Medical Devices
Do It By Design - An Introduction to Human Factors in Medical Devices - Appendix, Glossary and References
APPENDIX
Points to Consider
In considering the need for, and conduct of, human factors analysis and testing, there are a number of issues and questions to ask.
1. Does the device require user interaction with respect to operation, maintenance, cleaning, or parts installation? If so, do the technology and device functions permit alternative user interface designs?
2. Given the combination of user interface, user population, and operating conditions, are errors likely?
3. Could the consequences of error be serious for the patient or user?
4. In doing actual testing:
- Is someone integral to the design team focusing on the user-related issues?
- Are users involved?
- Are hardware and software designers, technical writers, and others coordinating their efforts with respect to human factors?
- Has a test plan been developed?
- Have user requirements been developed, and are they being updated?
5. Has the design team checked the literature and company files for useful human factors information?
6. What studies, analyses, and test steps are being performed? Are staff examining all relevant issues related to the installation of accessories and operation of the device?
7. Has the project team done testing in simulated and/or actual environments?
8. Have user requirements been met?
9. User interface changes can be inadvertantly introduced into production models during manufacturing. Have they been accounted for?
GLOSSARY
The definitions in this section are pertinent to this document. In some cases they are general and may not coincide with a specific usage or application.
Administration set (intravenous): A device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, flow regulator, drip chamber, filter, stopcock, fluid delivery tubing, connectors, capped side tube to serve as an injection site, and hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Anthropometry: The field that involves the measurement of the dimensions and other physical characteristics of people and the application of this information to the design of things they use.
Blood glucose monitor: A device that quantitatively measures glucose concentrations in the blood.
Calibration: To check, adjust, or standardize systematically the graduations of a quantitative measuring instrument.
Cardiac monitor (including cardiotachometer and rate alarm): A device used to measure the heart rate from an analog signal produced by an electrocardiograph, vector cardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Catheter: A tubular medical device for insertion into canals, vessels, passageways, or body cavities, usually to permit injection or withdrawal of fluids or to keep a passage open.
Coding: Identifying objects or events with the use of recognizable symbols, typically visual or auditory, utilizing readily apparent variables such as color, shape, size, direction, pitch, or duration.
Cognition: Processing information about the environment and oneself in conscious intellectual activity, as in thinking, reasoning, remembering, and imagining.
Default: Parameters that are automatically selected by a machine in case deliberate actions by the user do not occur.
DC-Defibrillator: A device that delivers an electrical shock for defibrillating (restoring to normal heart rhythm) the atria or ventricles of the heart or to terminate other cardiac arrhythmias. The device delivers the electrical shock through paddles placed directly across the heart or on the surface of the body.
Enteral Feeding Tube: A tube for passing of food or medicines into the stomach.
Function: The action or accomplishment intended of a system where the system consists of a device and a user. Alternatively, individual primary functions, such as installation, maintenance, operation, and monitoring, are needed to accomplish the intended use of the user-device system.
Guide wire: A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Human Factors: In the broadest sense, a discipline devoted to the effects of user interface design, job aiding, and personnel training in the operation, maintenance, and installation of equipment.
Heart valve leaflets: Any of the leaf-like flaps of the bicuspid or tricuspid valves of the heart
.
Infusion pump: A device used to pump fluids into a patient in a controlled manner. The device may use a piston pump, roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Infusion pump cassette: That part of the set of intravenous tubing that fits into an infusion pump. Each cassette is "dedicated" or designed to fit a specific pump.
Iterative Prototyping: Successive small-scale tests on variations of a limited function prototype. Such tests permit continual design refinements based upon human performance.
Interlock: To prevent initiation of new operations until current operations are completed (computer science). To connect in such a way that no part can operate independently.
MedWatch Form 3500A: A form that must be completed by user facilities and manufacturers to report device-related adverse events to FDA under the medical device reporting (MDR) system (21 CFR, Parts 803 and 804).
Mockup: Usually a full-sized scale model of a structure, used for demonstration, study, or testing.
Negative transfer: Transfer of training that results in increased likelihood of human error, due to changes in the user interface or situations that are not obvious to the user.
Oxygen concentrator: A device that produces a high concentration of oxygen (85% to 95%) at clinically useful flow rates (up to 5 L/min) by physical separation of oxygen from ambient air. Oxygen concentrators are commonly used in home healthcare and occasionally in general anesthesia.
Screen print: A static image, represented on paper, which is used to show how a computer program will appear on a monitor.
Storyboard: One page in a series of paper representations of the sequence of actions possible in a system. Story boards representing a computer program could show keys, prompts, and changes in status.
Task: The steps or work activities required of the user in order to perform functions.
Task analysis: Identification and analysis of the key user tasks and steps for a device. The analysis may be conducted as a paper-and-pencil exercise for a device concept, or by running through the procedures on a prototype or actual device.
Transfer-of-training: The automatic application of skills, habits, or expectations to a new situation that appears similar to the one in which the skills and expectations were originally developed.
Usability Test: A test of either an actual device or an advanced prototype with a fully functional user interface. Data obtained includes user performance (time, errors, and accuracy) and subjective responses of test participants.
User performance data: Information describing human behaviors and responses during task performance. Examples of the criteria measured are frequency of accomplishing a task, time required for task accomplishment, and changes in performance with practice.
Ventilator: A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.
REFERENCES FOR FURTHER READING
Standards
Association for the Advancement of Medical Instrumentation. (1994). Human factors engineering guidelines and preferred practices for the design of medical devices. ANSI/AAMI-1994. Arlington, VA: AAMI.
American National Standards Institue. (1988). American National Standard for human factors engineering of visual display terminal workstations. ANSI/HFS 100-1988. New York: ANSI.
American National Standards Institute. (1991). Safety Color Code. ANSI Z535.1-1991. New York: ANSI.
American National Standards Institute. (1991). Safety color code. ANSI Z535.1-1991. New York: ANSI.
American National Standards Institute. (1991). Criteria for safety symbols. ANSI-Z535.3-1991. New York: ANSI.
American National Standards Institute. (1991). Product safety signs and labels. ANSI Z535.4-1991. New York: ANSI.
American Society for Testing and Materials. (1988). Minimum performance and safety requirements for components and systems of anesthesia gas machines. ASTM F1161-1988. Philadelphia: ASTM.
American Society for Testing and Materials. (1990). Ventilators intended for use in critical care. ASTM F1100-1990. Philadelphia: ASTM.
American Society for Testing and Materials. (1993). Standard specification for alarm signals in medical equipment used in anesthesia and respiratory care. ASTM F1463-93-1993. Philadelphia: ASTM.
International Electrotechnical Commission. (1988). Graphical symbols for electrical equipment in medical practice. IEC 878-1988. Geneva: Switzerland.
National Committee for Clinical Laboratory Standards. (1996) Laboratory instruments and data management systems: Design of software user interfaces and end user software systems validation, operation, and monitoring. NCCLS GP-19-P. Villanora, PA.: NCCLS.
U.S. Department of Defense. (1991). Anthropometry of U.S. military personnel. DOD-HDBK-743A. Washington, D.C.: DOD.
U.S. Department of Defense. (1992). Human engineering design criteria for military systems, equipment, and facilities. MIL-STD-1472D. Washington, D.C.; DOD.
Textbooks, Reports, & Articles
Backinger, C., & Kingsley, P. (1993). Write it right: Recommendations for developing user instruction manuals for medical devices used in home health care, (HHS Pub. FDA 93-4258). Rockville, MD.: U.S. Department of Health and Human Services, Food and Drug Administration.
Bailey, R. W. (1989). Human performance engineering: using human factors/ergonomics to achieve computer system usability (2nd Edition). Englewood Cliffs: Prentice Hall.
Bias, R., & Mayhew, D. (1994). Cost justifying usability. Cambridge, MA: Academic Press.
Bogner, M.S. (Ed.). (1994). Human error In medicine. Hillsdale, NJ: Lawrence Erlbaum Associates.
Brennan, T.A. (1991). Incidence of adverse events and negligence in hospitalized patients: Results of the harvard medical practice study I. New England Jounel of Medicine, Vol. 324, No. 6.
Brown, C.M. (1989). Human-computer interface design guidelines. Norwood, NJ: Ablex Publishing Company.
Callan, J.R., & Gwynne, J.W. (1993). Human factors principles for medical device labeling. San Diego, CA: Pacific Sciences & Engineering Group.
Callan, J.R., Kelly, R.T., Gwynne, J.W., Muckler, F.A., Saunders, W.M., Lepage, R.P., Chin, E., Schoenfield, I., & Serig, D.I. (1995) Human factors evaluation of remote afterloading brachytherapy: function and task analysis. (NUREG/CR-6125, Vol 2.) Washington, D.C.: U.S. Nuclear Regulatory Commission.
Cangelosi R., Carstensen, P., Crowley, J., et al. (1989). An analysis of medical device-related reporting of anesthesiology-related deaths. Rockville, MD: Food and Drug Administration (Internal Report).
Carstensen, P. (1986). FDA Issues Pre-Use Checkout. Anesthesia Patient Safety Foundations Newsletter, 1,13-17.
Cook, R.I., Woods, D.D., Howie, M.B., Horrow, J.C., & Gaba, D.M. (1992). Unintentional delivery of vasoactive drugs with an electromechanical infusion device. Journal of Cardiothoracic and Vascular Anesthesia, 6 (2), 238-244.
Dray, S.M., & Karat, C. (1994). Human factors cost justification for an internal development project. In Bias, G., & Mayhew, D.J. (Eds.). Cost-justifying ssability (pp 111-122). Boston: Academic Press.
Dumas, J., & Redish, J.C. (1993). A practical guide to usability testing. Norwood, N.J.: Ablex Publishing Corporation.
Emergency Care Research Institute (ECRI). (1992). Critical alarms: Patients at risk. Technology for Critical Care Nurses, 2 (1), 1-5.
Fischer, S.C., Blowers, P.A., & Bakowski, L.S. (1993). Using FDA recommendations to develop and test a home-user manual. Medical Device Diagnostic Industry. 15(11); 110-114.
Gopher, D., M., Badihi, Y., Cohen, G., Donchin, Y., Bieski, M., and Shamay, C. (1989, October). The nature and cause of human errors in a medical intensive care unit. Proceedings of the 33rd Annual Meeting of the Human Factors Society, October, 1989. Denver, CO: Human Factors Society.
Helander, M. (Ed.). (1991). Handbook of human-computer interaction (2nd Edition). Amsterdam: North Holland.
Henriksen, K., Kaye, R.D., Jones, R., Morisseau, D.S., & Serig, D.I. (1995). Human factors evaluation of teletherapy: Summary and prioritization of issues. Volume I. (NUREG/CR-6277). Washington, D.C.: U.S. Nuclear Regulatory Commission.
Horton, W. (1990). Designing and writing online documentation. New York: John Wiley and Sons, Inc.
Hyman, W.A. (1988). Human factors in medical devices. In Webster, J.G. (Ed.) Encyclopedia of medical devices and instrumentation, (Vol. 3., pp. 1542-1553). New York: Wiley.
Karat, C. (1992). Cost-justifying support on software development projects. Human Factors Society Bulletin. 35 (11). Santa Monica, CA.: Human Factors Society.
Kelly, R.T., Callan, J.R., Kozlowski, T.A., Meringola, E.D., & Meadows, S.K. (1990). Human factors in self monitoring of blood glucose: Final report. (NTIS PB 90-260886/AS). San Diego, CA: Pacific Science & Engineering Group for the FDA.
Kortstra, J.R.A. (1995). Designing for the user. Medical Device Technology. Jan/Feb pp 22-28.
Le Cocq, D. (1987). Application of human factors engineering in medical product design. Journal of Clinical Engineering, 12 (4), 271-277.
Loeb, G.L, Jones, B.R., Leonard, R.A., & Kendra, B. (1992). Recognition accuracy of current operating room alarms. Anesthesia & Analgesia, 75, 499-505.
Meister, D. (1985). Behavioral analysis and measurement methods. N.Y.: Wiley.
Meister, D. (1976). Behavioral foundations of system development. N.Y.: Wiley.
Nielson, J. (1993). Usability engineering. Chestnut Hill, MA: Academic-AP Professional.
Pickett, R. M., and Triggs, T.J. (1975). Human Factors in Heath Care. (Papers presented at an international symposim held in Lisbon, Portugal, June 1974. D.C. Heath and Company. ISBN 0-669-95885-9.
Reason, J. (1990). Human error. Cambridge, England: Cambridge University Press.
Rouse, W.B. (1980). Systems engineering models of human-machine interaction. New York: North Holland
Salvendy, G. (Ed.) (1987). Handbook of human factors. New York: John Wiley & Sons.
Sanders, M.S., & McCormick, E.J. (1993). Human factors in engineering design. New York: McGraw-Hill, Inc.
Sawyer, D., & Lowery, A. (1994). CDRH's role in promoting user-oriented design. Medical Device & Diagnostic Industry. (MD&DI),16 (3), 72-82.
Schneiderman, B. (1992). Designing the user Interface: Strategies for effective human-computer Interaction. Reading, MA: Addison-Wesley.
Spranger, D.M. (1988, April). Developing new products: Meeting the users' needs is the key to profits. Health Industry Today. 40-44.
Van Cott, H.P., & Kinkade, R.G. (Eds.). (1972). Human engineering guide to equipment design. Washington, D.C.: U.S. Government Printing Office.
Wiklund, M.E. (1991). Usability tests of medical products as prelude to the clinical trial. Medical Device & Diagnostic Industry (MD&DI), 3 (7), 13(7), 68-73.
Wiklund, M.E. (1993). How to implement usability engineering. Medical Device & Diagnostic Industry. (MD&DI), 15 (9),68-73.
Wiklund, M.E. (1995). Medical device and equipment design. Buffalo Grove, IL: Interpharm Press Inc.
Wiklund, M.E. (Ed.). (1994). Usability in practice--how companies develop user-friendly products. Cambridge, MA: Academic Press.
Woodson, W.E. (1981). Human factors design handbook. New York: McGraw Hill.







