On June 7 through July 2, 2002, Investigator Cynthia Jim of our office conducted an inspection of Presbyterian Hospital's blood bank. Our inspection documented deviations from the Current Good Manufacturing Practice Regulations, Title 21, Code of Federal Regulations (21 CFR) Parts 600 - 680. These deviations cause your blood products to be adulterated within the meaning of Section 501 (a)(2)(B) of the Federal Food, Drug, and Cosmetic Act ("Act"). Deviations noted include:
Failure to review all records pertinent to the lot or unit maintained pursuant to the regulations before the release or distribution of a lot or unit of final product, as required by 21 C.F.R. 606.100(c) (Observation Number 1). For example:
Numerous errors were noted in the "Quality Control - Leukoreduced Blood Products - Whole Blood/Red Cells" records as well as m the Irradiation Logs reviewed by our investigator. Errors noted included miscalculations in the "% Product Recovered," and the "Residual WBC/uL." Also. there were many instances of blank spaces noted 111 the records. These errors and omissions were not detected by your quality control unit although a second review of the records had occurred
Failure to adequately train personnel responsible for the collection, processing, compatibility testing, storage, or distribution of blood or blood components to assure competent performance of their assigned functions and to ensure that the final product has the safety, purity, potency, Identity, and effectiveness it purports or is represented to possess, as required by 21 C.F.R. 606.20(b) (Observation Numbers 2, 4, 6, 11) For example:
A unit was collected from a donor listed in the [redacted] as temporarily deferred.
The database had not been referenced before initiating the donation. There is
no assurance that employees are verifying the deferral status prior to conducting
donations.
Also, employees processing whole blood units into Fresh Frozen Plasma and Leukoreduced Red Blood Cells documented the weight of the product in the [redacted] instead of the volume, as required.
The blood bag weight used in the calculations was also found to be inaccurate,
resulting in incorrect final volume results. The technician documented the empty
weight of the bag as [redacted] grams when, in actuality, the bag was found
to weigh [redacted] grams.
Failure to document the performance of each significant step in the collection,
processing, compatibility testin g, storage, and distribution of each unit of
blood and blood components so that all steps can be clearly traced, as required
by 21 C.F.R. 606.160(a)(1) (Observation Numbers 3, 5, 10, 12). For example:
A blood bank technician completed data entry in the [redacted] including the
time irradiated, before the actual irradiation.of the units.
There were numerous data entries that were found to be crossed-out in the log without an explanation.
The original documentation of the manufacturing steps was not entered directly into the [redacted] out was observed to be written on paper which was later transcribed to the log. There is no assurance that such information is transcribed accurately as the original documentation is discarded.
Failure to document that Fresh Frozen Plasma ("FFP") has been processed and placed m a freezer within 8 hours of collecnon, or within the timefiame specified for use for the blood collecting, processing, and storage system, as required by 21 C.F.R. 640.34(b) (Observation Number 6). For example:
Whole blood unit [redacted] was collected on June 3, 2002. There is no indlcatlon in the [redacted] that shows when the FFP was created. The [redacted] only shows that the products were available on June 6, 2002.
Whole blood unit [redacted] was collected on June 4, 2002. There is no indication
when the FFP was created.
Whole blood unit [redacted] was collected on June 6, 2002. There ts no indlcanon when the FFP was created. The [redacted] only indicates that the products were available on June 7, 2002.
Failure to mainiain adequate processmg records to demonstrate all steps in
blood processing, as required by 21 C F.R. 606.160(b) (Observation Number 7).
For example.
Review of your Records of Irradiated Blood Products revealed instances where
several units were processed at the same tune, although the irradiation contamer
cannot hold more than one unit at a tulle. Instances include May 10, 2002 [redacted],
irradiated at 2300); February 26, 2002 [redacted], irradiated at 1400); January
12, 2002 (units [redacted]) irradiated at 2050); and November 13, 2001 (units
[redacted] irradiated at 0830).
On several dates, the color change of the [redacted] was not documented or
was inadequately documented. These include September 25, 2001, September 27,
2001, October 7, 2001, October 26, 2001, October 31, 2001, November 1, 2001,
November 27.
2001, December 17, 2001, February 26, 2002, March 20, 2002, April 7, 2002, April
9, 2002, June 3, 2002, and June 10, 2002.
Failure to assure that the irradiator is perfomling in the manner for which
it was designed, as required by 21 C.F.R. $ 606,60(c) (Ob servation Number 8).
For example:
Our inspection revealed that the irradiation timer setting was calculated to give a minimum dose of [redacted] cGy of radiation to an empty container instead of a full container. There was no explanation why the timer setting was calculated for an empty container.
Failure to maintain standard operating procedures to ensure the proper collection,
processing, compatibility testing, quarantine, storage, handling, and disposition
of blood and blood components, as required by 21 C.F.R. 606.100(b) (Observation
Number 9). Your Standard Operating Procedures ("SOPs") have not been
updated to reflect current operation procedures
For example:
The SOP for Leukoreduced Red Cells requires a minimum of [redacted] residual
white blood cells per blood component instead of [redacted]. Also, your SOP
requires QC to be perfornied on the [redacted] units of red cells, but your
blood bank is currently
performing QC on [redacted] filtered units.
The SOP, "Irradiating Blood Components," does not indicate the maxmum
number of units of blood or blood components that can be irradiated at one time.
This procedure states, "If irradiating multiple products, place the units
in the canister so that the
indicators are in the middle."
We acknowledge receipt of your July 18, 2002, correspondence responding to the Form FDA-483 issued on July 2: 2002. Review of your response found it to be inadequate. The proposed corrective actions do not include an effective quality control system to assure that future deviations will be detected and corrected. Although in some cases; you have developed charts outlining the steps to be taken, there were no accompanying, written procedures. More specifically, we have the following comments:
-Attachment 7 includes a table showmg the tare weight of empty bags to be used
when computing the volume of blood components. This chart lists the weight of
Leultorrap SCRC bags as [redacted] grams. During the mspectlon, our mvestlgator
had your technician weigh an empty bag. The weight was found to be, [redacted]
grams. Please be sue that the weights listed reflect the correct value, as this
can have a significant impact on the final product.
- Regarding your procedure, "Irradiating Blood Components Using the [redacted]
your procedure references the maximum number of componenets per irradiation
cycle as "...based on [redacted]. The procedure lists the maximum number
as [redacted]. During our inspection, Investigator Jim was informed that only
[redacted] RBC unit collected internally could fit in the container and that
[redacted] units of RBCs obtained from UBS was a very tight squeeze. As discussed
during the inspection, the validation of your irradiator must reflect the actual
cappacity of the equipment and the operations employed by your facility.
- The majority of the attachments are identified as "draft" and do
not contain signatures indicating formal acceptance. The timelines given indicate
that all of the corrective actions should have been completed by September 30,
2002. Therefore it is expected that all
corrections have been finalized, approved, and are in place at this time.
- Many of the observations made by Investigator Jim involve recording errors or missing data: as well as lack of adherence to Standard Operating Procedures that should have been caught by your quality assurance review, but were not. It is imperative that your entire staff, including those who perfonn the quality review of records, be properly trained. Procedures are only effective if they are followed.
Your response states that [redacted] as been engaged to teach GMP training
to your employees. We suggest that you also obtain the services of an independent,
outside consultant to evaluate your procedures and to determine the compliance
status of your facility.
Our investigator also noted that the position of the QA Coordinator has been
vacant for approximately one year and that employees have been conducting the
quality control of each others' work. In order to assure objectivity, an independent
review of the records should be performed.
The above-identified deviations are not intended to be an all-inclusive list
of deficiencies at your facaclity. It is your responsibility to assure that
your establishment complies with all requirements of the federal regulations.
You should take prompt action to con-ect these violations. Failure to do so
may result in regulatory action without further notice, including seizure and/or
injunction.
You should notify this office in writing, within fifteen working days of receipt
of this letter: of the specific steps you have taken to correct the noted violations
and to prevent their recurrence. If corrective action cannot be completed within
fifteen working days, state the reason for the delay and the time withm which
the corrections will be completed.
Your response should be sent to Regina A. Barrell, Compliance Officer, Food
and Drug Administration. Denver District, P.O. Box 25087, Denver, Colorado 80225-0087.
If you have any further questions: please feel free to contact Ms. Barrell at
(303) 236-3043.