Elemental Analysis Manual: Section 4.1 Flame Atomic Absorption Spectrometric Determination of Lead and Cadmium Extracted from Ceramic Foodware

<< Elemental Analysis Manual (EAM) for Food and Related Products 


Version 1.1 (September 2010)
Author: Susan C. Hight (v1.0)
Stephen G. Capar (v1.1)

Table of Contents

4.1.1 SCOPE AND APPLICATION

4.1.2 SUMMARY OF METHOD

4.1.3 SAFETY

4.1.4 DEFINITIONS

4.1.5 INTERFERENCES

4.1.6 APPARATUS AND MATERIALS

4.1.7 REAGENTS

4.1.8 SAMPLE PREPARATION AND LEACHING

4.1.9 INSTRUMENT OPTIMIZATION

4.1.10 SCREENING OF LEACH SOLUTIONS AND PREPARATION OF TEST SOLUTIONS

4.1.11 CALIBRATION

4.1.12 ANALYSIS OF CHECK AND TEST SOLUTIONS

4.1.13 REPORT

4.1.14 CALCULATIONS

4.1.15 METHOD VALIDATION 

GLOSSARY


Note: This version is primarily a format change of version 1.0 (i.e., Revision 0, June 2000) to incorporate the new EAM style. This method does not utilize the new analytical attributes imposed by the EAM in methods issued after EAM Method 4.2. The use of EAM Method 4.6 instead of EAM Method 4.1 is encouraged.

4.1.1 SCOPE AND APPLICATION

This method describes procedures for using flame atomic absorption spectrometry (AAS) to quantitatively determine lead and cadmium extracted by acetic acid at room temperature from the food-contact surface of foodware. The method is applicable to food-contact surfaces of silicate-based materials (earthenware, glazed ceramicware, decorated ceramicware, decorated glass, and lead crystal glass) and is capable of determining lead concentrations greater than approximately 1.0 µg/mL and cadmium concentrations greater than approximately 0.1 µg/mL. This method also describes contamination control procedures which ensure that leach solutions are not contaminated and are suitable for subsequent analysis by graphite furnace AAS if lead and cadmium concentrations are too low to be determined by flame AAS (are less than 1.0 and 0.1 µg/mL, respectively.) This method describes a specific analytical sequence of measurements which demonstrates proper instrument operation during the time period in which test solutions are analyzed. Typical analytical solution quantification limits are listed in 4.1 Table 1.

back to top

4.1 Table 1. Analytical Limits

ElementSymbolASDL
(mg/L)
ASQL
(mg/L)
CadmiumCd0.1
LeadPb1.0

back to top

4.1.2 SUMMARY OF METHOD

Lead and cadmium are extracted from the food-contact surface of test vessels by filling them with 4% acetic acid to within 6-7 mm (¼") of overflowing and leaching them for 24 h at 20-24 °C (68-75 °F). Lead and cadmium are determined by flame AAS using instrumental background correction. Concentrations in leach solutions are calculated by using a calibration curve and linear least squares regression.

back to top

4.1.3 SAFETY

This method does not attempt to address all safety issues, if any, associated with its use. The user of this method must establish appropriate safety and health practices prior to use.

4.1.4 DEFINITIONS

Sample — six test vessels of identical size, shape, color, and decorative pattern.

Sub-sample — each of the 6 individual vessels that make up the sample.

Method blank — a contamination-free laboratory beaker or dish that is analyzed by the entire method including preparation, leaching, and solution analysis.

Leach solution — solution obtained by leaching a test vessel or method blank with 4% acetic acid for 24 h.

Test solution — solution aspirated into the flame for analysis. Test solutions are prepared by diluting leach solutions with known amounts of 4% acetic acid. Test solutions also include portions of undiluted leach, check, and independent check solutions aspirated into the flame.

Dilution factor (DF) — factor by which concentration in test solution is multiplied to obtain concentration in original leach solution. For test solutions prepared by mixing measured portions of leach solutions and diluent, DF = (V1+V2)/V1 where V1 and V2 are volumes of leach solution and diluent in test solution, respectively. For test solutions prepared in volumetric flasks, DF = V2/V1 where V1 and V2 are volumes of leach solution in volumetric flask and total volume of test solutions (volume of volumetric flask), respectively.

Calibration solutions — 4% acetic acid solutions containing known amounts of lead or cadmium which are used to calibrate the instrument.

Check solutions — 4% acetic acid solutions containing known amounts of lead or cadmium which are analyzed in the same time period and subjected to the same analytical conditions and calibration curve as sample solutions. Check solutions are analyzed to verify that carry-over did not occur and the instrument was operating correctly during the time period in which sample solutions were analyzed. Portions of calibration solutions analyzed as unknown test solutions (as opposed to analysis for calibrating the instrument) are used for this purpose.

Independent check solution — 4% acetic acid solution containing a known amount of lead or cadmium that is from a starting material that is different from the starting material used to prepare calibration solutions. Starting materials with different lot numbers are acceptable, but starting materials from different manufacturers are preferable. The independent check solution is analyzed to verify that calibration solutions have been prepared correctly. Independent check solutions must be used to verify calibrations until such time that a reference material certified for lead and cadmium leaching becomes available.

Fortified leach solution — a portion of leach solution to which a known amount of lead or cadmium is added. Fortified leach solutions are analyzed to calculate percent recovery. Stock, intermediate, and calibration solutions are used to fortify leach solutions.

Characteristic concentration (c0) — concentration (µg/mL) of lead or cadmium that produces instrument response (peak area) of 0.0044 absorbance. Characteristic concentration is a measure of instrument sensitivity and is a function of instrument and nebulizer design and operating conditions. Characteristic concentration is calculated from the response of a solution that gives instrument response in the middle of the working range (i.e., approximately 0.100 or 0.200 absorbance) or from the slope of the calibration curve. Characteristic concentration is compared to manufacturer specifications to verify that the instrument is optimized.

Working range — range of instrument response that may be described as a linear function of concentration. The linear region of flame AAS measurements is generally 0.050 to 0.350-0.400 absorbance. The range of linear response depends on the element and operating conditions and must be verified by analyzing calibration solutions each time the instrument is used.

Analyte concentration limit (ACL) — a low concentration (µg/mL) that can be reliably measured in leach solutions. In this method, the analyte concentration limit is the concentration of lead or cadmium that produces 0.050 absorbance. The value 0.050 absorbance is chosen to establish the limit of the method for two reasons; 0.050 absorbance is 10 times greater than the maximum response (0.005 absorbance) typically expected from periodic, repeated analysis of a contamination-free, 0 µg/mL solution and thus guarantees that concentrations in sample solutions are significantly (10 times) greater than those in a true blank; and percent relative standard deviation of instrument response (relative variability due to instrument precision) is better for 0.050 absorbance than for lower values. The analyte concentration limit depends on the characteristic concentration of the instrument; the numerical value of the limit increases as characteristic concentration increases.

Analyte mass limit (AML) — a low mass (µg) of extractable lead or cadmium that can be reliably measured by this method. The analyte mass limit is the product of the concentration limit times the volume of leach solutions.

Gravimetric dilution — practice of quantitatively preparing dilute solutions from more concentrated ones by combining known weights of diluent and solution of known concentration. Gravimetric dilution using contamination-free, disposable plasticware is recommended whenever possible because glass volumetric flasks require time-consuming, acid-cleaning procedures to eliminate contamination. Gravimetric dilution may be used when densities and major components of the diluent and concentrated solution are the same (i.e., both solutions contain 4% acetic acid). Volumetric flasks must be used when the densities are different (i.e., as when diluent contains 4% acetic acid and stock standards contain 2% nitric acid). Gravimetric dilution is accomplished as follows: Weigh necessary amount (≥1.0000 g) of solution with known concentration to nearest 0.0001 g in a tared, plastic container. Add 4% acetic acid so that weight of final solution provides required concentration.

Calculate concentration in final solution as:

EAM Equation 4.1.4: Calculate concentration in final solution

where
C2 = concentration in diluted (final) solution, µg/mL
C1 = concentration in initial solution, µg/m
W1 = weight of initial solution, g
W2 = weight of final solution, g 
 

4.1.5 INTERFERENCES

Nonspecific absorption and scattering of light due to concomitant species in leach solutions may produce erroneously high results. Instrumental background correction must be used to compensate for this interference.

Contamination from laboratory glassware, supplies, and environmental particulate matter (dust) may cause erroneously high results for solutions that require subsequent analysis by graphite furnace AAS. Contamination must therefore be minimized by keeping work areas and labware scrupulously clean, using plastic labware whenever possible, using acid-cleaning procedures when glass labware is required, and protecting samples and supplies from dust.

Note: Analysts must establish contamination control procedures before attempting sample analysis because correcting for lead and cadmium contamination that is sporadic (heterogeneous) by the practice of "blank subtraction" is not scientifically valid.

Spectral interferences due to direct line overlap are extremely rare when hollow cathode lamps are used and are not expected from leach solutions.

4.1.6 APPARATUS AND MATERIALS

Disclaimer: The use of trade names in this method constitutes neither endorsement nor recommendation by the Food and Drug Administration. Equivalent performance may be achievable using apparatus and materials other than those cited here.

Atomic absorption spectrometer — equipped with light sources (hollow cathode or electrodeless discharge lamps) specific for lead and cadmium and instrumental background correction. To determine lead, use wavelength 283.3 nm for solutions containing high concentrations and 217.0 nm for those containing either low or high concentrations. Use 228.8 nm for cadmium analyses. Record instrument response as absorbance.

Gas supply for flame — breathing quality air and welding or atomic absorption grade acetylene.

Adjustable macro- and micropipettes — Manually operated pipets with disposable, colorless, plastic tips and with capacity ranging from 10 µL to 10 mL are acceptable. Motorized pipets capable of automatic dilution are preferred.

Plastic labware — Use plastic or Teflon labware (graduated cylinders, beakers, stirrers, containers, pipet tips, autosampler cups) for all procedures except preparation of calibration solutions and diluting leach solutions with high concentrations. Disposable labware that does not need pre-cleaning is preferred. When pre-cleaning is necessary to eliminate contamination, rinse plastic labware with 10% (1+9) nitric acid followed by rinsing with copious quantities of reagent water. Air-dry the ware in a dust-free environment.

Note: Polypropylene centrifuge tubes with caps, 50 mL capacity (item no. 2068, Becton Dickinson and Co., Franklin Lakes, NJ) have been found suitable for holding solutions.

Glassware — Use volumetric flasks dedicated for use with only this method to prepare calibration solutions and test solutions. Do not use glassware used for other laboratory operations because potential for contamination is too great. Do not use glass pipets. Wash glassware with warm tap water and laboratory detergent followed by soaking over with 10% (1+9) nitric acid and rinsing with copious quantities of reagent water. Air-dry in dust-free environment. Dedicated glassware may be reused after rinsing with copious quantities of reagent water and repeating the acid-cleaning procedure.

Note: Micro Cleaner, a trademark of International Products Corp., Burlington, NJ, (catalogue number 6731) has been found suitable laboratory detergent to clean laboratory glassware.

Gloves, powder — free vinyl-Wear gloves when handling test vessels to prevent contamination.

Polyethylene bags, self-sealing — Cover or wrap labware with new plastic bags of suitable size to prevent contamination from dust during drying and storage.

Clean-air canopy — Laminar flow canopy equipped with high-efficiency particulate filters is recommended because it makes contamination control easier and analyses faster. Contamination can be controlled, however, without using a clean-air canopy if care is taken to prevent contamination from dust.

back to top

4.1.7 REAGENTS

Reagent grade chemicals may be used provided that they are of sufficiently high purity to permit their use without lessening the accuracy of the determination. The high sensitivity of graphite furnace AAS may require regents of higher purity than reagent grade.

Reagent water — Ultrapure, deionized, resistance ≥18 megohm-cm.

Detergent solution for cleaning samples (0.02%, by volume) — Mix 1 mL detergent with 5 L tap water. Use nonacidic, liquid detergent designed for washing household dishes by hand. Do not use chemicals or detergents designed for cleaning labware because such detergents may damage the ware.

Note: Ajax or Joy, trademarks of Colgate — Palmolive Co., New York, NY and Proctor and Gamble Co., Cincinnati, OH, respectively, have been found suitable for cleaning samples.

Acetic acid (4% by volume) — Mix 1 volume glacial acetic acid with 24 volumes reagent water. Prepare a quantity sufficient for leaching samples and preparing calibration and check solutions.

Stock lead and cadmium solutions — Use 1000 or 10,000 µg/mL single-element stock solutions in 2-10% nitric acid prepared specifically for spectrometric analysis. Do not use solutions containing hydrochloric, sulfuric, or phosphoric acid. Multi-element solutions may be used to prepare independent check solutions. Commercially prepared stock solutions are recommended.

Intermediate lead and cadmium solutions — Transfer by pipet ≥1000 µL stock solution to acid-cleaned volumetric flask and dilute to ≥100.0 mL with 4% acetic acid.

Calibration and independent check solutions — Prepare calibration solutions that produce responses of 0.000 absorbance (0 µg/mL) and approximately (±20%) 0.050, 0.100, 0.200, and 0.350-0.400 absorbance. Prepare an independent check solution that produces approximately 0.300 absorbance. Preparation of a calibration solution that produces approximately 0.300 absorbance is optional.

Note: Daily preparation of intermediate, independent check, and calibration solutions is recommended. Solutions may be stored for longer periods however, if stored in clean, plastic containers with tightly sealed caps. Calibration solutions alternatively may be prepared by instrument autosampler immediately before analysis of test solutions.

4.1.8 SAMPLE PREPARATION AND LEACHING

Wash method blank and test vessels for 30 s by immersing in 0.02% detergent solution (≤40 °C) and rubbing gently with soft cloth. Rinse with tap water (≤40 °C) followed by copious quantities of reagent water. Air-dry in dust-free environment.

Fill method blank and test vessels with 4% acetic acid to within 6-7 mm (¼") of the edge of the vessel measured along the surface. Record volume of extractant for each vessel.

Immediately cover vessels to minimize evaporation. Use opaque material or place vessel in dark location to prevent photo-oxidation of insoluble cadmium sulfide to soluble cadmium sulfate.

Note: Polystyrene culture dishes (item no. 25030-150, Corning Inc., Corning, NY and item no. 4014, Nalgene Nunc International, Naperville, IL) have been found suitable for covering test vessels.

Leach vessels for 24 h at 22 ± 2 °C.

At 24 h, visually observe level of leach solutions. If evaporative losses have occurred, add 4% acetic acid to within 6-7 mm of the edge of vessel. Proceed immediately to next step.

Gently stir leach solutions with plastic device and transfer by pipet to plastic container. Do not pour. For best results, analyze within 1 day. Leach solutions with no precipitate may be held longer if stored in clean containers with tightly sealed caps. Store in total darkness until analysis.

Precipitated matter, if present, may be removed from leach solutions by filtering with PTFE filters in natural (not colored) polypropylene housings attached to polypropylene syringes. Acid-clean filters and syringes with 4% acetic acid immediately before use.

Note: Item no. 6159-06N, Lida Corp., Kenosha, WI, has been found suitable for filtering and item no. 14-826-13, Fisher Scientific, Pittsburgh, PA, has been found a suitable polypropylene syringe.

back to top

4.1.9 INSTRUMENT OPTIMIZATION

Optimize spectrometer settings and nebulizer controls for each element so that characteristic concentration of lead and cadmium is ±20% of manufacturer specifications, precision of 10 measurements is ≤5% (preferably ≤2%) relative standard deviation. Use a calibration solution that produces approximately 0.100 or 0.200 absorbance for the optimization process.

4.1.10 SCREENING OF LEACH SOLUTIONS AND PREPARATION OF TEST SOLUTIONS

To prevent cross-contamination of bulk leach solutions by the nebulizer uptake tube, transfer a portion of leach solution to a 15 mL plastic container for flame AAS procedures. This contamination control precaution is essential if graphite furnace AAS procedures will be used for subsequent analysis of leach solutions that contain concentrations that are too low for measurement by flame AAS. Complete screening, calibration, and analysis procedures for lead first. Then repeat these procedures for cadmium. Hold test solutions in tightly sealed containers. Discard test solutions which have been held in unsealed containers for longer than 15-20 min.

Screening:

Screen leach solutions as follows. Analyze undiluted and diluted (with 4% acetic acid) leach solutions until a test solution which produces instrument response in the working range (0.050 to 0.350-0.400 absorbance) is found. Use this response and the dilution factor to calculate approximate concentration in each sub-sample leach solution. If undiluted leach solutions produce instrument response <0.050 absorbance or have lead or cadmium concentrations <1.0 or <0.1 µg/mL, respectively, do not complete the analyses using flame AAS. Instead, use graphite furnace AAS procedures in EAM Method 4.2 to analyze the remainder of the bulk leach solutions. Do not skip the screening step because it serves 2 purposes; (a) it determines appropriate dilutions for test solutions for the final analytical run and (b) it determines appropriate fortification levels. Do not report results of screening because the instrument (a) is not properly calibrated and (b) requires 20-30 min warm-up after igniting the flame.

Preparation of Fortified Leach and Test Solutions:

For each sample, prepare 1 fortified leach solution and appropriate test solutions to check for recovery and dilution error (test solutions a, b, and c). Use leach solution from the sub-sample which produced the highest concentration of lead or cadmium found by screening.

  • Prepare the fortified leach solution by adding a known amount of lead or cadmium to a portion (preferably ≥5 mL) of the leach solution. Fortify the leach solution so that the concentration added by fortification is approximately 90-110% of the concentration due to test vessel. If concentration in the leach solution is ≤2 times the analyte concentration limit, fortify the leach solution so that the concentration added is approximately equal to 2 times the analyte concentration limit.
  • Prepare test solution(s) from the unfortified leach solution. If the leach solution produces instrument response >0.350-0.400 absorbance, prepare 2 test solutions (a and b) from portions of unfortified leach solution by diluting with 4% acetic acid so that test solutions produce 0.050 to 0.350-0.400 absorbance and so that instrument response of test solution a is approximately half that of test solution b; i.e., test solution a produces 0.100 absorbance and test solution b produces 0.200 absorbance. If the leach solution produces instrument response <0.350-0.400 absorbance, analyze the undiluted leach solution as is (test solution a, DF = 1).
  • Prepare 1 test solution (c) from the fortified leach solution. If concentration added by fortification is approximately 90-110% of the concentration due to test vessel, dilute with 4% acetic acid so that test solution c produces an instrument response approximately equal to that of test solution b. Dilution factors of test solutions c and a will be equal if these fortification recovery instructions are followed. If concentration added by fortification is approximately 2 times the analyte concentration limit, analyze the fortified leach solution as is (test solution c, DF = 1).

The following are examples of preparation of test solutions a, b, and c. Instrument responses, dilution factors, and analyte concentration limits in the examples are applicable to instruments for which lead sensitivity (c0) is 0.07 µg/mL.

Example 1: If screening indicates that the highest concentration of lead is 30 µg/mL from sub-sample 1, fortify a portion of sub-sample 1 leach solution by adding 30 µg/mL (add 150 µL of a lead solution containing 1000 µg/mL to 5.0 mL of sub-sample 1 leach solution). Dilute 2 portions of sub-sample 1 leach solution so that test solution
a produces 0.100 absorbance (DF = 20) and test solution b produces 0.200 absorbance (DF = 10). Dilute 1 portion of fortified leach solution so that it produces 0.200 absorbance (test solution c, DF = 20).

Example 2: If screening indicates that the concentration of all sub-samples is ≤2 times the analyte concentration limit (≤1.2 µg/mL), fortify a portion of any sub-sample leach solution by adding 1.2 µg/mL (add 60 µL of a lead solution containing 100 µg/mL to 5.0 mL leach solution). Analyze test solutions
a and c as is (DF=1).

Preparation of Remaining Test Solutions:

For each of the 5 sub-sample leach solutions not used to prepare fortification recovery test solutions, prepare 1 test solution (test solutions d through h) that produces instrument response in the working range (0.050 through 0.350-0.400 absorbance) by diluting leach solutions with 4% acetic acid when necessary.

back to top

4.1.11 CALIBRATION

The analytical sequence which demonstrates that the instrument operated properly during the time leach solutions were analyzed is given in this Calibration section and the following section on Analysis of Check and Test Solutions. Do not vary the sequence. An example of the sequence is shown in Table 4.1 Table 2.

4.1 Table 2. Example of Analytical Sequencea

AnalysisTest solutionDFbPurpose of analysis
10.000 absorbance (0 µg/mL) calibration solution1calibrate instrument & check for contamination in reagents
20.050 absorbance calibration solution1calibrate instrument
30.100 absorbance calibration solution1calibrate instrument
40.200 absorbance calibration solution1calibrate instrument
50.300 absorbance calibration solution (optional)1calibrate instrument
60.350-0.400 absorbance calibration solution1calibrate instrument
7independent check solution1verify calibration solutions
80 µg/mL check solution (optional)1document absence of carry-over
9method blank solution1document absence of contamination
10spl 1 sub 1 (test solution a, example 1)20analyze leach solution
11spl 1 sub 1 (test solution b, example 1)10check for dilution error
12spl 1 sub 1 (test solution c, example 1)20check for recovery
13spl 1 sub 2 (test solution d)50analyze leach solution
14spl 1 sub 3 (test solution e)25analyze leach solution
15spl 1 sub 4 (test solution f)10analyze leach solution
16spl 1 sub 5 (test solution g)10analyze leach solution
17spl 1 sub 6 (test solution h)5analyze leach solution
180.200 absorbance check solution (optional)1check calibration/instrument performance
190 µg/mL check solution (optional)1check carry-over
20spl 2 sub 1 (test solution a, example 2)1analyze leach solution
21spl 2 sub 1 (test solution b, example 2)1check for dilution error
22spl 2 sub 1 (test solution c, example 2)1check for recovery
23spl 2 sub 2 (test solution d)1analyze leach solution
24spl 2 sub 3 (test solution e)1analyze leach solution
25spl 2 sub 4 (test solution f)1analyze leach solution
26spl 2 sub 5 (test solution g)1analyze leach solution
27spl 2 sub 6 (test solution h)1analyze leach solution
280.200 absorbance check solution1check calibration/instrument performance
290.000 absorbance (0 µg/mL) check solution1document absence of carry-over

aAnalyses 10-12 and 20-22 are of test solutions prepared as in Fortification Recovery Examples 1 and 2, respectively.
bDF indicates dilution factor.

Calibrate the instrument by analyzing calibration solutions that produce responses of 0.000 absorbance (0 µg/mL) and approximately (±20%) 0.050, 0.100, 0.200, and 0.350-0.400 absorbance. Analysis of a calibration solution which produces approximately 0.300 absorbance is optional. Evaluate calibration curve. If errors in preparation of calibration solutions, deviations from linearity, or contamination are observed, correctly prepare new solutions and repeat calibration with new solutions.

Use least squares regression to calculate slope (m) and intercept (b) of the linear equation (y = mx + b) that best fits data from calibration solutions. Do not force equation through zero; use instrument response obtained from 0 µg/mL calibration solution. Instrument software may be used if it satisfies requirements of this section. Proceed immediately to analysis of check and test solutions.

back to top

4.1.12 ANALYSIS OF CHECK AND TEST SOLUTIONS

Verify the calibration and absence of carry-over and contamination by analyzing independent check solution and method blank leach solution. The dilution factor of the method blank solution must equal 1. Absence of carry-over may also be demonstrated by analyzing a 0 µg/mL check solution in addition to, but not as a substitute for, the method blank leach solution. If carry-over is indicated (if instrument response of method blank or 0 µg/mL check solution is >0.005 absorbance), eliminate it and re-calibrate instrument and analyze test solutions. If concentration found in independent check solution does not agree with the actual concentration within approximately ±5% relative difference, calibration or independent check solutions, or both, have been prepared incorrectly. Determine source of error, prepare new solutions correctly, re-calibrate instrument and analyze test solutions. If contamination is found in method blank leach solution (if instrument response of method blank is greater than approximately 0.005 absorbance), eliminate source of contamination, obtain 6 additional sub-samples, and repeat analysis beginning with sample preparation.

Check for dilution error and recovery by analyzing test solutions a, b, and c. Calculate concentrations in unfortified and fortified leach solutions. If leach solution concentrations calculated from test solutions a and b agree within approximately ±5% relative difference and recovery is approximately 90-110%, solutions have been diluted with good precision and recovery is acceptable. If results do not meet this criteria, test solutions have been prepared incorrectly or an interference, possibly precipitate, is present. Filter leach solutions, prepare test solutions again with greater care, re-calibrate instrument and re-analyze test solutions.

Analyze remaining test solutions (d through h).

After all test solutions have been successfully analyzed, verify absence of carry-over and re-verify calibration by analyzing check solutions that produce 0.000 and approximately 0.100 (or 0.200-0.300) absorbance. Calibration and absence of carry-over may be verified periodically during the time test solutions are analyzed in addition to, but not as a substitute for, verification at the end of the analytical sequence. If carry-over is indicated (if instrument response of 0 µg/mL check solution is >0.005 absorbance) or calibration is no longer valid (if concentration found in check solution does not agree within approximately ±5% relative difference), discard all results obtained after last acceptable calibration and carry-over check. Eliminate source of error, re-calibrate instrument and analyze remaining test solutions.

Note: Reference 6 provides examples of analytical data obtained for lead and cadmium using the analytical sequence (4.1 Table 2) and flame AAS.

4.1.13 REPORT

For each sub-sample report the presence or absence of a spout or handle, internal height of vessel (length of a perpendicular line from lowest internal point to the plane defined by the top edge), mm, volume of leach solution, mL, concentrations of lead and cadmium in leach solution (Csub), µg/mL, and masses of lead and cadmium extracted (Msub), µg.

For the sample, report average of concentrations found in sub-sample leach solutions (CSPL) and average of masses extracted (µgSPL).

For leach solutions with concentrations that are less than the limits, report <X and <Y, where X and Y are the numeric values of the analyte concentration limit and analyte mass limit, respectively.

Report analyte concentration and mass limits for lead and cadmium; e.g., ACLPb = 0.020 µg/mL and AMLPb = (0.020 µg/mL) x 300 mL = 6 µg.

back to top

4.1.14 CALCULATIONS

Record and use 3 significant figures for all calculated values of analyte concentration and mass.

Concentration in Test Solution (Cts), µg/mL

Use slope and intercept determined from calibration data and instrument response from test solution to calculate concentration in test solution, µg/mL, as follows:

EAM Equation 4.1.14.1: Use slope and intercept determined from calibration data and instrument response from test solution to calculate concentration in test solution

where
Ats = instrument response of test solution, absorbance
b = intercept determined by linear least squares regression of calibration data, absorbance
m = slope determined by linear least squares regression of calibration data, (absorbance)/(µg/mL)

Alternatively, instrument software may be used to calculate Cts if it meets requirements in Calibration section (§4.1.11).

Concentration in Leach Solution Calculated from Result of a Single Test Solution (Cls), µg/mL

Use concentration found in test solution to calculate concentration in leach solution, µg/mL, as:

EAM Equation 4.1.14.2: Use concentration found in test solution to calculate concentration in leach solution

where
Cts-ls = concentration in test solution prepared from leach solution, µg/mL
DF = dilution factor of test solution
Cts-mb = concentration in method blank test solution, µg/mL. DFmb must = 1. If the absolute value of instrument response of method blank is less than approximately 0.005 absorbance, zero (0) may be substituted for Cts-mb.

Concentration in Leach Solution Calculated from Results of 2 Test Solutions (Cls-ab), µg/mL

Use concentrations calculated from results of single test solutions to calculate average concentration in leach solution, µg/mL. EAM Equation 4.1.14.3: Use concentrations calculated from results of single test solutions to calculate average concentration in leach solution

where
Cls-a = leach solution concentration calculated from 1 of the test solutions of a sub-sample, µg/mL
Cls-b = leach solution concentration calculated from the other test solution of the sub-sample, µg/mL 
 

Example: Cls-a and Cls-b are calculated from test solutions a and b.

Concentration in Sub-sample Leach Solution (Csub), µg/mL

For the leach solution used to prepare test solutions a and b,

EAM Equation 4.1.14.4: Concentration in Sub-sample Leach Solution for the leach solution used to prepare test solutions a and b

For leach solutions used to prepare test solutions d through h,

EAM Equation 4.1.14.5: Concentration in Sub-sample Leach Solution for the leach solution used to prepare test solutions d trhough h

Sample Concentration (CSPL), µg/mL

Use sub-sample concentrations to calculate average concentration released from sample as:

EAM Equation 4.1.14.6: Equation to calculate to use sub-sample concentrations to calculate average concentration released from sample

where
C1-C6 = sub-sample concentrations (Csub), µg/mL. For sub-sample concentrations <ACL, use Csub=ACL/2, where ACL is the analyte concentration limit calculated for lead or cadmium in 4% acetic acid.
 

Recovery of Fortified Analyte (Rec), %

Calculate percent recovery from fortified leach solution as follows:

EAM Equation 4.14.1.7: Equation to calculate percent recovery from fortified leach solution

where
A = µg/mL recovered from fortified leach solution
B = µg/mL added to fortified leach solution
 

Calculate A and B as:

EAM Equation 4.1.14.8: Equation to calculate A: amount recovered from fortified leach solution

EAM Equation 4.1.14.9 Equation to calculate B: amount added to fortified leach solution

where
C = concentration found in fortified leach solution, µg/mL
D = concentration found in unfortified leach solution, µg/mL. When using percent recovery to check for dilution error, calculate D from results of test solution a only. After dilution error has been shown to be absent, calculate D from the average of results from test solutions a and b.
E = volume of leach solution in fortified leach solution, mL
F = volume of fortification solution in the fortified leach solution, mL
G = concentration of fortification solution used to fortify leach solution, µg/mL

Mass of Analyte Extracted from Food-Contact Surface (M), µg

Multiply concentration in sub-sample leach solution by volume of leach solution to obtain mass extracted as follows:

EAM Equation 4.1.14.10: Equation to obtain mass extracted from food contact surface

where
Csub = concentration in sub-sample leach solution, µg/mL
V = volume of sub-sample leach solution, mL

Analyte Concentration Limit (ACL), µg/mL

Calculate from the slope of the calibration curve as:

EAM Equation 4.1.14.11c: Calculate Analyte Concentration Limit (ACL) from the slope of the calibration curve

where
0.050 = definition of analyte concentration limit, absorbance
m = slope of calibration curve determined by least squares regression of calibration data, (absorbance)/(µg/mL)

Analyte Mass Limit (AML), µg

Calculate from the analyte concentration limit and the volume of leach solution as:

EAM 4.1.14.12c: Equation to calculate from the analyte concentration limit and the volume of leach solution

where
ACL = analyte concentration limit, µg/mL
V = volume of sub-sample leach solution, mL 
 

back to top

4.1.15 METHOD VALIDATION

The 24-hour leaching procedure for ceramicware is officially recognized by the ASTM International1 and AOAC International2. Collaborative study results showed that interlaboratory precision was approximately 5% and 11% relative standard deviation for lead concentrations 4.5-83 µg/mL and 1-2 µg/mL, respectively3-4. Note that precision of sample results is limited by the ability to obtain a representative sample of the statistical universe being sampled and may be worse than precision of repeated flame AAS analysis of a single solution. Analysis of large populations has shown that sample results for lead and cadmium release conform to a Pearson III distribution with a coefficient of variation between 30% and 140%, typically 60%5. Contamination and quality control procedures were taken from Reference 6.

REFERENCES

  1. ASTM International (2006) ASTM C 738-94, "Standard Test Method for Lead and Cadmium Extracted from Glazed Ceramic Surfaces". ASTM disclaimer icon. 
  2. Official Methods of Analysis of AOAC INTERNATIONAL (2005) Official Method 973.32, Lead and Cadmium Extracted from Ceramicware-Atomic Absorption Spectroscopic Method. 18th Ed., AOAC INTERNATIONAL, Gaithersburg, MD, USA, AOAC OMA Online disclaimer icon. 
  3. Krinitz, B. and Franco, V. (1973) Collaborative Study of an Atomic Absorption Method for the Determination of Lead and Cadmium Extracted from Glazed Ceramic Surfaces. J. Assoc. Off. Anal. Chem. 56, 869-875. 
  4. Krinitz, B. and Holak, W. (1976) Collaborative Study of Effect of Light on Cadmium and Lead Leaching from Ceramic Glazes. J. Assoc. Off. Anal. Chem. 59, 158-161.
  5. Moore, F. (1977) Some Statistical Aspects of Metal Release Regulations. J. Br. Ceramic. Soc. 76, 52-57. 
  6. Hight, S. C. (1998) Flame Atomic Absorption Spectrometric Determination of Lead and Cadmium Extracted from Ceramic Foodware. FDA Laboratory Information Bulletin No. 4126, Food and Drug Administration, Division of Field Science, Rockville, MD. 

back to top  