Marin Breast Health Center 19-Oct-01
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
San Francisco District
1431 Harbor Bay Parkway
Alameda, CA 94502-7070
Telephone: 510-337-6700
WARNING LETTER
October 19, 2001
Via Federal Express
Our Reference: 2951753
MQSA Facility ID: 122051
Inspection ID: 1220510008
Denise Fong, Clinic Supervisor
Marin Breast Health Center
1240 South Eliseo Drive, Ste. 101
Greenbrae, CA 94904
Dear Ms. Fong,
We are writing to you because on October 2, 2001, your facility was inspected by a
representative of the State of California, acting on behalf of the Food and Drug
Administration (FDA). This inspection revealed a serious regulatory problem involving
the mammography at your facility.
Under a United States Federal law, the Mammography Quality Standards Act of 1992,
your facility must meet specific requirements for mammography. These requirements
help protect the health of women by assuring that a facility can perform quality
mammography. The inspection revealed the following: repeat Level 2 findings at your
facility:
1. Corrective action before further exams, for a failing image score, or a phantom
background optical density, or density difference outside the allowable regulatory
limits, was not documented for unit 3, [redacted] room 2
2. Corrective action before further exams, for a failing image score, or a phantom
background optical density, or density difference outside the allowable regulatory
limits, was not documented for unit 4, [redacted], [redacted] Room 1 [redacted].
3. Corrective action before further exams, for a failing image score, or a phantom
background optical density, or density difference outside the allowable regulatory
limits, was not documented for unit 5 [redacted], Room 3 [redacted].
The specific problems noted above appeared on your MQSA Facility Inspection Report,
which was issued to your facility at the close of the inspection.
Because these conditions may be symptomatic of serious underlying problems that could
compromise the quality of mammography at your facility, they represent a serious
violation of the law which may result in FDA taking regulatory action without further
notice to you. These actions include, but are not limited to, placing your facility under a
Directed Plan of Correction, charging your facility for the cost of on-site monitoring,
assessing civil money penalties up to $10,000 for each failure to substantially comply
with, or each day of failure to substantially comply with, MQSA Standards, suspension or
revocation of your facility?s FDA certificate, or obtaining a court injunction against further mammography.
In addition, your response should address the Level 2 findings that were listed on the
inspection report provided to you at the close of the inspection. These Level 2 findings
are:
- Phantom OC records were missing for at least two weeks but less than four weeks for unit 3, [redacted] Room 2 [redacted].
- Phantom OC records were missing for at least two weeks but less than four weeks for Unit 4, [redacted] Room 1 [redacted].
- Phantom OC records were missing for at least two weeks but less than four weeks for unit 5, [redacted] Room 3 [redacted].
4. Processor QC records in the month of 08/2001 were missing for at least 10% but less than 30% of operating days, for processor 1 [redacted] Other, room Daylight at site
Main Breast Health Center 3.
Processor QC records were missing at least 2 but less than 5 consecutive days for
processor 1, [redacted] Other, room Daylight at site Marin Breast Health Center.
It is necessary for you to act on this matter immediately. Please explain to this office in
writing within fifteen (15) working days from the date you received this letter:
- the specific steps you have taken to correct all of the violations noted in this letter;
- each step your facility is taking to prevent the recurrence of similar violations;
- equipment settings (including technique factors), raw test data, and calculated final results, where appropriate; and
- sample records that demonstrate proper record keeping procedures, if the findings relate to quality control or other records (Note: Patient names or identification should be deleted from any copies submitted).*
*This note is not applicable for letters which also address patient notification
Please submit your response to:
Russell A. Campbell, Compliance Officer
San Francisco District
U.S. Food and Drug Administration
1431 Harbor Bay Parkway
Alameda, CA 94502
Finally, you should understand that there are many FDA requirements pertaining to
mammography. This letter pertains only to findings of your inspection and does not
necessarily address other obligations you have under the law. You may obtain general
information about all of FDA?s requirements for mammography facilities by contacting
the Mammography Quality Assurance Program, Food and Drug Administration, P.O.
Box 6057, Columbia, MD 21045-6057 (1-800-838-7715) or through the Internet at
http://www.fda.gov.If you have more specific questions about mammography facility requirements, or about
the content of this letter, please feel free to contact Russell A. Campbell at 510-337-6861.
Sincerely yours,
Charles D. Moss, Acting for
Dennis K. Linsley
District Director