Vaccines, Blood & Biologics
Resources for You
Recall of RAD-SURE Blood Irradiation Indicators-International Specialty Products / Kent Adhesive Products Company
DATE RECALL INITIATED:
July 21, 2006
PRODUCT / LOT NUMBER / EXPIR
RAD-SURE XR Lot Number O15811XR September 2007
RAD-SURE XR Lot Number O15816XR October 2007
International Distribution:
RAD-SURE X-15 Lot Number O15811X15 September 2007
RAD-SURE X-15 Lot Number O15814X15 October 2007
RAD-SURE X-15 Lot Number O15819X15 November 2007
RAD-SURE X-15 Lot Number O15823X15 December 2007
RECALLING FIRM:
International Specialty Products (ISP)
Wayne, New Jersey
MANUFACTURER:
Kent Adhesive Products Company
Kent, Ohio
REASON:
ISP received two complaints from customers in which certain indicators from these lots failed to turn completely black after being irradiated. ISP believes that a portion of the active film laminate in these lots was exposed to excessive heat during the manufacturing process. Customers are asked to contact ISP to arrange for product return.