Medical Devices
Guidance Document For Testing Bone Anchor Devices
This guidance was written prior to the February 27, 1997 implementation of FDA's Good Guidance Practices, GGP's. It does not create or confer rights for or on any person and does not operate to bind FDA or the public. An alternative approach may be used if such approach satisfies the requirements of the applicable statute, regulations, or both. This guidance will be updated in the next revision to include the standard elements of GGP's.
DRAFT
April 20, 1996
PLEASE FORWARD YOUR COMMENTS TO:
Orthopedic Devices Branch
Division of General and Restorative Devices
Center for Devices and Radiological Health
U.S. Food and Drug Administration
9200 Corporate Blvd.
Rockville, MD 20850
301-594-2036
I. CONTENTS
-
CONTENTS
-
PREFACE
-
ESTIMATION OF CLINICAL LOADS APPLIED TO THE DEVICE FOR COMPARISONS TO BENCH TEST RESULTS
-
PRECLINICAL TEST DATA
-
APPENDIX I.
-
APPENDIX II.
- APPENDIX III.
II. PREFACE
The purpose of this document is to recommend to the device manufacturer or sponsor of a future premarket notification (510k), Investigational Device Exemption (IDE), Premarket Approval (PMA) application, reclassification petition, or master file important information that should be provided to the FDA so that the substantial equivalence and/or safety and effectiveness of bone anchor or suture anchor devices can be determined.
Bone anchors or suture anchors are devices which attach soft tissue to bone. This may be achieved by tieing one end of a suture to soft tissue and the other end to a device which "anchors" the suture to the bone. In this document, "anchor" refers to the device the suture is attached to and "bone anchor" refers to the anchor plus attached suture (if there is a suture).
Suggestions and recommendations presented in this document are not mandatory requirements, but reflect data and methodologies which the Office of Device Evaluation (ODE), Division of General and Restorative Devices (DGRD) has determined to be acceptable. In this context, several points should be remembered:
- The guidance document is primarily intended to include scientific recommendations. Therefore, it suggests some important evaluation criteria, test procedures and end points. There may be circumstances where an alternative method or additional information may be useful and this document has included some examples. If the manufacturer or sponsor can answer the same scientific issues by means other than those included in this guidance document, they should not refrain from doing so. Because the scope of this document does not specify any particular type of bone/suture anchor device, some of the recommended test methods may need modification to address the properties of a particular product.
- The guidance document should be viewed as a living document. As scientific knowledge changes and scientific techniques are improved, FDA will periodically revise the document.
III. ESTIMATION OF CLINICAL LOADS APPLIED TO THE DEVICE FOR COMPARISONS TO BENCH TEST RESULTS
The maximum expected clinical loads applied to the device and tissues should be measured, calculated, or obtained from other scientific sources and compared to the results of the fixation strength tests described below. If this data is impossible to obtain, the strength of each healthy tissue indicated for repair by the bone anchor should be determined or calculated. This provides an upper limit for the maximum expected loads on the device. For example, if the fixation strength of the bone anchor approaches this upper load limit, clinical data may not be required.
IV. PRECLINICAL TEST DATA
The fixation strength of the bone anchor device should be evaluated in a bone model which simulates the density and bone structure of the bony site for which it is indicated. The results should be compared to a legally marketed predicate device or an established and validated surgical procedure using a suture (control) for the same intended use as the test device.
The loads to displace the test and control bone anchor devices from the bone model (e.g., cadaver bone) should provide precise data on failure loads and mechanisms that clinical data are unable to provide. However, these in vitro test methods may not adequately simulate actual in vivo conditions because of tissue strength variability between different joints and bony sites and the variability resulting from different surgical techniques. The number of anchors, orientation of the device and suture, patient parameters (e.g., applied loads) and other unknown variables complicate the usefulness of the results. Therefore, the worst case situation should be tested (e.g., the suture loaded perpendicular to the surface of the bone test sample).
If a previously cleared suture is packaged and sterilized with a bone anchor, the following should be provided:
suture diameter, needle attachment strength and knotted tensile strength if the suture was not sterilized prior to packaging with the anchor;
if already sterilized before packaging, the method of initial sterilization and a statement from the suture company that the second sterilization method is acceptable.
Documentation should be maintained regarding vendor certification for raw or semiprocessed source material, all manufacturing and quality control release procedures, and validation of sterilization procedures used in the manufacture of the suture. Any deviation of the source material or processing requires submission of a new premarket notification and Food and Drug Administration (FDA) clearance prior to commercial distribution of the modified device.
Clinical data may be required if the intended use, materials, design or some combination of these differ substantially from a legally marketed predicate device and increase risks to health (e.g., see Appendix 1). Examples might include new absorbable materials not previously cleared for implant use or a new design with a pull-out strength that is less than the predicate device.
To help FDA in its review and to facilitate a determination of substantial equivalence, a very brief summary of all mechanical test data should be organized into a table as suggested in Appendix 2. Any additional and important information not specifically mentioned should be inserted into this organization where appropriate. Detailed test reports from which the summarized data originated should be organized in a similar manner and included in the submission to FDA. The detailed reports should include, but are not limited to, the information contained in Appendix 3.
V. APPENDIX 1
POSSIBLE FAILURE MECHANISMS OF BONE ANCHORS
Loss of fixation is the primary cause of bone anchor failure. Loss of fixation may occur due to inadequate:
SOFT TISSUE STRENGTH
SUTURE STRENGTH
BONE OR ANCHOR STRENGTH
SUTURE FATIGUE RESISTANCE
ANCHOR FATIGUE RESISTANCE
VI. APPENDIX 2
The materials, methods and results of fixation strength testing should be organized as follows:
fixation strength test:
anchor type;
suture:
name,
material,
size,
number,
knot type
sterility
test subject or substrate:
type (e.g., human cadaver tibia),
cortical bone thickness and depth of bone at anchor implant site
quantitative description of microstructure, porosity and defects;
distance between drill holes;
presoaking time and medium;
angle of applied loads;
displacement rate;
other methods;
fixation strength;
failure mode (e.g., see appendix 1);
summary:
statistical evaluation of the differences between the results for the test and control specimens;
the maximum possible loads that the suture and anchor could experience under clinical conditions;
the strength of the healthy tissues indicated for repair;
all references (authors, title, etc.)
VII. APPENDIX 3
Detailed reports should be organized and subdivided into separate sections (some sections may be combined to enhance clarification) having the following headings (if applicable):
- Report title
- Investigators' names
- Facility Performing the test
Name
Address
Phone Number
- Dates
Test initiation
Test completion
Final report completion
- Objectives/Hypothesis
- Test and control samples
Sample selection criterion
Design
Materials
Processing methods
Differences between test samples, control samples and marketed device
- Methods and Materials
Test setup schematic or photograph
Description of grips or potting medium interfacing with samples
List of dependent, independent and uncontrolled variables, e.g.:
Test and control sample parameters
Environment composition, pH, volume, flow, temperature, replacement
Electromagnetic fields, applied charges, irradiation
Load directions, points of application and magnitudes
Times (e.g. rates, frequencies, number of cycles)
Other
Methods of specimen examination (e.g., failure analysis)
Statistical justification for the number of samples
Chronological description of the test procedures
Deviations from referenced protocols and standards
- Results
Time from manufacturing till testing commences
Discussion of the data and possible mechanisms
List of conclusions
Discussion of the objective/hypothesis
Simplifications and assumptions made and clinical implications of results
- Appendices
Experimental data
Calculations
Bibliography of all references pertinent to the report

