WARNING LETTER

October 19, 2001

Via Federal Express

Our Reference: 2951753

MQSA Facility ID: 122051

Inspection ID: 1220510008

Denise Fong, Clinic Supervisor

Marin Breast Health Center

1240 South Eliseo Drive, Ste. 101

Greenbrae, CA 94904

Dear Ms. Fong,

We are writing to you because on October 2, 2001, your facility was inspected by a

representative of the State of California, acting on behalf of the Food and Drug

Administration (FDA). This inspection revealed a serious regulatory problem involving

the mammography at your facility.

Under a United States Federal law, the Mammography Quality Standards Act of 1992,

your facility must meet specific requirements for mammography. These requirements

help protect the health of women by assuring that a facility can perform quality

mammography. The inspection revealed the following: repeat Level 2 findings at your

facility:

1. Corrective action before further exams, for a failing image score, or a phantom

background optical density, or density difference outside the allowable regulatory

limits, was not documented for unit 3, [redacted] room 2

2. Corrective action before further exams, for a failing image score, or a phantom

background optical density, or density difference outside the allowable regulatory

limits, was not documented for unit 4, [redacted], [redacted] Room 1 [redacted].

3. Corrective action before further exams, for a failing image score, or a phantom

background optical density, or density difference outside the allowable regulatory

limits, was not documented for unit 5 [redacted], Room 3 [redacted].

The specific problems noted above appeared on your MQSA Facility Inspection Report,

which was issued to your facility at the close of the inspection.

Because these conditions may be symptomatic of serious underlying problems that could

compromise the quality of mammography at your facility, they represent a serious

violation of the law which may result in FDA taking regulatory action without further

notice to you. These actions include, but are not limited to, placing your facility under a

Directed Plan of Correction, charging your facility for the cost of on-site monitoring,

assessing civil money penalties up to $10,000 for each failure to substantially comply

with, or each day of failure to substantially comply with, MQSA Standards, suspension or

revocation of your facility?s FDA certificate, or obtaining a court injunction against further mammography.

In addition, your response should address the Level 2 findings that were listed on the

inspection report provided to you at the close of the inspection. These Level 2 findings

are:

  1. Phantom OC records were missing for at least two weeks but less than four weeks for unit 3, [redacted] Room 2 [redacted].
  2. Phantom OC records were missing for at least two weeks but less than four weeks for Unit 4, [redacted] Room 1 [redacted].
  3. Phantom OC records were missing for at least two weeks but less than four weeks for unit 5, [redacted] Room 3 [redacted].

4. Processor QC records in the month of 08/2001 were missing for at least 10% but less than 30% of operating days, for processor 1 [redacted] Other, room Daylight at site

Main Breast Health Center 3.

Processor QC records were missing at least 2 but less than 5 consecutive days for

processor 1, [redacted] Other, room Daylight at site Marin Breast Health Center.

It is necessary for you to act on this matter immediately. Please explain to this office in

writing within fifteen (15) working days from the date you received this letter:

  • the specific steps you have taken to correct all of the violations noted in this letter;
  • each step your facility is taking to prevent the recurrence of similar violations;
  • equipment settings (including technique factors), raw test data, and calculated final results, where appropriate; and
  • sample records that demonstrate proper record keeping procedures, if the findings relate to quality control or other records (Note: Patient names or identification should be deleted from any copies submitted).*

*This note is not applicable for letters which also address patient notification

Please submit your response to:

Russell A. Campbell, Compliance Officer

San Francisco District

U.S. Food and Drug Administration

1431 Harbor Bay Parkway

Alameda, CA 94502

Finally, you should understand that there are many FDA requirements pertaining to

mammography. This letter pertains only to findings of your inspection and does not

necessarily address other obligations you have under the law. You may obtain general

information about all of FDA?s requirements for mammography facilities by contacting

the Mammography Quality Assurance Program, Food and Drug Administration, P.O.

Box 6057, Columbia, MD 21045-6057 (1-800-838-7715) or through the Internet at

http://www.fda.gov.

If you have more specific questions about mammography facility requirements, or about

the content of this letter, please feel free to contact Russell A. Campbell at 510-337-6861.

Sincerely yours,

Charles D. Moss, Acting for

Dennis K. Linsley

District Director

 

 

-
-