Vaccines, Blood & Biologics

Urgent Product Correction: LeukoSure Enumeration Kit

Urgent Product Correction

LeukoSure Enumeration Kit

DATE CORRECTION INITIATED :

June 1, 2011

PRODUCT:

Product
Product Name Part Number Lot Number Exp. Date
LeukoSure
Enumeration Kit
175621 5621012K
5621013K
5621014K
02/15/2011
05/17/2011
07/26/2011

MANUFACTURER:

Beckman Coulter, Inc.
Miami , Florida

REASON:

The Beckman Coulter LeukoSure Enumeration Kits have an incorrect activity on their label due to the activity value for the raw material Ribonuclease A type III-A, utilized as a component of the LeukoSure Stain Reagent. The activity value for these lots is 3.25KU/ml, as compared to 4KU/ml as stated in the product labeling. Customers with test results that exceeded their laboratory established specifications are asked to reprocess the units of blood or test by another method permitted by their laboratory protocol.


The information in this listing reflects CBER's best efforts to communicate information that has been reported to FDA. Its accuracy and comprehensiveness cannot be guaranteed.

Page Last Updated: 08/21/2014
Note: If you need help accessing information in different file formats, see Instructions for Downloading Viewers and Players.