URGENT PRODUCT CORRECTION NOTIFICATION - MTS™ A/B/D Monoclonal and Reverse Grouping Cards & MTS™ A/B Monoclonal Grouping Cards


DATE NOTIFICATION INITIATED: November 5, 2012
 
PRODUCT / LOT NUMBER / EXPIRATION DATE:
 
 
Product
Code
 
Product Name
 
Affected Lot No.
 
Expiry Date
 
 
 
 
 
MTS080515
 
 
 
 
 
MTS™ A/B/D Monoclonal and Reverse Grouping Card
070312037-30
10 Jun 2013
070312037-32
11 Jun 2013
070312037-34
12 Jun 2013
070312037-35
12 Jun 2013
070312037-36
13 Jun 2013
091312037-02
17 Jun 2013
091312037-06
02 Jul 2013
MTS080017
MTS™ A/B Monoclonal Grouping Card
060612057-03
13 Jun 2013
 
 
MANUFACTURER:
 
Micro Typing Systems
Pompano Beach, FL
 
REASON:
 
Ortho-Clinical Diagnostics, Inc. (OCD) has received complaints of an incomplete foil seal for specific lots of ID-Micro Typing System™ Gel Cards. They have isolated the issue to the lots listed above.  Their investigation confirmed that some MTS Gel Cards exhibited a weaker or a loose foil seal resulting in the potential for drying of the first microtube. Consignees are asked to immediately discontinue using and discard remaining inventory of the MTS Gel Cards listed above. 
 

 