Medical Devices
16. Servicing
INTRODUCTION
Interfaces
SERVICE PERSONNEL
SERVICE REQUIREMENTS
INSTALLATION
CONTRACT SERVICE
SERVICE EQUIPMENT
SERVICE PROCEDURES
ACCEPTANCE STATUS
SERVICE REPORTS
SERVICE REPORT ANALYSIS
Parts Shipping Trends
COMPLAINTS
CORRECTIVE AND PREVENTIVE ACTION
EXHIBIT
INTRODUCTION
The requirements in the Quality System (QS) regulation govern the methods used in, and the facilities and controls used for, the design, manufacture, packaging, labeling, storage, installation, and servicing of all finished devices intended for human use. Servicing covers the maintenance and repair of finished, distributed devices.
The intent of the quality system regulation is to assure that servicing is correctly performed and verified according to company specified requirements such that the serviced device is suitable for the intended use and that service information is collected and analyzed to help correct any quality system problems and device design, manufacturing, labeling, or packaging problems.
The basic servicing requirements are in 820.200, Servicing. However, there are related requirements throughout the QS regulation. For example, service procedures are documented per 820.181, Device Master Record; and servicing activities and/or data may lead to complaint analysis per 820.198, Complaint Files, or require corrective and preventive action per 820.100.
When a finished device manufacturer contracts with another supplier to perform their servicing, such service (or service contractor) must meet the applicable purchasing and servicing requirements in the QS regulation.
Interfaces
There are interface requirements in the QS regulation that apply to service functions. Section 820.30(b), Design and Development Planning, requires that each manufacturer shall establish and maintain plans that describe or reference the design and development activities and define responsibility for implementation. The plans shall identify and describe the interfaces with different groups or activities that provide, or result in, input to the design and development process.
The preamble clarifies the fact that these requirements extend to service functions by stating: the plan shall identify and describe the interfaces with different groups or activities that provide, or result in, input to the design process. Many organization functions, both inside and outside the design group, may contribute to the design process. For example, interfaces with marketing, purchasing, regulatory affairs, manufacturing, service groups, or information systems may be necessary during the design development phase. To function effectively, the design plan should establish the roles of these groups in the design process and describe the information that should be received and transmitted.
Therefore, for medical devices that require servicing, during appropriate activities such as design input and design reviews, service requirements and ease of service should be considered; and service managers, senior service technicians, etc., may need to participate in these design functions. Such participation may reduce the:
- need for maintenance and repairs;
- time to perform repairs,
- need for special tools; and
- cost of repairs.
Reducing the time for repairs and the need for special tools usually reduces production assembly time and manufacturing costs.
SERVICE PERSONNEL
Service shall be conducted by appropriately trained and experienced service personnel (820.25) in order to:
- assure, to the extent feasible, that the process of device handling, diagnosis, and repair will not compromise the determination of the root cause of the device failure;
- identify and correct the failure;
- correctly report the service information; and
- identify and report data related to a serious incident or adverse event.
The repair diagnosis should also try to determine, and/or provide adequate data to assist analysts in determining, the actual failure mechanism to the objective level necessary to correct or reduce the problem.
Thus, service personnel must be trained to adequately perform their assigned maintenance, repair, and reporting responsibilities. Such training shall be documented (820.25). The training is also performed in accordance with the instructions and procedures established under 820.200 for performing and verifying that servicing meets the specified requirements. Because servicing must be verified, service personnel must be made aware of defects and errors that may be encountered as part of their job functions (820.25). This training requirement usually does not require separate or additional training because basic training to perform repairs emphasizes the identification of defects and errors.
SERVICE REQUIREMENTS
The QS regulation does not require that a manufacturer service a device. The decision to service or have their devices serviced is left to the manufacturer. When a manufacturer specifies that they will perform service or contract to have service performed, such service must meet all of the applicable QS regulation requirements. Such a manufacturer shall establish and maintain instructions and procedures for performing and verifying that the servicing meets the manufacturers specified requirements. Section 820.200, Servicing, states:
(a) Where servicing is a specified requirement, each manufacturer shall establish and maintain instructions and procedures for performing and verifying that the servicing meets the specified requirements.
(b) Each manufacturer shall analyze service reports with appropriate statistical methodology in accordance with 820.100.
(c) Each manufacturer who receives a service report that represents an event which requires reporting to FDA under part 803 or 804 of this chapter shall automatically consider the report a complaint and shall process it in accordance with the requirements of 820.198.
(d) Service reports shall be documented and shall include:
(1) The name of the device serviced;
(2) Any device identification(s) and control number(s) used;
(3) The date of service;
(4) The individual(s) servicing the device;
(5) The service performed; and
(6) The test and inspection data.
INSTALLATION
Where service and installation are required by a manufacturer, both of these product activities are related and so are the QS requirements for both. Section 820.170 Installation states:
(a) Each manufacturer of a device requiring installation shall establish and maintain adequate installation and inspection instructions, and where appropriate test procedures. Instructions and procedures shall include directions for ensuring proper installation so that the device will perform as intended after installation. The manufacturer shall distribute the instructions and procedures with the device or otherwise make them available to the person(s) installing the device.
(b) The person installing the device shall ensure that the installation, inspection, and any required testing are performed in accordance with the manufacturer's instructions and procedures and shall document the inspection and any test results to demonstrate proper installation.
Some manufacturers use their service department or a service contractor to install their medical devices. Servicing may also include re-installing a device. As shown in Table 16.1, Comparison Of Servicing And Installation Requirements, the QS requirements for installation in 820.170 essentially parallel the requirements for service in section 820.200. For example, The QS regulation includes detail of the reporting requirements for servicing; for installation, the manufacturer chooses the information to document. However, from a practical viewpoint, each manufacturer would choose to have the same information documented. Thus, as appropriate, a manufacturer may combine most of their service and installation QS activities.
CONTRACT SERVICE
When a finished device manufacturer contracts with another supplier to perform their servicing, such service (or service contractor) must be obtained per the applicable requirements in 820.50, Purchasing. Each manufacturer shall establish and maintain procedures to ensure that all purchased or otherwise received services conform to specified requirements. Each manufacturer shall establish and maintain the requirements, including quality requirements, that are to be met by contractors. Each manufacturer shall:
(1) Evaluate and select potential contractors on the basis of their ability to meet specified requirements, including quality requirements. The evaluation shall be documented.
(2) Define the type and extent of control to be exercised over the contractors based on the evaluation results.
(3) Establish and maintain records of acceptable contractors.
Each manufacturer shall establish and maintain data that clearly describe or reference the specified service requirements, including quality requirements. Purchasing data shall be approved in accordance with 820.40.
A major portion of the purchasing requirements are met when the manufacturer meets the servicing requirements in section 820.200(a) which states:
(a) Where servicing is a specified requirement, each manufacturer shall establish and maintain instructions and procedures for performing and verifying that the servicing meets the specified requirements....
That is, these specified requirements, service instructions and procedures, and device verification procedures may be used together with other information such as the finished device description to help show a prospective contractor the scope and expected quality of the servicing that is being contracted.
Table 16.1 COMPARISON OF SERVICING AND INSTALLATION REQUIREMENTS
| SERVICING | INSTALLATION |
| (a) ... establish and maintain instructions and procedures for performing and verifying that the servicing meets the specified requirements. | (a) ... establish and maintain adequate installation and inspection instructions, and where appropriate test procedures. |
| (a) .... and verifying that the servicing meets the specified requirements. |
(a) Instructions and procedures shall include directions for ensuring proper installation so that the device will perform as intended after installation. (b) The person installing the device shall ensure that the installation, inspection, and any required testing are performed in accordance with the manufacturer's instructions and procedures and ... |
| (b) Each manufacturer shall analyze service reports with appropriate statistical methodology in accordance with 820.100. | |
| (a) ... establish and maintain instructions and procedures for performing and verifying that the servicing meets the specified requirements. | (a) ... The manufacturer shall distribute the instructions and procedures with the device or otherwise make them available to the person(s) installing the device. |
|
(d) Service reports shall be documented and shall include: (1) The name of the device serviced; (2) Any device identification(s) and control number(s) used; (3) The date of service; (4) The individual(s) servicing the device; (5) The service performed; and (6) The test and inspection data. |
(b) The person installing the device shall ensure that the installation, inspection, and any required testing are performed in accordance with the manufacturer's instructions and procedures and shall document the inspection and any test results to demonstrate proper installation. |
| (c) Each manufacturer who receives a service report that represents an event which requires reporting to FDA under part 803 or 804 of this chapter shall automatically consider the report a complaint and shall process it in accordance with the requirements of 820.198. |
SERVICE EQUIPMENT
Section 820.20(b)(2) Resources requires each manufacturer to provide adequate resources, including the assignment of trained personnel, for management, performance of work, and assessment activities to meet the requirements of this part. As appropriate, adequate resources include service instructions, service procedures, supporting DMR drawings, and service equipment. Service equipment includes equipment to perform the repair and to verify the proper performance of the serviced devices. Service equipment may include complex apparatus; however, it also includes any simple jigs, test cables, special hand tools, etc., as needed to meet the service needs of specific medical devices.
Servicing and Installation both require verifying that the device meets acceptance criteria. Therefore, appropriate and calibrated test equipment should be used. Section 820.72, Inspection, Measuring, and Test Equipment, requires that each manufacturer to ensure that all inspection, measuring, and test equipment, including mechanical, automated, or electronic inspection and test equipment, is suitable for its intended purposes and is capable of producing valid results. Each manufacturer shall establish and maintain procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained. The procedures shall include provisions for handling, preservation, and storage of equipment, so that its accuracy and fitness for use are maintained.
When establishing service and installation procedures, each manufacturer needs to comply with the Inspection, Measuring, and Test Equipment requirements, as appropriate, in order to assure that the serviced/installed device performs as intended. For example, a manufacturer may need to determine which service equipment, if any, needs to be calibrated in a laboratory and which, if any, may be calibrated using the self-contained internal calibrators. Also, the manufacturer may need to select equipment that is capable of producing valid results after being subjected to repetitive and demanding service.
SERVICE PROCEDURES
If any are required, maintenance needs, schedules, and procedures are developed as part of the device design program. Some preventive maintenance tasks and their schedules may result from reliability studies performed during design development.
Repair procedures are based in part on design verification and finished device test and inspection procedures, production procedures, and rework procedures. Other aspects of repair procedures are developed by qualified technical personnel and senior repair technicians. The development of procedures may involve inserting failures or defects and having another person find and repair them. The problems, discovery methods, and rework techniques are documented.
For redesigns, existing maintenance and/or repair procedures that are known to be current and correct may be referenced in the new service procedures or these may be renumbered and copied into the new procedures.
Identifying defective subassemblies or modules in the device and replacing them with good modules is a common servicing practice. The defective assembly is discarded, sent to be investigated, or is repaired at a designated facility with the necessary environmental conditions and facilities; test equipment and tools; component availability; trained rework employees; etc. This approach should also be covered by appropriate procedures.
The development of service procedures includes the development of appropriate service reporting forms.
Service instructions and procedures must be documented per 820.40. They are part of the device master record (DMR). Typical DMR documents (820.181) that are needed for service or that may be modified for service include:
- device specifications including appropriate drawings, component specifications, and software specifications;
- quality assurance procedures and specifications including acceptance criteria and the quality assurance equipment to be used; and
- installation, maintenance, and servicing procedures and methods.
These DMR documents usually cover:
- what the device does;
- theory of operation;
- operating instructions;
- safety;
- device specifications;
- component specifications, identification, and nomenclature;
- test apparatus, jigs, and special tools;
- typical failure modes and conditions;
- how to identify and isolate failures;
- test points where specific parameters may be measured;
- removal and replacement of parts;
- testing and inspecting (verifying) the repaired device;
- re-installation procedures, if applicable; and
- reporting forms.
ACCEPTANCE STATUS
Each manufacturer shall identify by suitable means the acceptance status of devices to indicate whether it has been service and whether it conforms with the acceptance criteria. The conformance is determined by the procedures established in accordance with 820.200(a).
The identification of acceptance status shall be maintained throughout servicing of the device to ensure that only devices which have passed the required acceptance activities are distributed, used, or installed (820.86). Identification is usually done by appropriate information on a decal, tag, or an attached pouch that contains the service request and/or report.
If a decal is left on the serviced device, for the next request and service, the decal can provide immediate information about the last service date, etc.
SERVICE REPORTS
Service activities shall be documented by service personnel and sent to the manufacturer according to the manufacturer's established procedures. As mentioned, service reports shall include:
(1) the name of the device serviced;
(2) any device identification(s) and control number(s) used;
(3) the date of service;
(4) the individual(s) servicing the device;
(5) the service performed; and
(6) the test and inspection data.
The device identification should be specific regarding the revision level, modification version, software version, etc., of the device in order to support analysis of the service data.
The test and inspection data should verify that the servicing meets the manufacturer's specified requirements. That is, the serviced device did, or did not, meet the acceptance criteria. (See Acceptance Criteria below.)
The service reports should also include information such as:
- device owner, address and phone number;
- specific location of the device;
- any unusual environmental conditions;
- any evidence of damage or misuse;
- if the device failed to meet specifications;
- if the device was being used for treatment or diagnosis;
- relationship, if any, of device to a death or serious injury; and
- date of last service and service report number, if known.
SERVICE REPORT ANALYSIS
Each manufacturer that performs servicing shall analyze service reports with appropriate statistical methodology in accordance with 820.100, Corrective and Preventive Action, which requires manufacturers to establish and maintain procedures for implementing corrective and preventive action. The procedures shall include requirements for analyzing service records and other sources of quality data to identify existing and potential causes of nonconforming product or other quality problems. The primary intent is to identify the action(s) needed to correct and prevent recurrence of nonconforming product and other quality problems; and to verify or validate the corrective and preventive action to ensure that such action is effective and does not adversely affect the finished device.
Failure and service report analysis should be conducted by appropriately trained and experienced personnel (820.25). Such personnel are also one of the resource requirements in 820.20(b)(2).
The analysis of service reports or subsequent analysis of the same or equivalent device(s) should be designed to determine the actual failure mechanism or quality problem to the objective level necessary to correct the problem.
When systematic failure has been diagnosed and corrective action established, a manufacturer need not analyze all additional devices that are serviced or returned with the same symptoms.
The analysis of service reports is totally dependent on the quality of the data in the reports. Therefore, it is very important that service training cover reporting so that the resulting reports are correct, complete, understandable, and easy to analyze.
The service reports for routine service requests for maintenance, adjustment, or repair of damage or failure resulting from long use, misuse or accident, usually do not need the same level of analysis as for other failures. However, some requests for service may appear to be routine when, in fact, they may be for unusual conditions that warrant attention. For example, service requests because of rapid wear, unusual problems, unusual maintenance, or development of hazardous conditions should receive a complete analysis in order to determine if corrective action is needed in the preventive maintenance procedures, design, labeling, manufacturing processes, etc. Enough information should be obtained from the customer to determine whether the request is for routine maintenance or the device is to be serviced for other reasons.
Parts Shipping Trends
As appropriate manufacturers should periodically (e.g., monthly) examine shipping records for repair parts. Any increases in shipment of specific parts due to unknown reasons should be analyzed to determine if a significant failure problem exists. Manufacturers have identified quality problems by this simple, low-cost technique.
COMPLAINTS
Service requests for repairing or investigating an event that allegedly resulted in a death or serious injury shall also be investigated as a complaint. Section 820.200(c) requires that each manufacturer who receives a service report that represents an event which requires reporting [Medical Device Reporting (MDR)] to FDA under part 803 or 804 of this chapter shall automatically consider the report a complaint and shall process it in accordance with the requirements of 820.198.
Section 820.198(d) requires that any complaint that represents an event which must be reported to FDA under part 803 or 804 of this chapter shall be promptly reviewed, evaluated, and investigated by a designated individual(s) and shall be maintained in a separate portion of the complaint files or otherwise clearly identified. In addition to the information required by 820.198(e), records of investigation under this paragraph shall include a determination of:
(1) whether the device failed to meet specifications;
(2) whether the device was being used for treatment or diagnosis; and
(3) the relationship, if any, of the device to the reported incident or adverse event.
Because of the MDR and complaint relationship, manufacturers should have the service personnel collect as much as is possible of the information required to complete the records of investigation, steps 1 - 3 listed above. Thus, the service form for some devices may need blanks/areas to support the collection of the needed information.
The service requirements AND complaint requirements shall be met for such combination service/MDR/complaint events. If the death or serious injury was caused by a design error, it may not be possible to perform a repair.
CORRECTIVE AND PREVENTIVE ACTION
A major intent of the service requirements is to look for quality problems during servicing and analysis of service data, and, if problems are found that affect or could affect safety or performance, to mandate appropriate corrective action. Thus, the collection (820.200) and analysis (820.100) of servicing data are required and these are part of the quality feedback system. Without the feedback provided by the quality audit and other information sources, such as complaints and service records, manufacturers operate in an open loop system with no assurance that the process used to design and produce devices is operating in a state of control.
Section 820.100, Corrective and Preventive Action, requires the analysis of quality records, service records, complaints, returned product, and other sources of quality data to identify existing and potential causes of nonconforming product, or other quality problems. Appropriate statistical methodology shall be employed where necessary to detect recurring quality problems.
In-warranty or out-of-warranty are not factors to be considered when collecting or analyzing data regarding servicing.
When nonconformities are found they are to be investigated to determine the cause, such as an inadequate quality system or a defect in the design, component(s), assembly, processing, labeling, packaging, installation method, service technique, etc.
The collection and analysis of service data should be broad based because the root cause may be: an inappropriate component; a bad component; early wear out; poor maintenance; compatibility; human factors, and safety; misuse; misuse due to inadequate labeling; poor workmanship; incorrect assembly; etc.
The investigation and corrective actions should continue until valid actions are identified and implemented to correct and prevent recurrence of nonconforming product and other quality problems.
If service instructions, techniques, equipment, etc., contribute to a quality problem, make a quality problem worse, destroy valuable data, etc., then such items are also subject to investigation and corrective action.
EXHIBIT
An example of a service request form is included in this chapter. This form may be modified to match individual needs as appropriate.
*** SAMPLE *** Page 1 of 2
| CUSTOMER SERVICE REQUEST | Service Report No.: | |
| Device Name: | ||
| ID/Cat. No.: | Lot Number: | |
| Any Specific Rev/Mod. #: | Date placed in use: | |
| Manufacturer/Distributor: | ||
| Account Name: | ||
| Account Address: | ||
| Phone Number: | Fax Number: | |
| Last Service Report (optional): | Date Of Last Service: | |
| Last service comments: | ||
| Technician: | Date: | |
| Device Location: | ||
| Customer Description Of Device Problem: |
||
|
Were you told that there was a serious injury or death associated with this problem? NO YES If yes, describe relationship to incident or event. Was device being used for treatment or diagnosis? |
||
| Describe any unusual environment: |
||
| Describe any evidence of misuse: |
||
| If a repair, device problem(s) you found: |
||
| Maintenance and/or repairs performed: |
||
| Service Completed: | Service not completed: | |
| Reason Not Completed: |
||
*** SAMPLE *** Page 2 of 2
| CUSTOMER SERVICE REQUEST | |
| In charts below, write identification code of components replaced. Write in first letter of ID code where necessary. Be SURE to put drawing number & rev. level in top row. | |
| Drawing number that shows replaced components listed below: rev. | |||
| CAP | CAP | CAP | CAP |
| RES | RES | RES | RES |
| IC | IC | IC | IC |
| TRANSISTOR | TRANSISTOR | TRANSISTOR | TRANSISTOR |
| DIODE | DIODE | DIODE | DIODE |
| BAT | LED | COIL | COIL |
| FUSE | FUSE | LAMP | LAMP |
| CABLE | CABLE | DET | DET |
| Drawing number that shows replaced components listed below: rev. | |||
| CAP | CAP | CAP | CAP |
| RES | RES | RES | RES |
| IC | IC | IC | IC |
| TRANSISTOR | TRANSISTOR | TRANSISTOR | TRANSISTOR |
| DIODE | DIODE | DIODE | DIODE |
| BAT | LED | COIL | COIL |
| FUSE | FUSE | LAMP | LAMP |
| CABLE | CABLE | DET | DET |
| Drawing number that shows replaced components listed below: rev. | |||
| CAP | CAP | CAP | CAP |
| RES | RES | RES | RES |
| IC | IC | IC | IC |
| TRANSISTOR | TRANSISTOR | TRANSISTOR | TRANSISTOR |
| DIODE | DIODE | DIODE | DIODE |
| BAT | LED | COIL | COIL |
| FUSE | FUSE | LAMP | LAMP |
| CABLE | CABLE | DET | DET |

