Kaiser Permanente-East 15-Oct-03
|
Public Health Service |
|
Cincinnati District Office |
WARNING LETTER
CIN-03-19031
October 15, 2003
Sent by Federal Express
Dr. Antonio Motta
Medical Director
Kaiser Permanente - East
12301 Snow Rd.
Parma, OH 44130
RE: Facility MQSA I.D. #: 221837
Cleveland Heights Medical Center
10 Severance Circle
Cleveland Heights, OH 44118
Dear Dr. Motta:
A representative from the State of Ohio acting on behalf of the
Food and Drug Administration (FDA)
inspected your facility (Cleveland Heights Medical Center, 10 Severance Circle,
Cleveland Heights, OH
44118) on September 8, 2003. This inspection revealed a serious regulatory problem
involving the
mammography at your facility.
Under the Mammography Quality Standards Act (MQSA) of 1992, 42 U.S.C. § 263b, and the regulations set forth in Title 21, Code of Federal Regulations (C.F.R.), Part 900, your facility must meet specific requirements for mammography. These requirements help protect the health of women by assuring that a facility can perform quality mammography.
The inspection revealed the following Level 1 finding at your
facility (identified on your inspection
report):
Medical records and mammography reports - Communication of mammography results to the patients (21 C.F.R § 900.12(c)(2))
Your records revealed that your facility failed to provide lay
summary letters to patients that included the results of each mammography examination
written in lay terms within 30 days of the examinations. During the inspection,
there were at least three patient film records waiting for comparison films
whose mammography examination dates had been more than 30 days. The mammography
records indicated the following examination dates: July 29, August 5 & 6,
2003. Your facility failed to ensure that these mammograms are interpreted and
results are communicated to the patients within thirty days of the date of examinations.
Your facility failed to follow your own Mammography Policy/Procedure which states:
In the event that outside comparison images must be obtained, a letter
of results is sent within 30 days of the mammogram.
The inspection also revealed the following Level 2 violation at your facility (identified on your inspection report):
Quality Assurance - Equipment (21 CFR 900.12(e)(8)(ii)(A), as required in 21 CFR 900.12(e)(2)(ii))
Your 2003 phantom quality control record revealed on July 8, 2003, the background density was
charted outside the allowable regulatory limit and your facility failed to document corrective action taken before further mammography exams.
Because these conditions may be symptomatic of serious underlying problems
that could compromise
the quality of mammography at your facility, they represent serious violations
of the law that may result
in FDA initiating regulatory action. These actions include, but are not limited
to, placing your facility
under a Directed Plan of Correction; charging your facility for the cost of
on-site monitoring; seeking
civil money penalties of up to $10,000 for each failure to substantially comply
with, or each day of
failure to substantially comply with, the MQSA Standards; seeking suspension
or revocation of your
facilitys FDA certificate; or seeking a court injunction against performing
further mammography (see
42 USC § 263b(h) through (j) of the MQSA). Before initiating any enforcement
action against you,
FDA may perform a follow-up inspection of your facility to determine whether
these problems have
been corrected. FDA assesses fees for MQSA inspections, including follow-up
inspections 42 USC
263b(r).
It is necessary to act on these matters immediately. Please explain to this office in writing within fifteen
(15) working days from the date you received this letter:
1. The specific steps you have taken to correct the violations noted in this letter; and
2. Each step your facility is taking to prevent the recurrence of similar violations.
Please also provide the following in answering this letter: (Note: Patient
names or patient social security
numbers should be deleted from any copies submitted)
1. A copy of your corrected written procedure for communication of mammography results ensuring that patients will receive a lay summary letter within 30 days of mammography examinations. The written procedure shall also include communication of Suspicious and Highly suggestive of malignancy mammography results to the patients as soon as possible (normally within three to five days) from the date of examination. The written procedures shall cover all mammography results for patients with or without comparison films. Also, include a copy of sample lay summary letters for each final assessment categories.
2. A copy of your written procedure for documentation of corrective actions when the phantom
QC parameter(s) is operating out of limits.
Please submit your response to:
Mr. R. Terry Bolen
MQSA Compliance Officer
Food & Drug Administration
6751 Steger Drive
Cincinnati, OH 45237-3097
FAX: 513-679-2772
Also, please send a copy to the State radiation control offtce:
Ms. Terri Eckert
Ohio Department of Health
Radiological Technology Section
161 South High St., Suite 400
Akron, OH 44308-1616
Finally, you should understand that there are many FDA requirements pertaining
to mammography. This letter pertains only to certain findings of your inspection
and does not necessarily address other obligations you have under the law. You
may obtain general information about all of FDAs requirements for mammography
facilities by contacting the Mammography Quality Assurance Program, Food and
Drug Administration, P.O. Box 6057, Columbia, MD 21045-6057, (1-800-838-7715)
or through the Internet at http://www.fda.gov/cdrh/mammography.
If you have more specific questions about mammography facility requirements,
or about the content of this letter, please feel free to contact R. Terry Bolen at 513-679-2700, extension
138.
Sincerely yours,
/S/
Carol A. Heppe
District Director
Cincinnati District Office





Department
of Health and Human Services