M-b-346: Honeywell DR4500A Truline Series HTST, STLR and Pasteurization Flow Circular Chart Recorder/Controller

<< Back to Index of Memoranda of Information (M-b) Main Page  

March 30, 2004

TO: All Regional Food and Drug Directors
Attn: Regional Milk Specialists

FROM: Milk Safety Branch (HFS-626)

SUBJECT: Honeywell DR4500A Truline Series HTST, STLR and Pasteurization Flow Circular Chart Recorder/Controller

The Honeywell DR4500A Truline HTST, STLR and Pasteurization Flow Circular Chart Recorder/Controller has been reviewed and found to comply with the applicable provisions of the Grade "A" Pasteurized Milk Ordinance (PMO) when used both as a Pasteurization and Flow Recorder/Controller.

Compliance with the PMO is based upon construction, installation and operation, as described in the Manufacturer's Manual, (Doc No. 44-45-25-30) DR 4500A Truline Circular Chart Recorder Product Manual and Addendum (#44-45-99-08), revised on 3/2004 and the following provisions:

  1. The recorder/controller shall be installed in compliance with the criteria of Appendix H. Pasteurization Equipment and procedures of the PMO. Included in the Truline DR 4500AP Series ERC for use as a Flow Recorder/Controller, is the Loss (Low Flow) of Signal and High Flow Alarm and the 15 or 25 second Legal Flow Delay Relay. The 10 minute CIP Time Delay Relay is externally mounted in an enclosure, under regulatory seal, or located in the Flow Diversion Device panel. The control element signal output is 4 to 20 mA. A signal can be used directly to control an AC variable speed drive for the centrifugal pump or converted to 20.7 to103 kPa (3 to15 psig) for the throttling valve.
  2. This recorder-controller shall not contain a communications option.
  3. Upon installation, the configured values for Loss (Low Flow) of Signal and High Flow Alarm settings, 15 or 25 second Legal Flow Delay Relay, the chart and pen parameters, and the EPROM identification number will be placed on a document and maintained at the plant, so that the operator and/or Regulatory Agency can verify these values. The Regulatory Agency should maintain a copy for their files. After the configuration values have been verified, access to the recorder/controller shall be restricted by the application of a regulatory seal through the drilled cover bolt provided on the chart plate.
  4. When installed, the Recorder/Controller shall comply with the applicable test requirements in Appendix I., 11.2A-E magnetic flow meter based timing systems continuous flow-holding time of the PMO.

The technical information that was reviewed constitutes the Honeywell DR4500A Truline Series HTST, STLR and Pasteurization Flow Circular Chart Recorder/Controller Engineering Design Technical and Construction file (EDTCF).

For additional information regarding this equipment, please contact:

Douglas Bradway, P.E.
Honeywell Sensing and Control
1100 Virginia Drive
Fort Washington, PA 19034
Phone: (215) 641-3001
Fax: (215) 641-3599
E-mail: douglas.e.bradway@honeywell.com

FDA's review and acceptance of this piece of equipment does not constitute FDA endorsement or approval of this equipment. Any representation on a label or in printed literature citing or indicating as "FDA Approved" is false and misleading.

Copies of this memorandum are enclosed for distribution to Regional Milk Specialists, State Milk Regulatory Agencies and State Milk Sanitation Rating Officers in your region. This memorandum should be widely distributed to representatives of the dairy industry and other interested parties and will also be available on the FDA Web Site at http://www.cfsan.fda.gov at a later date.

If you would like an electronic version of this document prior to it being available on the CFSAN Web Site, please e-mail your request to Robert.Hennes@fda.hhs.gov.

/s/

Jonathan M. Gardner
FDA/MSB Milk Sanitation Officer

 

Attachments:

Equipment Literature
Locking and Sealing Procedures
Sealing Procedures
Configuration Values

  

Honeywell DR4500A Truline Circular Chart Recorder

 

DR4500A outside view

 

 

Honeywell DR4500A Truline Circular Chart Recorder

DR4500A inside view

 

Manufacturer information on Locking and Sealing the recorder

 

Manufacturer information on configuration

 

Manufacturer information on Locking and Sealing the recorder: Diagram of the recorder with Pasteurization control

Manufacturer information on Locking and Sealing the recorder, continued

 

Manufacturer information on Locking and Sealing the recorder, continued

 

 

Manufacturer information on Locking and Sealing the recorder, continued