Urgent Product Correction - Bio-Rad Multispot HIV-1/HIV-2 Rapid Test

DATE NOTIFICATION INITIATED: October 18, 2012

PRODUCT / LOT NUMBER:

Bio-Rad Multispot HIV-1/HIV-2 Rapid Test
Catalog Number 25228, Lot number 114284

MANUFACTURER:

Bio-Rad
Redmond, Washington

REASON:

Bio-Rad Laboratories has received reports of decreased reactivity accompanied by a faint speckling appearance on cartridge membranes when using some vials of Conjugate in the Multispot HIV-1/HIV-2 Rapid Test, kit lot number 114284. Our internal investigation has verified that some Conjugate vials contained in this kit lot (Conjugate lot 114280) can produce reactions of reduced intensity accompanied by a faint speckling appearance on cartridge membranes. Customers are asked to refer to the Urgent Product Correction Notification demonstrating the appearance of “Normal” and “Abnormal” results.