Inspections, Compliance, Enforcement, and Criminal Investigations
Plus Orthopedics 03-Oct-03
|
Public Health Service |
|
Center for Devices and Radiological Health |
Via Federal Express
October 3, 2003
WARNING LETTER
David R. Lorenzi
President/CEO
Plus Orthopedics
6055 Lusk Boulevard
San Diego: California 92121
Dear Mr. Lorenzi:
The purpose of this letter is to inform you of objectionable practices and activities found during a Food and Drug Administration (FDA) inspection and to acknowledge your June 1, 2003: written response addressed to Alonza Cruse, Los Angeles District Director, FDA. Your letter was in response to the Form FDA 483 Inspectional Observations that was issued at the close of the inspection of Plus Orthopedics. The inspection took place during the period of April 14 through 23, 2003, and was conducted by Ms. Yvette E. Guillermo, an investigator from FDAs Los Angeles District Office. The purpose of the inspection was to determine whether your firm's activities as sponsor of an investigational study of [redacted] comply with applicable FDA regulations. This product is a device as defined in section 201 (h) of the Federal Food, Drug, and Cosmetic Act.
Our review of the inspection report prepared by the district office revealed serious violations of the requirements of Title 21 Code of Federal Regulations (21 CFR), Part 812 - Investigational Device Exemptions. You received a Form FDA 483. Inspectional Observations. at the conclusion of-the inspection that listed the deviations noted and discussed with you; [redacted].
We acknowledge the corrective actions taken as indicated in !your annual IDE report G010120/S6. dated January 20, 2003 - before FDAs inspection - and in your written response to the Form FDA 483, dated June 1, 2003. You also submitted a draft Audit/Corrective Action Operating Plan on July 21, 2003. However, some of your responses to date do not describe an adequate corrective action plan with specific times for completion and supporting documentation for corrections already made. The deviations noted on the Form FDA 4X3 and our subsequent review of the inspection report and your responses are discussed below.
Failure to obtain IDE and Investigational Review Board (IRB) approval before initiating the study. (21 CFR 812.20(a), 812.30(a), 812.40, and 812.42)
An investigation involving a significant risk device, such as the [redacted] cannot be conducted without an approved IDE. (21 CFR 812.20(a), 812.30(a), and 812.42.) The sponsor is also responsible for ensuring that necessary IRB approval is obtained. (21 CFR 812.40.)
You failed to submit an IDE application to FDA or obtain approval before shipping some of the investigational devices. You also failed to ensure that clinical investigators participating in the study obtained IRB approval prior to the use of the investigational device. As a result, thirty-one investigational devices were implanted without FDA and IRB approval of the study.
Failure to adhere to investigational device exemption (IDE) requirements
including ensuring control of the investigational device, monitoring the study
properly,and obtaining investigator agreements. (21 CFR 812.40, 812.43, and
812.45)
As the sponsor of an investigation, you are responsible for shipping the device
only to qualified participating investigators (21 CFR 812.43(b)), monitoring
the study properly (21 CFR 812.40), selecting qualified investigators and providing
them with sufficient information to conduct the investigation properly (21 CFR
812.40 and 812.45), obtaining financial disclosure agreements from each investigator
(21 CFR 812.43(c)(5)), and ensuring investigators commitment to satisfy
applicable informed consent requirements
(21 CFR 812.43(c)(4)(iii)).
Your firm failed to do the following: ensure proper monitoring of the study; and obtain commitments that investigators would comply with the investigation plan and informed consent requirements. As a result:
-
twelve investigational devices were implanted using a [redacted] that was not part of the test device and not included in the investigational plan;
-
nine investigational devices were implanted in patients not enrolled in the study (five of the nine devices were implanted at unapproval clinical sites);
-
some investigators participating, in the Investigation did not receive current copies of the investigational plan:
-
the investigator agreement for one investigator was signed after implantation of four subjects with the investigational device
-
seventeen patients signed an informed consent form that deviated from the IRB-approved version; and
-
subjects at four different clinical sites signed an informed consent form that failed to include randomization information.
In addition, you failed to obtain financial disclosure agreements for each investigator.
During the inspection, [redacted] indicated that [redacted] at the time of
patient implants with the unapproved [redacted] resigned from Plus Othropedics.
[redacted] also stated that Plus Orthopedics received FDA approval on February
20, 2003 to use the [redacted] in the study. [redacted] stated that your company
is currently ensuring compliance at the clinical sites through monitoring, and
[redacted] stated that the [redacted] site closed its Orthopedic Department
on April 1, 2003 and is no longer enrolling patients.
In your June 1, 2003 written response, you attributed some of the deficiencies noted above to a lack of understanding of FDA's IDE and device regulations by prior management. You stated that management was replaced, and you committed to temporarily suspend enrollment of new subjects while you work toward compliance with FDA regulations. You also committed to multiple corrective actions, including having the study audited by a third party auditor and notifying patients who received the investigational device outside of the study. You also stated that several corrective actions have already been implemented: unused investigational devices have been removed from investigation sites; employees have received training on the [redacted], [redacted] and [redacted] study monitoring is currently being performed by an outside consultant; and all investigational sites have received the latest investigational plan.
Your responses to observations 1, 2, 3(a), 4 and 5 states the steps you plan
to take or have taken to prevent future occurrences, but in most cases your
responses do not adequately address how you will correct and prevent future
deviations. For example, during the
inspection close-out discussion, Ms. Guillermo discussed your failure to follow
your new procedure for distributing the investigational device to investigators
[redacted]. The discrepancies identified included the failure to ship some test
devices in kits, the dates on some [redacted] forms not matching dates on the
[redacted] log, and the completion of some [redacted] forms incorrectly by warehouse
personnel. You must provide additional information to confirm that these problems
have been addressed, such as the latest, approved SOP for this process, and
copies of your device accountability log.
Your responses to Form FDA 483 observations 3(b) and 3(c) state that your procedures
for monitoring the clinical trial have been revised that your [redacted] has
completed training in monitoring, and that an outside consultant is currently
monitoring the study.
Your response, however, does not specifically address how you plan to ensure
that the study is adequately monitored. You must identify the specific steps
you plan to take to ensure proper monitoring, submit documentation of the monitoring
visits for each clinical site, and verify that the current consultant is adequately
monitoring the study.
In addition, during the inspection closeout discussion and in your written
response, it was stated that a modified investigator agreement, including a
financial disclosure agreement, has been sent to all investigators. You also
stated during the close-out
discussion that the original investigator agreement for [redacted] was signed
before patients were enrolled, but that it was misplaced. You stated that [redacted]
signed another agreement on December 20, 2001, and that a memo was placed in
[redacted] file. Please submit copies of each of these signed investigator agreements.
Based on follow-up conversations with [redacted] and on your June 1, 2003 response, we also understand that the company is preparing a notification to be sent to patients who did not participate in the investigation but who were implanted with the investigational device. Please submit copies of this notification, including copies signed by the recipients, as well as the well as the re-signed informed consent forms for the [redacted] a site and the IRB-approved informed consent forms referent to observation 5 in the Form FDA 483, In addition, please submit copies of any IRB correspondence indicating approval of your revised informed consent forms and any other pertinent documentation.
During the inspection and close-out discussion, Ms. Guillermo discussed with
you and [redacted] several protocol deviations, including the treatment of two
patients not meeting the specified age criteria of 18-75 years, three patients
above the maximum body
mass index limit of 40, and three patients who received treatment of both [redacted]
with the investigational device. [redacted] stated that there was confusion
among clinical investigators concerning treatment of subjects for [redacted],
and that the company intends to submit another IDE supplement to FDA. Please
describe the steps taken to minimize such protocol deviations in the future,
and any steps taken to address confusion regarding use of the device for [redacted].
Failure to maintain accurate, complete, and current records relating to the investigation (21 CFR 812.140(b))
It is the sponsors responsibility to maintain accurate, complete and current records relating to an investigation. 21 CFR 812.140. Your firm did not maintain,adequate documentation of pre-study monitoring visits. For example, there is a lack of documentation of these visits for [redacted]. There is also incomplete documentation of pre-study monitoring visits at [redacted] since the clinical investigator did not sign the necessary do at these visits.
The violations listed above are not intended to be an all-inclusive list of
objectionable practices that may exist. The sponsor is responsible for adhering
to each applicable requirement of the Federal Food, Drug, and Cosmetic Act (21
USC 321 et seq.) and all
pertinent federal regulations.
We recognize that you have been working with the Office of Device Evaluation
in FDAs Center for Devices and Radiological Health (CDRH) and the Institutional
Review Boards to implement corrective actions, and that your efforts to complete
your corrective action
plan are ongoing. This letter does not supercede the obligations you previously
committed to take to solve data integrity problems. Plus Orthopedics should
continue to work with FDA and the IRBs to satisfactorily complete the required
auditing, implementation of the corrective action plan, and other responsibilities
arising in connection with this study.
You must provide this office with written documentation of the additional specific
steps you have taken or plan to take to correct these violations and bring your
study activities into compliance with FDA regulations and to prevent recurrence
of similar violations.
Failure to do so could result in regulatory action without further notice.
Please address your correspondence to the Food and Drug Administration, Center for Devices and Radiological Health, Office of Compliance, Division of Bioresearch Monitoring, Program Enforcement Branch II (HFZ-312), 2098 Gaither Road, Rockville, Maryland 20850, Attention: Linda Godfrey.
A copy of this letter has been sent to FDAs Los Angeles District Office, 19900 MacArthur Blvd., Suite 300, Irvine, California 92612. We request that copies of your response be sent to FDAs Los Angeles District Office and the Office of Device Evaluation, Attention: Carl DeMarco, 9200 Corporate Boulevard, HFZ-400, Rockville, Maryland 20850.
If you have any questions, feel free to contact Linda Godfrey at (301) 594-4723 extension 134.
Sincerely yours,
/s/
Michael Marcarelli for Timothy A. Ulatowski
Director
Office of Compliance
Center for Devices and Radiological Health





Department
of Health and Human Services