Vaccines, Blood & Biologics
Resources for You
Important Notification: Blood Grouping Reagent Anti-Lea BioClone 2.0-Ortho Clinical Diagnotics, Inc.
DATE NOTIFICATION INITIATED:
October 30, 2007
PRODUCT / LOT NUMBER / EXPIRATION DATE:
Blood Grouping Reagent Anti Lea BioClone 2.0
| Product Code 715280 Lot Number |
Product Code 6901859 Lot Number |
Expiration Date |
|---|---|---|
| LAB129A | LAB129AX | 3/10/2008 |
| LAB130A | LAB130AX | 6/29/2008 |
| LAB131A | LAB131AX | 12/27/2008 |
MANUFACTURER:
Ortho Clinical Diagnostics, Inc. (OCD)
Raritan, New Jersey
REASON:
OCD has received complaints of weakened reactivity or false negative reactions when using the referenced lots. Their investigation has confirmed decreased potency, weakened reactivity and, in some cases, false negative results. OCD is requiring customers to discontinue using and discard any remaining inventory.
Contact Us
Consumer Affairs Branch (CBER)
- (800) 835-4709
- (301) 827-1800
- ocod@fda.hhs.gov
Division of Communication and Consumer Affairs
Office of Communication, Outreach and Development
Food and Drug Administration