authorization.
- Only DIATHERIX Laboratories, Inc., may request changes to the authorized DIATHERIX H1N1-09 Influenza Test Fact Sheet for Healthcare Providers or the authorized DIATHERIX H1N1-09 Influenza Test Fact Sheet for Patients. Such requests will be made by contacting FDA concerning FDA review and approval.
- DIATHERIX Laboratories, Inc., will perform the assay on the Applied Biosystems 9700 Thermocycler, coupled to the Qiagen Luminex LiquiChip 100 detection platform.
- DIATHERIX Laboratories, Inc., will ensure that any records associated with this EUA are maintained until notified by FDA. Such records will be made available to FDA for inspection upon request.
The emergency use of the authorized DIATHERIX H1N1-09 Influenza Test as described in this letter of authorization must comply with the conditions above and all other terms of this authorization.
V. Duration of Authorization
This EUA will be effective until the declaration of emergency is terminated under section 564(b)(2) of the Act or the EUA is revoked under section 564(g) of the Act.
Margaret A. Hamburg, M.D.
Commissioner of Food and Drugs
1 Memorandum, Determination Pursuant to § 564 of the Federal Food, Drug, and Cosmetic Act (April 26, 2009).
2 No other criteria of issuance have been prescribed by regulation under section 564(c)(4) of the Act.
3 This EUA does not authorize the DIATHERIX H1N1-09 Influenza Test to be sold or distributed to or used by other laboratories.