Medical Device Notification Transfer Set with 150 Micron Filter-Charter Medical
DATE NOTIFICATION INITIATED:
October 13, 2005
PRODUCT / LOT NUMBER / EXPIRATION DATE:
Transfer Set with 150 micron filter with distal slip lock patient adapter
Lot Number: 110220
Catalog Number: 03-960-57
Expiration Date: June, 2009
MANUFACTURER:
Charter Medical, Ltd
Winston-Salem, NC 27103
REASON:
Charter Medical, Ltd. has initiated a voluntary recall of Lot 110220 (Transfer Set with 150 micron filter with distal slip lock patient adapter). The product is labeled sterile. However, the devices were not subjected to ethylene oxide sterilization processing and are therefore, non-sterile. Consignees have been asked to return any affected product to the manufacturer.