Topical Oxygen Chamber for Extremities - Class II Special Controls Guidance Document - Draft Guidance for Industry and FDA Staff

PDF Printer VersionDRAFT GUIDANCE

This guidance document is being distributed for comment purposes only.
Document issued on: April 6, 2006

Comments and suggestions regarding this draft document should be submitted within 90 days of publication in the Federal Register of the notice announcing the availability of the draft guidance. Submit written comments to the Division of Dockets Management (HFA-305), Food and Drug Administration, 5630 Fishers Lane, rm. 1061, Rockville, MD 20852. Alternatively, electronic comments may be submitted to Regulations.gov. All comments should be identified with the docket number listed in the notice of availability that publishes in the Federal Register.

For questions regarding this draft guidance document, contact Charles Durfor, Ph.D., at 240-276-3600, or by email at charles.durfor@fda.hhs.gov.

CDRH Logo

U.S. Department of Health and Human Services
Food and Drug Administration
Center for Devices and Radiological Health

Plastic and Reconstructive Surgical Devices Branch
Division of General, Restorative, and Neurological Devices
Office of Device Evaluation

Contains Nonbinding Recommendations
Draft - Not for Implementation

Preface

Additional Copies

Additional copies are available from the Internet, or to receive this guidance document via your fax machine, call the CDRH Facts-On-Demand system at 800-899-0381 or 301-827-0111 from a touch-tone telephone. Press 1 to enter the system. At the second voice prompt, press 1 to order a guidance document. Enter the guidance document number ( 1582) followed by the pound sign (#). Follow the remaining voice prompts to complete your request.

Table of Contents

  1. INTRODUCTION
  2. BACKGROUND
  3. THE CONTENT AND FORMAT OF AN ABBREVIATED 510(K) SUBMISSION
  4. SCOPE
  5. DEVICE DESCRIPTION
  6. RISKS TO HEALTH
  7. STERILITY
  8. FIRE AND EXPLOSION CONTROL
  9. OXYGEN PRESSURE CONTROL
  10. BIOCOMPATIBILITY
  11. ELECTRICAL SAFETY TESTING
  12. CLINICAL STUDIES
  13. LABELING

Draft Guidance for Industry and FDA Staff

Class II Special Controls Guidance Document: Topical Oxygen Chamber for Extremities

This draft guidance document, when finalized, will represent the Food and Drug Administration's (FDA's) current thinking on this topic. It does not create or confer any rights for or on any person and does not operate to bind FDA or the public. You can use an alternative approach if the approach satisfies the requirements of the applicable statutes and regulations. If you want to discuss an alternative approach, contact the FDA staff responsible for implementing this guidance document. If you cannot identify the appropriate FDA staff, call the appropriate number listed on the title page of this guidance document.

1. Introduction

This draft guidance document was developed as a special control guidance document to support the reclassification of the topical oxygen chamber for extremities (TOCE) into class II. The TOCE, as proposed, is intended to aid in the healing of chronic skin ulcers such as bedsores. This draft guidance will be issued in conjunction with a Federal Register notice announcing the proposal to reclassify this device type. This guidance is issued for comment purposes only. If a final rule to reclassify this device type is not issued, this guidance document will not be issued as a special control.

Following the effective date of a final rule reclassifying the device, any firm submitting a premarket notification (510(k)) submission for a TOCE will need to address the risks identified in the special control guidance. However, the firm need only show that its device meets the recommendations of the guidance document or in some other way provides equivalent assurances of safety and effectiveness.

 FDA's guidance documents, including this guidance document, do not establish legally enforceable responsibilities. Instead, guidance documents describe the Agency's current thinking on a topic and should be viewed only as recommendations, unless specific regulatory or statutory requirements are cited. The use of the word should in Agency guidance documents means that something is suggested or recommended, but not required.

The Least Burdensome Approach

This draft guidance document reflects our careful review of what we believe are the relevant issues related to the TOCE device and what we believe would be the least burdensome way of addressing these issues. If you have comments on whether there is a less burdensome approach, however, please submit your comments as indicated on the cover of this document.

2. Background

FDA believes that special controls, when combined with the general controls, will be sufficient to provide reasonable assurance of the safety and effectiveness of the TOCE. Thus, a manufacturer who intends to market a device of this generic type should (1) conform to the general controls of the Federal Food, Drug, and Cosmetic Act (the act), including the premarket notification requirements described in 21 CFR 807 Subpart E, (2) address the specific risks to health associated with TOCE identified in this guidance, and (3) obtain a substantial equivalence determination from FDA prior to marketing the device.

This special control guidance document identifies the classification regulation and product code for the TOCE (Please refer to Section 4. Scope). In addition, other sections of this special control guidance document list the risks to health identified by FDA and describe measures that, if followed by manufacturers and combined with the general controls, will generally address the risks associated with these TOCE and lead to a timely 510(k) review. This document supplements other FDA documents regarding the content requirements of a premarket notification submission. You should also refer to 21 CFR 807.87 and “ How to Prepare a 510(k) Submission” on FDA Device Advice.

As described in the guidance document entitled, The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications; Final Guidance, a manufacturer may submit a Traditional 510(k) or has the option of submitting either an Abbreviated 510(k) or a Special 510(k). FDA believes an Abbreviated 510(k) provides the least burdensome means of demonstrating substantial equivalence for a new device, particularly once a Class II Special Controls Guidance Document has been issued. Manufacturers considering certain modifications to their own cleared devices may lessen the regulatory burden by submitting a Special 510(k).

3. The Content and Format of an Abbreviated 510(k) Submission

An Abbreviated 510(k) submission must include the required elements identified in 21 CFR 807.87, including the proposed labeling for the device sufficient to describe the device, its intended use, and the directions for its use. In an Abbreviated 510(k), FDA may consider the contents of a summary report to be appropriate supporting data within the meaning of 21 CFR 807.87(f) or (g); therefore, we recommend that you include a summary report. The report should describe how this special control guidance document was used during the device development and testing and should briefly describe the methods or tests used and a summary of the test data or description of the acceptance criteria applied to address the risks identified in this document, as well as any additional risks specific to your device. This section suggests information to fulfill some of the requirements of section 807.87 as well as some other items that we recommend you include in an Abbreviated 510(k).

Coversheet

The coversheet should prominently identify the submission as an Abbreviated 510(k) and cite the title of this special controls guidance document.

Proposed labeling

Proposed labeling should be sufficient to describe the device, its intended use, and the directions for its use. (Please refer to Section 12. Labeling for specific information that should be included in the labeling for devices of the types covered by this guidance document.)

Summary report

We recommend that the summary report contain:

A description of the device and its intended use

We recommend that you describe the performance specifications and, when appropriate, include detailed, labeled drawings of the device. (Please refer to Section 5.Device Description for specific information that we recommend you include in the device description for devices of the type covered by this guidance document.) You should also submit an “indications for use” enclosure.1

Description of device design requirements

We recommend that you include a brief description of the device design requirements.

 Identification of the risk analysis method

We recommend that you identify the risk analysis methods you used to assess the risk profile, in general, as well as the specific device’s design and the results of this analysis. (Please refer to Section 6. Risks to Health for the risks to health generally associated with the use of this device that FDA has identified.)

Discussion of the device characteristics

We recommend that you discuss the device characteristics that address the risks identified in this class II special controls guidance document, as well as any additional risks identified in your risk analysis.

Description of the performance aspects

We recommend that you include a brief description of the test methods you have used or intend to use to address each performance aspect identified in Sections 7-11 of this class II special controls guidance document. If you follow a suggested test method, you may cite the method rather than describing it. If you modify a suggested test method, you may cite the method but should provide sufficient information to explain the nature of and reason for the modification. For each test, you may either (1) briefly present the data resulting from the test in clear and concise form, such as a table, or (2) describe the acceptance criteria that you will apply to your test results.2 (See also 21 CFR 820.30, Subpart C - Design Controls for the Quality System Regulation.)

Reliance on standards

If any part of the device design or testing relies on a recognized standard, we recommend that you include either:

  • a statement that testing will be conducted and meet specified acceptance criteria before the device is marketed, or
  • a declaration of conformity to the standard.3

Because a declaration of conformity is based on results from testing, we believe you cannot properly submit a declaration of conformity until you have completed the testing the standard describes. For more information, please refer to section 514(c)(1)(B) of the act and the FDA guidance, Use of Standards in Substantial Equivalence Determinations; Final Guidance for Industry and FDA.4

If it is not clear how you have addressed the risks identified by FDA or additional risks identified through your risk analysis, we may request additional information about aspects of the device’s performance characteristics. We may also request additional information if we need it to assess the adequacy of your acceptance criteria. (Under 21 CFR 807.87(l), we may request any additional information that is necessary to reach a determination regarding substantial equivalence.)

As an alternative to submitting an Abbreviated 510(k), you can submit a Traditional 510(k) that provides all of the information and data required under 21 CFR 807.87 and described in this guidance. A Traditional 510(k) should include all of your methods, data, acceptance criteria, and conclusions. Manufacturers considering certain modifications to their own cleared devices should consider submitting Special 510(k)s.

The general discussion above applies to any device subject to a special controls guidance document. The following is a specific discussion of how you should apply this special controls guidance document to a 510(k) for a TOCE.

4. Scope

The scope of this document is limited to the TOCE, in the proposed revision to 21 CFR 878.5650, product code to be designated, if a final rule is published.

21 CFR 878.5650 – Topical oxygen chamber for extremities.

  1. Identification. A topical oxygen chamber for extremities is a device that is intended to surround a patient’s limb and apply humidified oxygen topically at a pressure slightly greater than atmospheric pressure to aid healing of chronic skin ulcers such as bedsores.
  2. Classification. Class II (special controls). The special control for the device is FDA’s “Class II Special Controls Guidance Document: Topical Oxygen Chamber for Extremities.”

The full body hyperbaric chamber is not within the scope of this guidance document. It is a class II device, classified under 21 CFR 868.5470.

5. Device Description

We recommend that you identify your device, by the regulation and product code described in Section 4. Scope and include:

  • the composition for all device components
  • engineering drawings of the device
  • device design, including dimensions, shape, and final device specifications
  • a description of the device operation principles (e.g., the method of patient attachment, methods for controlling and monitoring gas pressure and methods for device sterilization or disinfection, as applicable, after patient use)
  • the method for assembling the device (e.g., the procedures for securing an air tight chamber and the attachment of any gas tubing or patient cushions to the chamber)
  • the methods for introducing and removing oxygen from the chamber in a safe and controlled manner
  • the methods for humidifying incoming oxygen
  • a description of any accessories used with the device (e.g., gas regulators, swage lock connectors, software control devices and gas pressure monitors)
  • a description of how the device is provided (e.g., sterile, assembled, for single use).

6. Risks to Health

In the table below, FDA has identified the risks to health generally associated with the use of a TOCE addressed in this document. The measures recommended to mitigate these identified risks are given in this guidance document, as shown in the table below. We recommend that you also conduct a risk analysis, before submitting your 510(k), to identify any other risks specific to your device and include the results in your 510(k). If you elect to use an alternative approach to address a particular risk identified in this guidance document, or have identified risks additional to those in the guidance, you should provide sufficient detail to support the approach you have used to address that risk.

Identified Risk

Recommended Mitigation Measures

Infection

Section 7. Sterility

Section 12. Clinical Studies

Section 13. Labeling

Fire and explosion

Section 8. Fire and Explosion Control

Section 13. Labeling

Local tissue damage

Section 9. Oxygen Pressure Control

Section 13. Labeling

Adverse tissue reaction Section 10. Biocompatibility
Electrical shock

Section 11. Electrical Safety Testing

Section 13. Labeling

7. Sterility

Wound infection impairs ulcer healing. Therefore, FDA believes that adequate sterilization of the initial product and sterilization or disinfection of devices after use or between uses is essential.

Regarding initial device sterilization, we recommend that the 510(k) contain the sterilization information requested in Updated 510(k) Sterility Review Guidance K90-1; Final Guidance for Industry and FDA.

For reusable devices, concerns exist about infection of a single patient during multiple uses with the same device or disease transmission after use of a single device on multiple patients. As discussed in Section 13. Labeling, the labeling should contain instructions for cleaning and disinfecting or sterilizing the patient-contacting materials after each use. Your 510(k) should describe how you validate the cleaning and disinfecting or sterilizing procedures. See also Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA Reviewer Guidance.

Additionally, because the medical grade oxygen introduced into the topical chamber may contain bacteria, fungi, or other infectious organisms, the 510(k) should describe the methods for sterile filtration of incoming gas.

8. Fire and Explosion Control

The oxygen-enriched atmosphere of a TOCE has an increased risk of fire and explosion. Sparks causing combustion can arise from:

  • the device’s electrical components
  • static electricity arising from the device components, a patient’s clothing or the treatment environment (e.g., excess oxygen exiting the chamber).

Therefore, we recommend you identify the methods used to reduce the risk of fire and explosion and the methods for introducing and removing oxygen gas from the chamber.

9. Oxygen Pressure Control

Excessive oxygen pressure (i.e., greater than 22mm of Hg) can occlude arterial circulation leading to decreased local tissue circulation and local tissue damage. Therefore, we recommend that you describe the methods used to control oxygen pressure within the chamber and demonstrate pressures do not exceed a safe limit.

If software is a part of the method for controlling oxygen pressure within the chamber, we recommend that you provide the information recommended in Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices(the Software Guidance), for a “moderate level of concern” device.

See also:

10. Biocompatibility

FDA recommends that you conduct biocompatibility testing as described in the FDA guidance Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part-1: Evaluation and Testing, for the patient-contacting components of the device. We recommend that you select biocompatibility tests (Parts 5 and 10 of ISO-10993) appropriate for the duration and level of contact consistent with the intended use of your device. We recommend you conduct testing on the final device (i.e., after manufacture, sterilization, and packaging for commercial distribution). If identical materials with identical material processing are used in a similar predicate device with the same type and duration of patient contact, you may identify the predicate device in lieu of providing biocompatibility testing.

11. Electrical Safety Testing

Inadequate electrical shielding and grounding can result in an electrical shock to the patient or fire/explosion.

We recommend you evaluate the electrical safety of your device, as well as its ability to function after exposure to environmental handling hazards. We recommend that you evaluate your device according to one or more of the standards below or equivalent methods.

  • International Electrotechnical Commission (IEC) 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety
  • Underwriters Laboratory (UL) 2601-1 Amendment 1 Medical Electrical Equipment: General Requirements for Safety

12. Clinical Studies

In accordance with the Least Burdensome provisions of act, the agency will rely upon well-designed bench, animal testing or both, rather than requiring clinical studies for new devices unless there is a specific justification for asking for clinical information to support a determination of substantial equivalence. While, in general, clinical studies will not be needed for most TOCE devices, FDA may recommend that you collect clinical data for a TOCE device with any one of the following:

  • indications for use dissimilar from legally marketed TOCE devices, of the same type  
  • designs dissimilar from legally marketed designs  
  • new technology, i.e., technology different from that used in legally marketed TOCE devices.

Examples of indications for use dissimilar from legally marketed TOCE devices at the time FDA issued this guidance are improved incidence of wound healing and prophylaxis or treatment of wound infection or pain reduction. Information about study design and interpretation of results for wound care products is described in the draft, Guidance for Industry Chronic Cutaneous Ulcer and Burn Wounds Developing Products for Treatment, which when finalized will represent FDA’s current thinking on wound care products..

FDA will always consider alternatives to clinical testing when the proposed alternatives are supported by an adequate scientific rationale. If a clinical study is needed to demonstrate substantial equivalence (i.e., conducted prior to obtaining 510(k) clearance of the device), the study must be conducted under the Investigational Device Exemptions (IDE) regulation, 21 CFR Part 812. In addition, sponsors of such trials must comply with the regulations governing institutional review boards (21 CFR Part 56) and informed consent (21 CFR Part 50).

After FDA determines that the device is substantially equivalent, clinical studies conducted in accordance with the cleared indications, including clinical design validation studies conducted in accordance with the quality systems regulation, are exempt from the investigational device exemptions (IDE) requirements. However, such studies must be performed in conformance with 21 CFR Part 56 and 21 CFR Part 50.

  13. Labeling

The 510(k) should include labeling in sufficient detail to satisfy the requirements of 21 CFR 807.87(e). The following suggestions are aimed at assisting you in preparing labeling that satisfies the requirements of 21 CFR 807.87(e).5

Directions for Use

As a prescription device, under 21 CFR 801.109, the device is exempt from having adequate directions for lay use. Nevertheless, we recommend including clear and concise instructions that delineate the technological features of the specific device and how the device is to be used on patients. Instructions should encourage institutional training programs designed to familiarize users with the features of the device and how to use it in a safe and effective manner.

Labeling should provide instructions for cleaning and disinfecting or sterilizing the device after each use. Refer to Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: FDA Reviewer Guidance. Such information can include, but may not be limited to:

  • pre-processing device handling methods (e.g., addressing biohazard concerns)
  • disassembly/reassembly methods
  • cleaning methods
  • cleaning/lubricating agents
  • rinsing techniques
  • disinfection or sterilization methods
  • special post-process handling
  • reuse life
  • warnings and precautions.

Labeling should also describe methods for venting oxygen from the TOCE to control the risk of fire or explosion.

Warnings

Labeling should warn that inadequate cleaning and disinfection or sterilization of the device after use may lead to transmission of infectious disease.

The label on the device should warn that inappropriate venting of oxygen from the TOCE can lead to fire or explosion.


 1. Refer to Indications for Use Form for the recommended format.

 2. If FDA makes a substantial equivalence determination based on acceptance criteria, the subject device should be tested and shown to meet these acceptance criteria before being introduced into interstate commerce. If the finished device does not meet the acceptance criteria and, thus, differs from the device described in the cleared 510(k), FDA recommends that submitters apply the same criteria used to assess modifications to legally marketed devices (21 CFR 807.81(a)(3)) to determine whether marketing of the finished device requires clearance of a new 510(k).

 3. See Required Elements for a Declaration of Conformity to a Recognized Standard (Screening Checklist for All Premarket Notification [510(K)] Submissions).

 4. Use of Standards in Substantial Equivalence Determinations; Final Guidance for Industry and FDA.

 5. Although final labeling is not required for 510(k) clearance, final labeling must comply with the requirements of 21 CFR Part 801 before a medical device is introduced into interstate commerce. In addition, final labeling for prescription medical devices must comply with 21 CFR 801.109. Labeling recommendations in this guidance document are consistent with the requirements of Part 801.

    
-
-
-
-
-