Medical Devices
Draft Guidance for Industry and Food and Drug Administration Staff - Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection of Antibodies to Borrelia burgdorferi
DRAFT GUIDANCE
This guidance document is being distributed for comment purposes only.
Document issued on: January 5, 2011
You should submit comments and suggestions regarding this draft document within 90 days of publication in the Federal Register of the notice announcing the availability of the draft guidance. Submit written comments to the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852. Submit electronic comments to http://www.regulations.gov. Identify all comments with the docket number listed in the notice of availability that publishes in the Federal Register.
For questions regarding this document contact Prasad Rao, Ph.D. at 301-796-6203 or by email at prasad.rao@fda.hhs.gov.
![]() |
U.S. Department of Health and Human Services |
Preface
Additional Copies
Additional copies are available from the Internet. You may also send an e-mail request to dsmica@fda.hhs.gov to receive an electronic copy of the guidance or send a fax request to 301-827-8149 to receive a hard copy. Please use the document number (1721) to identify the guidance you are requesting.
Table of Contents
- INTRODUCTION
- BACKGROUND
- SCOPE
- RISKS TO HEALTH
- ESTABLISHING PERFORMANCE CHARACTERISTICS
- CONTROLS
- PERFORMANCE STUDIES
- CLIA WAIVER
- REFERENCES
Draft Guidance for Industry and Food and Drug Administration Staff
Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection of antibodies to Borrelia burgdorferi
| This draft guidance, when finalized, will represent the Food and Drug Administration's (FDA's) current thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. You can use an alternative approach if the approach satisfies the requirements of the applicable statutes and regulations. If you want to discuss an alternative approach, contact the FDA staff responsible for implementing this guidance. If you cannot identify the appropriate FDA staff, call the appropriate number listed on the title page of this guidance. |
I. Introduction
FDA is issuing this draft guidance to provide industry and agency staff with recommendations for studies to establish the analytical and clinical performance of in vitro diagnostic devices (IVDs) intended for the detection of antibodies to Borrelia burgdorferi. These devices are used to aid in the diagnosis of Lyme disease.
FDA's guidance documents, including this guidance, do not establish legally enforceable responsibilities. Instead, guidances describe the Agency's current thinking on a topic and should be viewed only as recommendations, unless specific regulatory or statutory requirements are cited. The use of the word should in Agency guidances means that something is suggested or recommended, but not required.
II. Background
This document recommends studies for establishing the performance characteristics of in vitro diagnostic devices for the detection of antibodies to B. burgdorferi in human serum, plasma, and blood. Serological testing for antibodies to B. burgdorferi in Lyme disease diagnostics is a two step procedure [Ref. 1, 2]. Initial testing is done by an enzyme immunoassay (EIA) or immunofluorescent assay (IFA); specimens yielding positive or equivocal results are tested further by using a Western immunoblot assay. Specimens negative by a sensitive EIA or IFA do not need further testing. Results from Western blot assays for antibodies to B. burgdorferi are supplemental rather than confirmatory. Two-step positive results provide supportive evidence of exposure to B. burgdorferi, which could support a clinical diagnosis of Lyme disease but should not be used as a sole criterion for diagnosis.
A manufacturer who intends to market an in vitro diagnostic device for detection of antibodies to B. burgdorferi must conform to the general controls of the Federal Food, Drug, and Cosmetic Act (FD&C Act) and, unless exempt, obtain premarket clearance or approval prior to marketing the device (sections 510(k), 513, 515 of the FD&C Act; 21 U.S.C. 360(k), 360c, 360e). This document is intended to supplement 21 CFR 807.87 (information required in a premarket notification) and other FDA resources such as “Premarket Notification 510(k).” Guidance on the content and format for abbreviated and traditional 510(k)s is available in the guidance entitled “Format for Traditional and Abbreviated 510(k)s.”
III. Scope
This document recommends studies for establishing the performance characteristics of EIA or Western blot devices for the detection of antibodies to B. burgdorferi in human serum, plasma, and blood. The following is the product code for B. burgdorferi (Lyme disease) Class II devices cleared under 21 CFR 866.3830:
LSR - Reagent, Borrelia Serological Reagent
This document does not apply to B. burgdorferi nucleic acid amplification assays. Please contact the Division of Microbiology Devices in the Office of In Vitro Diagnostic Device Evaluation and Safety for further information on nucleic acid amplification devices.
IV. Risks to Health
Lyme disease is a significant public health concern. It is the most common vector-borne infectious disease in North America. The infectious agent in Lyme disease is the spirochete B. burgdorferi. Definitive diagnosis of B. burgdorferi infection is complicated by the varied nature of the clinical symptoms and the overlap of these symptoms with numerous other infectious and non-infectious diseases.
Failure of devices for the detection of B. burgdorferi antibodies to perform as expected, or failure to correctly interpret results may lead to incorrect patient management decisions and inappropriate public health disease reporting. In the context of individual patient management, a false negative report could lead to a delay or failure to provide treatment. A false positive report could lead to unnecessary or inappropriate treatment. Therefore, establishing the performance of these devices and understanding the risks that might be associated with their use is critical to their safe and effective use.
The studies conducted by manufacturers to establish the performance of B. burgdorferi antibody detection devices are the basis for determining the safety and effectiveness and substantial equivalence of these devices.
V. Establishing Performance Characteristics
A. Controls
When conducting the performance validation studies described below, we recommend that you run appropriate controls every day of testing for the duration of the analytical and clinical studies. We recommend that you provide a description of quality control recommendations in the labeling. Information should include a clear explanation of what controls and calibrators are to be used with the assay and how these should perform.
B. Performance Studies
We recommend you perform the following studies:
1. Cross-reactivity
Specimens known to contain potentially cross reactive antibodies to B.burgdorferi should be evaluated. We recommend that a minimum of ten specimens should be tested for each type. Examples of these include but are not limited to: (1) specimens from patients with infections such as tick-borne relapsing fever; syphilis and other treponemal infections; rickettsial diseases; ehrlichiosis; babesiosis; leptospirosis; infections with parvovirus B19, influenza viruses, Epstein-Barr virus, cytomegalovirus, and H. pylori; and (2) patients with diagnoses that can be confused with the late manifestations of Lyme disease such as chronic fatigue syndrome, fibromyalgia, rheumatoid arthritis, autoimmune diseases, and multiple sclerosis.
Interference
We recommend that you characterize the effects of potential interferents on assay performance. Examples of experimental designs, including guidelines for selecting interferents for testing, are described in CLSI documents EP7-A2 [Ref. 3]. Potential sources of interference can include high levels of compounds normally found in serum, such as triglycerides, hemoglobin (i.e., hemolyzed samples), bilirubin, protein, and lipids. We recommend that you state the criteria or level at which non-interference is determined.
2. Precision
Within-Laboratory Precision/Repeatability
We recommend that you conduct within-laboratory precision studies for devices that include instruments or automated components. You may perform these studies in-house, i.e., within your own company.
We recommend that you test sources of variability (such as operators, days, assay runs, etc.) for a minimum of 12 days (not necessarily consecutive), with 2 runs per day, and 2 replicates of each sample per run. These test days should span at least two calibration cycles. The test panel should consist of 3-6 patient samples at three levels of antibodies that include:
- A negative sample: a sample with no analyte such that results of repeated tests of this sample are negative 100% of the time.
- A “high negative” sample (C 5 concentration) : a sample with an analyte concentration below the clinical cut-off such that results of repeated tests of this sample are negative approximately 95% of the time (and results are positive approximately 5% of the time).
- A “low positive” sample (C 95 concentration) : a sample with a concentration of analyte just above the clinical cut-off such that results of repeated tests of this sample are positive approximately 95% of the time.
- A “moderate positive” sample: a sample with a concentration at which one can anticipate positive results approximately 100% of the time (e.g., approximately two to three times the concentration of the clinical cut-off).
CLSI documents EP5-A2 [Ref. 4] and EP12-A [Ref. 5] contain further information about designing and performing precision studies.
Reproducibility
- The protocol for the reproducibility study may vary slightly depending on the assay format. As a general guide, we recommend the following protocol:
- Evaluate the reproducibility of your test at 3 testing sites (for example, two external sites and one in-house site).
- Use a five day testing protocol, including a minimum of two runs per day, (unless the assay design precludes multiple runs per day) and three replicates of each panel member per run.
- Each day, have at least two operators at each facility perform the test. We recommend that, for rapid testing or point-of-care (POC)1 devices, you include a larger number of devices in your evaluation, in order to best represent the settings in which the devices will be used.
- Use the same sample panel as described in the repeatability study above.
- CLSI document, EP15-A2 [Ref. 6], contains additional information on reproducibility study design .
3. Specimen Collection and Handling Conditions
We recommend that you substantiate statements in your labeling about specimen storage and transport by assessing whether the device can maintain acceptable performance over the range of storage times and temperatures recommended to users. For example, an appropriate study would include an analysis of aliquots stored under the recommended conditions of time, temperature, or specified number of freeze/thaw cycles. We recommend that you state the criteria for an acceptable range of recoveries under the recommended storage and handling conditions as described in the CLSI document, H18-A [Ref. 7].
4. Assay Cut-Off Point
We suggest that data be furnished to explain how the assay cut-off point was selected and established. If appropriate, information should be provided on the use of an equivocal zone for testing. If data suggests that an equivocal zone is not appropriate, this should be explained.
5. Clinical Performance Studies
Current recommendation for Lyme disease serology testing is a two step procedure [Ref. 1, 2]:
Step I: EIA tests for IgM and/or IgG antibodies
Step II: Western blot tests for IgM or IgG antibodies when EIA tests are positive or equivocal.
Since the Borrelia burgdorferi strains from the United States (U.S.) differ markedly from Europe and other non-U.S. endemic regions, all antigenic reagents used in Lyme disease devices to be marketed in the U.S. should be manufactured from the U.S. strain sensu stricto B31 Borrelia burgdorferi spirochete and not from other non-U.S. strains. Similarly, non-U.S. clinical testing site data are not acceptable.
Clinical studies are necessary to establish the performance characteristics for both EIA and Western blot devices. Generally, we recommend that performance be assessed in the testing environment in which the device will ultimately be used (i.e., clinical laboratory) by individuals who will use the test in clinical practice (e.g., trained technologists). The following sets forth what is expected in terms of establishing the performance characteristics for each type of device.
Study Protocol
We recommend that you develop a detailed study protocol that includes specific patient inclusion and exclusion criteria, the type and number of specimens needed, directions for use, and a statistical analysis plan that accounts for variances to prevent data bias. You need to provide a statistical justification to support the sample size of the study population. We recommend that you include this and any other relevant protocol information in your premarket submission.
We encourage sponsors to contact the Division of Microbiology Devices to request a review of their proposed studies and selection of specimen types prior to study initiation.
Study Sites
We recommend that you conduct your studies at a minimum of three separate sites, one of which may be in-house. Clinical investigations of unapproved and uncleared in vitro diagnostic devices, including diagnostic devices for Lyme disease, are subject to the investigational device exemption (IDE) provisions of Section 520(g) of the Federal Food, Drug, and Cosmetic Act (21 U.S.C. 360j) and the implementing regulations. You should consider how 21 CFR part 812 (IDEs) applies to your particular study and refer to 21 CFR part 50 (informed consent), and 21 CFR part 56 (institutional review board review) for other applicable requirements.
We recommend that the performance evaluation for devices intended for point-of-care (POC) use include, at a minimum, one site at a clinical laboratory as well as 2 or more sites representative of non-laboratory settings in which the device is intended to be used (e.g., an emergency department).
Study Design
The following studies should be conducted and submitted with a 510(k) application for either a Lyme disease EIA device or a Lyme disease Western blot device. Study designs are similar for both IgM and IgG tests.
- Sensitivity Study : A study consisting of a minimum of 100 well characterized clinically or culture confirmed Lyme disease specimens should be conducted with the test device. These archived specimens should contain samples from early, early disseminated, and late phases of the disease. The sensitivity of the test device on these specimens should be compared to a predicate device. A pedigree for each characterized patient contributing samples for the studies should be included; this pedigree should encompass available clinical and laboratory testing information.
The following breakout of the patient population is suggested for sensitivity testing:- Initial (acute) samples from patients with documented erythema migrans (EM) or culture positive disease (<1 month, 1-2 months, 2–3 months after symptom onset).
- Convalescent samples from patients with documented EM or culture positive results, stratified by time of draw after initial appearance of symptoms (3-12 months).
- Testing of known Lyme disease patients with presentations other than EM, e.g., neuro-, arthritic, etc., in which the time interval from infection to symptom onset may be unknown or more than 1 year.
- Prospective Study: A study designed to test non-selected, prospectively collected consecutive samples should be conducted at a minimum of two geographically distinct locations within the U.S. using the test device and a comparison device.
- For EIA devices: The results should be compared to a predicate device. All positive and equivocal samples by the test device and the predicate device should be tested by an FDA-cleared IgM and/or IgG Western blot assay as applicable.
- For Western blot devices: The prospective study samples should be tested initially by an FDA cleared first-step EIA. All EIA positive and equivocal samples should be tested by the test device and the predicate Western blot. Positive and negative % agreement between the two devices should be provided.
Interpretation of Western blot results: The interpretation of Western blot results should follow the recommended criteria described by the Centers for Disease Control (CDC) [Ref. 1] and the Second National Conference on Serological Diagnosis of Lyme Disease [Ref. 2]. An IgM immunoblot is considered positive if two of the following three bands are present: 23 kDa (OspC), 39 kDa (BmpA), and 41 kDa (Fla). An IgG immunoblot is considered positive if five of the following 10 bands are present: 18 kDa, 23 kDa (OspC), 28 kDa, 30 kDa, 39 kDa (BmpA), 41 kDa (Fla), 45 kDa, 58 kDa, 66 kDa, and 93 kDa.
Analytical Specificity Testing : For the determination of analytical specificity, testing of samples from both endemic and non-endemic regions is recommended (minimum of 100 samples from each region). These samples should be obtained from an asymptomatic population but should not include prescreened blood donors. The results, expressed as % positives and % negatives, should be presented separately as endemic and non-endemic subjects.
CDC Reference Panel Testing : A standard panel of positive and negative specimens provided by the CDC for testing Lyme disease detection devices should be tested by the new device and the results provided as % agreement with the expected results. The data should be stratified by disease stage; early, intermediate, and late and be compared to the predicate device.
6. Labeling
Draft labeling for the device must comply with the requirements of 21 CFR 809.10.
C. CLIA Waiver
If you are seeking waiver for your device under the Clinical Laboratory Improvement Amendments of 1988 (CLIA),2 we recommend that you consult with the Division of Microbiology Devices staff regarding the design of specific studies to support the CLIA waiver application for your device. “Guidance for Industry and FDA Staff, Recommendations: Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices.”
VI. References
- CDC. Recommendations for test performance and interpretation from the Second National Conference on Serologic Diagnosis of Lyme Disease. MMWR 1995;44:590-591.
- Association of State and Territorial Public Health Laboratory Directors and the Centers for Disease Control and Prevention. Recommendations. In: Proceedings of the Second National Conference on Serologic Diagnosis of Lyme Disease (Dearborn, Michigan). Washington, DC: Association of State and Territorial Public Health Laboratory Directors 1995; 1-5.
- Clinical and Laboratory Standards Institute. 2005. Interference Testing in Clinical Chemistry; Approved Guideline. EP7-A2. Clinical and Laboratory Standards Institute, Wayne PA.
- Clinical and Laboratory Standards Institute. 2004. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline. EP5-A2. Clinical and Laboratory Standards Institute, Wayne PA.
- Clinical and Laboratory Standards Institute. 2002. User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline. EP12-A. Clinical and Laboratory Standards Institute, Wayne PA.
- Clinical and Laboratory Standards Institute. 2006. User Verification of Performance for Precision and Trueness; Approved Guideline. EP15-A2. Clinical and Laboratory Standards Institute, Wayne PA.
- Clinical and Laboratory Standards Institute. 1990. Procedures for Handling and Processing of Blood Specimens; Approved Guideline. H18-A. Clinical and Laboratory Standards Institute, Wayne PA.
1 Point-of-care tests, also known as bedside or near-patient tests, is a term that encompasses any diagnostic testing near the site of patient care regardless of whether the device is intended for use by a trained medical professional or by a lay user. The person conducting the diagnostic testing near the point of care, whether a trained medical professional or a lay user, is the caregiver in that instance as that person is the person providing care.
2 See 42 U.S.C. § 263a(d)(3).








