FDA Public Health Web Notification*: Information for Physicians on Sub-acute Thromboses (SAT) and Hypersensitivity Reactions with Use of the Cordis CYPHER™ Coronary Stent (1st archival)

This is an archived document. It is no longer current information and has been superseded. The current document can be found at http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/PublicHealthNotifications/ucm062120.htm

Issued 10/29/2003

Why is the Food and Drug Administration (FDA) issuing this notification?

Shortly after the April 24, 2003, approval of the Cordis CYPHER™ Coronary Stent, we began receiving adverse event reports through the Medical Device Reporting (MDR) system. We are still analyzing the information contained in these reports, but wanted to share with you the information we have at this time and encourage you to report your experience (see instructions and links below).

What is the CYPHER™ stent?

The CYPHER™ stent is a coronary stent coated with a thin polymer containing the drug sirolimus. The CYPHER™ stent releases an immunosuppressant drug, sirolimus, intended to reduce restenosis. For more information on the CYPHER™ stent, see the physician Instructions for Use at http://www.accessdata.fda.gov/cdrh_docs/pdf3/P020026d.pdf. For more information on the drug sirolimus, which is marketed as Rapamune™, see the physician Instructions for Use at  

 * CDRH Web Notifications are intended to augment the existing Safety Notification program by providing a mechanism to quickly disseminate device–related safety information that may be beneficial to healthcare providers. Unlike other forms of Notifications, such as Safety Alerts or Public Health Advisories, Web Notifications usually do not make specific recommendations and are typically used in situations where the available information and our understanding of an issue are still evolving.

 

What are FDA and Cordis doing about the SATs and hypersensitivity reactions?

After receiving a number of reports of thrombotic events, Cordis, in conjunction with FDA, issued a letter on July 7, 2003 (see http://www.fda.gov/bbs/topics/NEWS/2003/NEW00919.html) reminding physicians to follow the instructions for use and to report any adverse events to the Agency.

FDA and Cordis are gathering as much information as we can about the circumstances surrounding both SATs and hypersensitivity reactions. FDA is also working with the regulatory bodies of other countries to gain more information about foreign experience with the product.

What is causing these events?

While the etiology of reported adverse events associated with the CYPHER™ stent has not been identified, we are exploring a number of possibilities. We do not know whether product characteristics, patient characteristics (e.g. concomitant medications or illnesses), procedural factors, or a combination of these may have contributed to the events.

What can be done to reduce the likelihood that an SAT or hypersensitivity reaction will occur?

Until we can identify the cause(s) of the events, we cannot make specific recommendations. As we learn more about the mechanisms for these events, we will share that information with you. We do encourage you to follow the instructions for use.

Because of the uncertainty of the cause of reported hypersensitivity reactions, we urge you to be vigilant for any patient symptom that may be attributed to hypersensitivity.

In the meantime, we would like your help in gathering more information about these types of events. Historically, underreporting has made it difficult for FDA to properly assess the incidence of adverse events. You can help us by following your facility reporting procedures as noted below. We also appreciate any voluntary reports submitted by clinicians and patients.

How do I report adverse events to FDA?

The Medical Device Reporting (MDR) regulation, 21 CFR Part 803, requires hospitals and other user facilities, as defined by Part 803.3, to report adverse events to FDA and the device manufacturer when the event is associated with the use of the medical device and the device may have caused or contributed to a patient death or serious injury. You may refer to our MDR web site at: Reporting Adverse Events (Medical Devices).

You can also report to MedWatch, the FDA’s voluntary reporting program. You may submit reports to MedWatch one of four ways: online at http://www.accessdata.fda.gov/scripts/medwatch/; by telephone at 1-800-FDA-1088; by FAX at 1-800-FDA-0178; or by mail to MedWatch, Food and Drug Administration, 5600 Fishers Lane, Rockville, MD 20857.