URGENT PRODUCT CORRECTION NOTIFICATION-MTS™ Anti-IgG Cards

 


DATE NOTIFICATION INITIATED: November 5, 2012
  
PRODUCT / LOT NUMBER / EXPIRATION DATE:

Product Code
Product Name
Affected Lot No.
Expiry Date
 
 
 
 
 
 
 
 
 
MTS084024
 
 
 
 
 
 
 
 
 
MTS™ Anti-IgG Card
080712001-24
07 Jun 2013
080712001-25
07 Jun 2013
080712001-26
10 Jun 2013
080712001-27
10 Jun 2013
080712001-28
11 Jun 2013
080712001-29
11 Jun 2013
080712001-30
13 Jun 2013
080712001-31
14 Jun 2013
080712001-32
14 Jun 2013
080712001-33
17 Jun 2013
080712001-36
18 Jun 2013
090612001-01
18 Jun 2013
090612001-03
19 Jun 2013

 
 
MANUFACTURER:
 
Micro Typing Systems
Pompano Beach, FL
 
REASON:
 
Ortho-Clinical Diagnostics, Inc. (OCD) has received complaints of an incomplete foil seal for specific lots of ID-Micro Typing System™ Gel Cards.  They have isolated the issue to the lots listed above. Their investigation confirmed that some lots of MTS™ Anti-IgG Cards exhibited a weaker or a loose foil seal resulting in the potential for drying of the first microtube.  Using gel cards that exhibit drying in microtube(s) may cause false positive reactions leading to erroneous results. A positive result using MTS™ Anti-IgG Cards will lead to additional testing for confirmation of the result. Consignees are asked to follow the instructions as outlined in the Urgent Product Correction Notification.