Changes to an Approved Application: Biological Products: Human Blood and Blood Components Intended for Transfusion or for Further Manufacture; Guidance for Industry

PDF Printer Version

Reference Information:

November 2014


Table of Contents

I. INTRODUCTION

II. BACKGROUND

III. DEFINITIONS

IV. RECOMMENDATIONS
A. Changes under 21 CFR 601.12(b) - Changes Requiring Supplement Submission and Approval Prior to Distribution of the Product Made Using the Change (Major Changes) - Reporting Category (PAS)
B. Changes under 21 CFR 601.12(c) - Changes Requiring Supplement Submission at Least 30 Days Prior to Distribution of the Product Made Using the Change (Moderate Changes) - Reporting Category (CBE30)
C. Changes under 21 CFR 601.12(c)(5) - Changes Requiring Supplement Submission Before Distribution of the Product Made Using the Change But Such Product May be Distributed Immediately Upon FDA's Receipt of the Supplement - Reporting Category (CBE)
D. Changes under 21 CFR 601.12(d) - Changes to be Described in an Annual Report (Minor Changes) - Reporting Category (AR)

V. COMPARABILITY PROTOCOL (CP) UNDER 21 CFR 601.12(E)
A. Description of a Comparability Protocol
B. Applicability of a Comparability Protocol
C. Content of a Comparability Protocol Submission
D. Failure to Meet the Criteria of an Approved Comparability Protocol
E. Additional Considerations of a Comparability Protocol

VI. LABELING CHANGES UNDER 21 CFR 601.12(F)
A. Labeling Changes Requiring FDA Approval Prior to Product Distribution (21 CFR 601.12(f)(1)) - Reporting Category (PAS)
B. Labeling Changes Requiring FDA Approval but Product May Be Distributed Prior to FDA Approval (21 CFR 601.12(f)(2)) - Reporting Category (Special Labeling Supplement - Changes Being Effected)
C. Labeling Changes Requiring Submission in an Annual Report

VII. SUBMISSION OF CHANGES TO FDA 

VIII. FAILURE TO COMPLY UNDER 21 CFR 601.12(G) 

IX. REFERENCES
APPENDIX A: TYPES OF ESTABLISHMENTS/FACILITIES AND MANUFACTURING STEPS PERFORMED
APPENDIX B: REPORTING FACILITY CHANGES
APPENDIX C: REPORTING CHANGES IN FACILITY RELOCATIONS
APPENDIX D: REPORTING CHANGES ASSOCIATED WITH BLOOD ESTABLISHMENT COMPUTER SOFTWARE (BECS)
APPENDIX E: REPORTING CHANGES IN APHERESIS OPERATIONS AND AUTOMATED BLOOD SEPARATOR DEVICES
APPENDIX F: REPORTING CHANGES IN SOURCE PLASMA/IMMUNIZATION PROGRAMS FOR SOURCE PLASMA DONORS
APPENDIX G: REPORTING CHANGES IN THE MANUFACTURE OF LEUKOCYTES REDUCED BLOOD COMPONENTS
APPENDIX H: REPORTING CONTRACT CHANGES
APPENDIX I: REPORTING CHANGES IN DONOR HISTORY QUESTIONNAIRES
APPENDIX J: REPORTING CHANGES IN STANDARD OPERATING PROCEDURES (SOPs)
APPENDIX K: REPORTING CHANGES IN EQUIPMENT
APPENDIX L: REPORTING PRODUCT MANUFACTURING/PROCEDURAL CHANGES 



I . INTRODUCTION

We, FDA, are providing you, manufacturers of licensed Whole Blood and blood components intended for transfusion or for further manufacture, including Source Plasma, with recommendations intended to assist you in determining which reporting mechanism is appropriate for submission of changes to an approved Biologics License Application (BLA) in accordance with the requirements under Title 21 of the Code of Federal Regulations (CFR) 601.12 (21 CFR 601.12), including recommendations in connection with the applicability and content of comparability protocols under 21 CFR 601.12(e) and labeling changes under 21 CFR 601.12(f).

This guidance finalizes the draft guidance of the same title dated June 2013 (78 FR 32668, May 31, 2013) and supersedes the guidance entitled, “Changes to an Approved Application:  Biological Products:  Human Blood and Blood Components Intended for Transfusion or for Further Manufacture” dated July 2001 (66 FR 41247, August 7, 2001) (Ref. 1).

FDA’s guidance documents, including this guidance, do not establish legally enforceable responsibilities.  Instead, guidances describe the FDA’s current thinking on a topic and should be viewed only as recommendations, unless specific regulatory or statutory requirements are cited.  The use of the word should in FDA’s guidances means that something is suggested or recommended, but not required.

Return to Table of Contents


II. BACKGROUND

Frequently, a manufacturer of a licensed product determines that it is appropriate to make a change in its product, production process, quality controls, equipment, facilities, responsible personnel, or labeling.  Section 601.12 (21 CFR 601.12) states the requirements for you to report such changes for your licensed biological products to FDA.

Under 21 CFR 601.12, you must report each change in the product, production process, quality controls, equipment, facilities, responsible personnel or labeling established in the approved license application to FDA in one of the following categories of submissions:

  • a supplement which must be approved before distribution of the product made using the change;
  • a supplement submitted at least 30 days before distribution of the product made using the change; or
  • as a change described in an Annual Report (AR). 

The submission category depends on the potential for the change to have an adverse effect on the “identity, strength, quality, purity, or potency of the product as they may relate to the safety or effectiveness of the product” (Ref. 2) (referred to in this document as “the safety or effectiveness of the product”).  Before distributing a licensed product made using a change from the approved BLA, you are required to assess the effects of the change and demonstrate, through appropriate validation and/or other clinical and/or non-clinical laboratory studies, the lack of an adverse effect of the change on the safety or effectiveness of the product (21 CFR 601.12(a)(2)).

The three reporting categories into which a change to an approved application may be placed are defined in 21 CFR 601.12 and are as follows:

A.  Major Change.  A change that has a substantial potential to have an adverse effect on the safety or effectiveness of the product.  Major changes require the submission of a Prior Approval Supplement (PAS) to FDA, which FDA must approve before you distribute the product made using the change (21 CFR 601.12(b)).

B.  Moderate Change.  A change that has a moderate potential to have an adverse effect on the safety or effectiveness of the product.  Moderate changes require the submission of a Changes Being Effected in 30 Days Supplement (CBE30) to FDA at least 30 days before you distribute the product made using the change(21 CFR 601.12(c)).  In certain circumstances, FDA may determine that the product made using the change may be distributed immediately upon receipt of the Changes Being Effected Supplement (CBE) by FDA (21 CFR 601.12(c)(5)).

C.  Minor Change.  A change that has a minimal potential to have an adverse effect on the safety or effectiveness of the product.  Minor changes must be described by the applicant in an AR (21 CFR 601.12(d)).

If you make a change to an approved BLA, you must also conform to other applicable laws and regulations, including the current good manufacturing practice (cGMP) requirements of the Federal Food, Drug, and Cosmetic Act (FD&C Act) (21 U.S.C. 351(a)(2)(B)) and the regulations in 21 CFR Parts 210, 211 and Parts 600 through 680.  For example, you must comply with the relevant recordkeeping requirements (21 CFR 600.12) and ensure that the relevant records are readily available for examination by authorized FDA personnel during an inspection 21 CFR 600.22).  The recommendations contained in this guidance reflect current FDA and industry experience with reporting changes to an approved BLA, including reporting the implementation of new technologies.  Based on our experiences, we have revised recommendations that were in the July 2001 Guidance for reporting categories for certain changes to an approved application.  In addition, this guidance has been updated to include recommendations for reportable changes that were not included in the July 2001 Guidance.

If you have questions concerning which reporting category is appropriate, you may contact FDA prior to submitting a supplement.  

Return to Table of Contents


III. DEFINITIONS

The following terms are defined for the purpose of this document:

Acquisition:  The purchase of a facility previously operated under one United States (U.S.) license number by a manufacturer holding a different U.S. license number or by a new applicant.  This also includes combining facilities operating under different licenses owned by the same corporation so that there is only one surviving license number.  NoteOnce the acquisition is complete, the acquired facility will no longer be associated with the original U.S. license number.  We will either revoke or modify the original license to delete the facility.  The license application for the legal entity acquiring the facility will be supplemented to include the manufacture of products at the acquired facility.  If a new applicant has acquired a facility, we will grant a new license (Ref. 3).  See Appendix B of this guidance. 

ApplicantAny person or legal entity that has submitted an application to manufacture a product subject to licensure under section 351 of the Public Health Service Act (PHS Act) (42 U.S.C. 262).  Note:  The applicant assumes responsibility for compliance with the applicable product and establishment standards and for quality assurance (QA) oversight of all manufacturing steps (Ref. 4).

AR (Annual Report):  A report describing changes in the product, production process, quality controls, equipment, facilities, or responsible personnel that have a minimal potential to have an adverse effect on the safety or effectiveness of the product (21 CFR 601.12(d)).  Labeling changes requiring submission in an AR are described in 21 CFR 601.12(f)(3).

Authorized Official:  A person designated by the applicant to communicate with FDA on behalf of the applicant.  Note:  An authorized official can initiate applications or supplements to a license application, discuss submissions and product correspondence with FDA representatives, provide additional information in support of the submissions, and withdraw applications or supplements (Ref. 5).  The applicant should immediately notify FDA in writing if there is a change in the authorized official. 

BLA (Biologics License Application):  An application for a biologics license to manufacture and distribute in interstate commerce a product subject to licensure under section 351 of the PHS Act, including supportive documentation (Ref. 6).

Blood Product:  A drug which consists of human whole blood, plasma or serum or any product derived from human whole blood, plasma or serum (21 CFR 607.3(b)).

Circular of Information:  Required labeling that must be available for distribution with Whole Blood or blood components intended for transfusion.  The circular of information must provide adequate directions for the use of blood and blood components intended for transfusion, including the following information:  a description of the blood product; information on the tests performed for communicable disease agents; indications for use; contraindications; side effects and hazards; dosage and administration recommendations (21 CFR 606.122) (Ref. 7).

CBE (Changes Being Effected Supplement):  A supplement submission for a change that has a moderate potential to have an adverse effect on the safety or effectiveness of the product which may be implemented any time after FDA receives the submission describing the change.

CBE30 (Changes Being Effected in 30 Days Supplement):  A supplement submission for certain changes that have a moderate potential to have an adverse effect on the safety or effectiveness of the product which FDA receives at least 30 days before the distribution of the product made using the change. 

Computer-Assisted Interactive Interview:  The administration of questions to potential donors using a computer system without direct oral questioning by donor screening personnel.  The software program may make donor eligibility decisions based on the potential donor’s responses. There are non-web-based computer-assisted interactive interview software programs, which can be accessed at a firm’s fixed locations or mobiles and web-based computer-assisted interactive interview software programs, which permit potential donors to have remote access to the donor history questionnaire via the Internet (Ref. 8).  Note:  This definition does not apply to computer programs used to display questions to a blood establishment’s donor screening personnel who administer the questions to the potential donor by direct oral questioning and enter the donor’s responses into the computer.

Contractor:  Any person or entity, other than the applicant, that performs part or all of the manufacturing of a licensed product as a service to the applicant under a contract.  Note:  The applicant is responsible for determining that a contractor is in compliance with applicable FDA requirements (Ref. 9).  All contractors performing a manufacturing step for a licensed product must be registered with FDA, unless they are exempt from registration (21 CFR Part 607).

Contractual Agreement:  Written legal agreement between a manufacturer and a contractor that describes the manufacturing steps performed by the contractor.  Note:  Although the firm does not need to include the specific legal contract in the submission, the submission should include a description of the services performed by the contractor for each manufacturing step, for example, the performance by an outside testing laboratory of routine donor/product testing or confirmatory testing, the irradiation of products, the supply of red blood cells for immunization, or the provision of storage services.  The agreement should also be available for review during inspections.

Core Personnel:  Establishment/facility management, medical personnel and staff responsible for quality oversight.

Comparability Protocol (CP):  A PAS describing the specific tests and validation studies and acceptable limits to be achieved to demonstrate the lack of adverse effects for specified types of manufacturing changes on the safety or effectiveness of the product, which must be approved by FDA before distribution of the product (21 CFR 601.12(e)).

Disease-Associated IgG Antibody Donor:  A donor who meets all the required or recommended normal Source Plasma donor eligibility criteria and whose plasma contains pre-existing IgG antibodies as a result of previous exposure to certain diseases or cellular antigens (Ref. 10).

Disease-State/High-Risk Donor:  A donor whose plasma contains or lacks a specific property (for example, a protein, antibody, inherited trait) as a result of the donor’s disease.  Note:  This donor may not meet all the required or recommended Source Plasma donor eligibility criteria but nonetheless may be an acceptable donor under certain circumstances.

eSubmitter:  An electronic submissions program that is currently available for voluntary use by sponsors, manufacturers, and importers to submit a variety of submission types for FDA-regulated products.  eSubmitter is an acceptable mechanism for the submission of BLAs, BLA supplements, annual reports and amendments to pending eSubmitter regulatory submissions by licensed blood establishments that collect Whole Blood and blood components, including Source Plasma (Ref. 11).

Establishment/Facility:  A place of business under one management at one general physical location.  The term includes, among others, human blood and plasma donor centers, blood banks, transfusion services, other blood product manufacturers and independent laboratories that engage in quality control and testing for registered blood product establishments.  Note:  The facility in which a biological product is manufactured, processed, packed or held must meet standards designed to assure that the biological product continues to be safe, pure and potent (PHS Act) (42 U.S.C. 262(a)(2)(C)(i)(II)).  The term “establishment” has the same meaning as “facility” and includes all locations (21 CFR 600.3(w)) (Ref. 12).

Types of Facilities:

  • Collection Facility:  A facility that collects Whole Blood and/or apheresis products, and/or performs plasmapheresis collected by manual or automated methods, but does not perform FDA required or recommended blood and plasma donor testing.  Collection facilities may also label, store, and distribute blood products.
  • Community Blood Bank:  A commercial or non-profit blood collection/processing facility, not located in a hospital, that may perform manual and/or automated blood collection, prepare components from Whole Blood, perform FDA required or recommended blood and plasma donor testing, perform compatibility testing and that routinely labels, stores, and distributes blood and/or blood products to one or more hospitals.  Community blood banks may also prepare irradiated, frozen, deglycerolized and/or leukoreduced products, pre-storage pooled Platelets and pre-storage pooled Cryoprecipitated AHF.
  • Component Preparation Facility:  An intermediate processing facility that prepares components from Whole Blood or further processes apheresis components collected at a mobile or fixed collection site but does not perform FDA required or recommended blood and plasma donor testing.  Component preparation facilities may also label, store, and distribute blood products and/or may prepare irradiated, frozen, deglycerolized and/or leukoreduced products, pre-storage pooled Platelets and pre-storage pooled Cryoprecipitated AHF.
  • Hospital Blood Bank:  A facility located within a hospital or associated with a hospital system that routinely performs manual and/or automated blood collection and processes Whole Blood into blood components.  A hospital blood bank may also prepare irradiated, frozen, deglycerolized and/or leukoreduced products, pre-storage pooled Platelets and pre-storage pooled Cryoprecipitated AHF, distribute blood products to other hospitals and may perform FDA required or recommended blood and plasma donor testing and compatibility testing.
  • Plasmapheresis CenterA facility that collects Source Plasma by manual and/or automated methods.  Plasmapheresis centers may also perform FDA required or recommended plasma donor testing.
  • Product Testing Laboratory:  A facility that performs routine FDA required or recommended blood and plasma donor testing.
  • Distribution Center:  A facility that stores Whole Blood or blood components under specific controlled conditions prior to shipment to the final user, including suppliers of source material for further manufacture, such as recovered plasma and Source Plasma or Whole Blood, Red Blood Cells, or Platelets for diagnostic use; for example, Source Plasma warehouses that intend to redistribute the product to fractionators or recovered plasma holding facilities or brokers that intend to redistribute the product to diagnostic device manufacturers or fractionators.  This term does not include transfusion services that occasionally ship excess products as a means to manage the blood supply in a specific region or during a rare emergency.

Inspection:  A careful, critical, official evaluation conducted by FDA personnel of operations at an establishment to assess whether the establishment is in compliance with applicable laws and regulations and the standards established in the BLA.  Note:  FDA inspections may be conducted for several purposes.

Types of Inspections:

  • Pre-License Inspection:  An announced inspection conducted by a team of the Center for Biologics Evaluation and Research (CBER) and/or Office of Regulatory Affairs (ORA) investigators.  This comprehensive inspection is conducted as part of the review of an application for a U.S. license for a new product by an establishment that has not previously manufactured a licensed product. 
  • Pre-Approval Inspection:  An announced or unannounced inspection conducted by inspectors from CBER and/or investigators from ORA, FDA.  This inspection is conducted as part of the review of a supplement to an approved BLA.  Note:  Examples of supplements that would require a pre-approval inspection include submissions for an additional new facility operating under an existing U.S. license, the manufacture of irradiated blood products, or implementation of a red blood cell immunization program under a Source Plasma license.
  • Post-Approval Inspection:  A periodic unannounced inspection conducted by ORA investigators.  This inspection is a surveillance activity, conducted to assess whether the operations of blood establishments are in compliance with applicable laws and regulations and with commitments made in the approved license application.  Examples of post-approval inspections include the annual/biennial inspections and directed inspections.  Directed inspections are unannounced “for cause” inspections conducted by ORA investigators who inspect selected operations of an establishment.  Examples of directed inspections include a follow-up to a fatality report or consumer compliant.

Manufacture:  The collection, preparation, processing or compatibility testing by chemical, physical, biological or other procedures of any blood product which meets the definition of a drug, as defined in section 201(g) of the FD&C Act, including manipulation, sampling, testing or control procedures applied to the final product or to any part of the process.  The term includes packaging, labeling, repackaging or otherwise changing the container, wrapper, or labeling of any blood product package in furtherance of the distribution of the blood product from the original place of manufacture to the person who makes final delivery or sale to the ultimate consumer (21 CFR 607.3(d)).  The term means all steps in propagation or manufacture and preparation of products and includes, but is not limited to, filling, testing, labeling, packaging, and storage by the manufacturer (21 CFR 600.3(u)).

Manufacturer:  Any legal person or entity engaged in the manufacture of a product subject to license under the PHS Act.  The term also includes any legal person or entity who is an applicant for a license where the applicant assumes responsibility for compliance with the applicable product and establishment standards (21 CFR 600.3(t)).

Manufacturer’s Instructions:  Instructions for use of equipment, test kits, reagents, supplies, etc., used in the manufacture of Whole Blood and blood components that are prepared by the manufacturer of the equipment, test kits, reagents, supplies, etc.  The term may also be used to describe the instructions in equipment operator’s manuals and product labels.

Merger:  A union of two or more licensed manufacturers to form a new legal entity (See SOPP 8403:  Issuance and Reissuance of Licenses for Biological Products) (Ref. 13).

Prior Approval Supplement (PAS):  A supplement submission for a major change that has a substantial potential to have an adverse effect on the safety or effectiveness of the product and for which distribution of the product made using the change cannot occur before FDA approval is obtained (21 CFR 601.12(b)).

Product Correspondence:  Any communication from a manufacturer to FDA, excluding ARs, that is not related to a pending supplement or application.  Examples of product correspondence include changes in corporate mailing addresses or a change in or addition of, an authorized official.  Note:  Product correspondence is not a reporting category under 21 CFR 601.12.

Self-Administered Questionnaire:  Questionnaire in which the donor reads or listens to the medical/health history questions and/or high-risk questions and documents his/her answers.  A self-administered questionnaire process allows a donor to answer the pre-donation screening questions without direct oral questioning by collection personnel (Ref. 8).

Supplement:  Request submitted to the Director, CBER, to approve a change in an approved license application (21 CFR 600.3(gg)).

Return to Table of Contents


IV. RECOMMENDATIONS

In this guidance document, we provide examples of changes to be submitted in each reporting category described in 21 CFR 601.12.  These examples are not intended to be all-inclusive and the recommended reporting categories may change as new products and technologies become available.

When you implement recommendations contained in a specific guidance document, you should report the changes to FDA in accordance with the instructions in that specific guidance document.  

A. Changes under 21 CFR 601.12(b) - Changes Requiring Supplement Submission and Approval Prior to Distribution of the Product Made Using the Change (Major Changes) - Reporting Category (PAS) 

Under 21 CFR 601.12(b), if you make any change to your product, production process, quality controls, equipment, facilities or responsible personnel that has a substantial potential to have an adverse effect on the safety or effectiveness of the product, you must submit a supplement and receive approval from FDA prior to distribution of the product made using the change.  You must submit product manufacturing/procedural changes, equipment changes, contractor changes and facility changes in a PAS, as required under 21 CFR 601.12(b), unless this guidance specifies that the change may be reported in another category.  Labeling changes requiring submission in a PAS are described in 21 CFR 601.12(f)(1).

Under 21 CFR 601.12(b)(3), for a change under this category, you must submit a supplement to your approved BLA that includes the following:

  • a detailed description of the proposed change;
  • the product(s) involved;
  • the manufacturing site(s) or area(s) affected;
  • a description of the methods used and studies performed to evaluate the effect of the change on the product’s safety or effectiveness;
  • the data derived from such studies;
  • relevant validation protocols and data; and
  • a reference list of relevant Standard Operating Procedures (SOPs).  In some cases, we may recommend submission of a complete set of SOPs describing all manufacturing impacted by the proposed change.  Please refer to the FDA guidance to industry entitled, "Guidance for Industry: For the Submission of Chemistry, Manufacturing and Controls and Establishment Description Information for Human Blood and Blood Components Intended for Transfusion or for Further Manufacture and For the Completion of the Form FDA 356h 'Application to Market a New Drug, Biologic, or an Antibiotic Drug for Human Use'" dated May 1999 (May 1999 Guidance)(Ref. 14).

In addition, if the change involves a change in labeling, you must follow the requirements in 21 CFR 601.12(f), discussed in section VI of this guidance.

Refer to the following appendices for examples of applicable changes:

  • Appendix B: Reporting Facility Changes 
  • Appendix C: Reporting Changes in Facility Relocations
  • Appendix D: Reporting Changes Associated with Blood Establishment Computer Software (BECS)
  • Appendix E: Reporting Changes in Apheresis Operations and Automated Blood Separator Devices
  • Appendix F: Reporting Changes in Source Plasma/Immunization Programs for Source Plasma Donors
  • Appendix G: Reporting Changes in Manufacture of Leukocytes Reduced Blood Components
  • Appendix H: Reporting Contract Changes
  • Appendix I: Reporting Changes in Donor History Questionnaires
  • Appendix J: Reporting Changes in Standard Operating Procedures (SOPs)
  • Appendix K: Reporting Changes in Equipment
  • Appendix L: Reporting Product Manufacturing/Procedural Changes 

B. Changes under 21 CFR 601.12(c) - Changes Requiring Supplement Submission at Least 30 Days Prior to Distribution of the Product Made Using the Change (Moderate Changes) - Reporting Category (CBE30)

Under 21 CFR 601.12(c), if you make any change to your product, production process, quality controls, equipment, facilities or responsible personnel that has a moderate potential to have an adverse effect on the safety or effectiveness of the product, you must submit a supplement to FDA at least 30 days prior to distribution of the product made using the change.  The information contained in the supplement is the same as for a PAS
(See 21 CFR 601.12(b)(3)).

You must specify that the changes are being reported in this category by labeling the submission:  “Supplement - Changes Being Effected in 30 Days” (21 CFR 601.12(c)).  Within 30 days of the date we receive the submission, we will determine if the change or changes have been reported in the proper category and will notify you if they have not.  If we have not notified you otherwise within 30 days after we receive the supplement, you may distribute your product under licensure, using the change described in your supplement.  You do not have to wait for our written approval before distributing a product made using a change reported in this category. 

Note:  If we do not notify you, it does not mean that we have approved the changes reported in your supplement, merely that you have reported the changes in the proper category.  Our review of your submission will proceed after we have determined that the changes have been reported in the proper category.

We will not notify you when we receive your CBE30 supplement; instead, we recommend that you have a mechanism to track the date we received the supplement submission; for example, a courier service that will return confirmation of the receipt date.

If we determine that the information submitted in your supplement fails to demonstrate the continued safety or effectiveness of the product made using the change, we will try to resolve the problem(s) with you.  In assessing your plans to correct the problem, we will consider your reasons for making the change and the available alternatives to the change. If we find that your product in distribution poses a danger to public health, or if we determine that there are unresolved issues, we may require that you cease distribution of the product made using the change or that you remove the product from distribution pending resolution of the issues related to the change.

Refer to the following appendices for examples of applicable changes:

  • Appendix B: Reporting Facility Changes 
  • Appendix C: Reporting Changes in Facility Relocations
  • Appendix D: Reporting Changes Associated with Blood Establishment Computer Software (BECS)
  • Appendix E: Reporting Changes in Apheresis Operations and Automated Blood Separator Devices
  • Appendix F: Reporting Changes in Source Plasma/Immunization Programs for Source Plasma Donors
  • Appendix G: Reporting Changes in Manufacture of Leukocytes Reduced Blood Components
  • Appendix H: Reporting Contract Changes
  • Appendix I: Reporting Changes in Donor History Questionnaires
  • Appendix J: Reporting Changes in Standard Operating Procedures (SOPs)
  • Appendix K: Reporting Changes in Equipment
  • Appendix L: Reporting Product Manufacturing/Procedural Changes 

C. Changes under 21 CFR 601.12(c)(5) - Changes Requiring Supplement Submission Before Distribution of the Product Made Using the Change But Such Product May Be Distributed Immediately Upon FDA’s Receipt of the Supplement - Reporting Category (CBE)

As described in 21 CFR 601.12(c)(5), in certain circumstances, we may determine that, based on our experience with a particular type of change, the supplement for such change is usually complete and provides the proper information.  Likewise, there may be particular assurances that the proposed change has been appropriately submitted, such as when the change has been validated in accordance with a previously approved protocol.  In these circumstances, we may determine that the product made using the change may be distributed under licensure at the time we receive your supplement.  We recommend that you have a mechanism to track the date we received your CBE submission, as mentioned in section IV.B. of this guidance.  Labeling changes requiring submission in a CBE are described in 21 CFR 601.12(f)(2).

You should specify that the changes are being reported in this category by labeling the submission:  “Supplement - Changes Being Effected.”  We will determine if the change has been reported in the proper category and will notify you if it has not.  You do not have to wait for our written approval before distributing a product made using a change in this category.  

Note:  If we do not notify you, it does not mean that we have approved the changes reported in your supplement, merely that you have reported the changes in the proper category.  Our review of your submission will proceed after we have determined that the change is reported in the proper category.

If we determine that the information submitted in your supplement fails to demonstrate the continued safety or effectiveness of the product made using the change, we will try to resolve the problem(s) with you.  In assessing your plans to correct the problem, we will consider your reasons for making the change and the available alternatives to the change.  If we find that your product in distribution poses a danger to public health, or if we determine that there are unresolved issues, we may require that you cease distribution of the product made using the change or that you remove the product from distribution pending resolution of the issues related to the change.

Refer to the following appendices for examples of applicable changes:

  • Appendix B: Reporting Facility Changes 
  • Appendix C: Reporting Changes in Facility Relocations
  • Appendix D: Reporting Changes Associated with Blood Establishment Computer Software (BECS)
  • Appendix E: Reporting Changes in Apheresis Operations and Automated Blood Separator Devices
  • Appendix F: Reporting Changes in Source Plasma/Immunization Programs for Source Plasma Donors
  • Appendix G: Reporting Changes in Manufacture of Leukocytes Reduced Blood Components
  • Appendix H: Reporting Contract Changes
  • Appendix I: Reporting Changes in Donor History Questionnaires
  • Appendix J: Reporting Changes in Standard Operating Procedures (SOPs)
  • Appendix K: Reporting Changes in Equipment
  • Appendix L: Reporting Product Manufacturing/Procedural Changes 

D. Changes under 21 CFR 601.12(d) - Changes to be Described in an Annual Report (Minor Changes) - Reporting Category (AR)

Under 21 CFR 601.12(d), you must document changes to the product, production process, quality controls, equipment, facilities or responsible personnel that have minimal potential to have an adverse effect on the safety or effectiveness of the product in an AR submitted within 60 days of the anniversary date of approval of your first product application in each year you have changes to report in this category.  You must include a list of all licensed products involved, and a full description of the manufacturing and controls changes including:  the manufacturing site(s) or area(s) involved; the date each change was made; and a cross-reference to relevant validation protocol(s) and/or approved SOP(s) in your AR.  Labeling changes requiring submission in an AR are described in 21 CFR 601.12(f)(3).

Your first anniversary date is based on your first product approval.  You may request an alternative date for the purpose of combining ARs for multiple approved applications into a single AR submission; such a request must be in writing (21 CFR 601.12(d)(1)).  We will notify you if your alternative date is acceptable.  If so, the alternate date will become your new anniversary date and the combined report must be submitted within 60 days of this date.

You should submit one original AR and two copies to FDA for review.  If there are no minor changes to report, you need not submit an AR.

We will review the AR to determine if the changes were reported in the proper category.  If the AR contains changes that should have been reported as supplements, we will notify you in writing and by telephone of those changes that should be submitted as supplements. In such cases, we will try to resolve problems with you concerning your AR.  If we find that your product in distribution poses a danger to public health, or if we determine that there are unresolved issues, we may require that you cease distribution of the product made using the change or that you remove the product from distribution pending resolution of the issues related to the change.

Refer to the following appendices for examples of applicable changes:

  • Appendix B: Reporting Facility Changes 
  • Appendix C: Reporting Changes in Facility Relocations
  • Appendix D: Reporting Changes Associated with Blood Establishment Computer Software (BECS)
  • Appendix E: Reporting Changes in Apheresis Operations and Automated Blood Separator Devices
  • Appendix F: Reporting Changes in Source Plasma/Immunization Programs for Source Plasma Donors
  • Appendix G: Reporting Changes in Manufacture of Leukocytes Reduced Blood Components
  • Appendix H: Reporting Contract Changes
  • Appendix I: Reporting Changes in Donor History Questionnaires
  • Appendix J: Reporting Changes in Standard Operating Procedures (SOPs)
  • Appendix K: Reporting Changes in Equipment
  • Appendix L: Reporting Product Manufacturing/Procedural Changes 

1.   Information that should not be included in the AR:

a.   Major or moderate changes that have received FDA approval as supplements during the reporting period, unless they are included in the organizational changes.

b.   Major or moderate changes submitted as supplements and currently under our review.

c.   Shipment of inadvertently collected Source Plasma that is repeatedly reactive or positive for an infectious disease marker and is to be used in the manufacture of vaccines and in vitro diagnostic biological products.  You must report these shipments in the manner stated in
21 CFR 610.40(h)(2).

d.   Notification of the development of unexpected antibodies in donors participating in red blood cell immunization programs.  You should keep this information available so that it may be reviewed during FDA inspections.  If the development of unexpected antibodies is due to an error in immunization practices, you must report this under 21 CFR 600.14.

e.   Biological product deviation reports, and fatalities as a result of complications of blood collection or transfusion.  You must report these events as provided under the reporting requirements (21 CFR 606.170 and 606.171).

f.    Validation data compiled during the installation and qualification of new or upgraded equipment, computer systems or software.  You should keep this information available so that it may be reviewed by FDA during inspections, unless FDA has requested it for a specific submission, for example, a CP.

g.   Information usually contained in a product correspondence (PC), for example, a change in corporate mailing address of the legal entity, a change in, or addition of, an authorized official, or a request to change an AR date.

2.   Reporting Format for the AR

We recommend that you use the following reporting format or eSubmitter for the AR.  However, you may choose to use a different format.  In either case, you should include the information listed below in your report. 

a.   A cover letter summarizing the contents of your AR (for hard copy submissions only).

b.   A completed Form FDA 356h, “Application to Market a New Drug, Biologic, or an Antibiotic Drug for Human Use” (Ref. 6).  The form should include, reflect or identify: 

i.          Your U.S. license number; and
ii.         The time period covered in the report.

c.   A description of the current organizational systems involved in the manufacture of Whole Blood and blood components, including any quality assurance activities.  This description should include the following information:

i.          If organizational changes have occurred since the last report, a current organization chart (optional) with descriptive job titles; and
ii.         A list of the licensed products (optional) you are currently approved to distribute in interstate commerce.

d.   A full description of minor changes reported to approved applications.  This description should include:        

i.          The products affected by each change;
ii.         The facility or facilities where the change was implemented; including their registration numbers;
iii.        The date the change became effective;
iv.        The Submission Tracking Number(s) (STNs) for any approved CPs used to implement the change; and
v.         A description of the SOP or process affected by the change.

Return to Table of Contents


V. COMPARABILITY PROTOCOL UNDER 21 CFR 601.12(e)

A. Description of a Comparability Protocol

The CP described in 21 CFR 601.12(e) and submitted in a PAS, establishes the specific tests and validation studies to be done and acceptable limits to be achieved to demonstrate the lack of adverse effect for specific types of manufacturing changes on the safety or effectiveness of a product.  The purpose of a CP is to allow for a more expedient distribution of product by permitting you to submit a protocol for a change, which, if approved, may justify a reduced reporting category for the particular change at the time the change is implemented (as approved in the CP).  A new CP, or a change to an existing one, requires approval prior to implementation because it may result in a decreased reporting category for the changes covered in the CP (for example, PAS to CBE30).  The reporting category for the changes themselves will be established at the time that the CP is approved. 

B. Applicability of a Comparability Protocol

1. You should assess whether the use of a CP is appropriate for the specific type of manufacturing change.  Generally, the change should be a discrete, specific manufacturing change in a facility, equipment, or process.  There should be sufficient manufacturing experience and acceptance criteria available to demonstrate that the change does not have an adverse effect on the safety or effectiveness of the product.  A CP should only be considered if:

    1. The product manufactured using the change will meet approved product standards;
    2. The manufacturing process has been validated and all equipment has been qualified; and
    3. Appropriate validated assays are available to evaluate the effect of the change on the product.

2. We have experience with several types of traditional Chemistry, Manufacturing and Controls (CMC) protocols that would be appropriate for a CP.  The CMC protocols include a change with a long planning or development cycle but a short implementation window or a change that will be repeated several times by the applicant in a similar, but not identical, way.  Examples of changes for which a CP might be useful are:

    1. Acquisition of facilities operating under one manufacturer’s license by another licensee (See Appendix B of this guidance);
    2. Single change in the manufacture of a product that will be implemented in multiple facilities under a single license, for example, plateletpheresis; and
    3. A change that has a long planning stage but a quick implementation turn-around time, for example, the relocation of a testing laboratory.

3. The use of a CP is not appropriate for all manufacturing changes.  Certain changes may be too critical, complex, or of such a magnitude that a CP cannot be designed to adequately evaluate the effect of the change on the safety or effectiveness of the product.  In such cases, you would need to submit a PAS under 21 CFR 601.12(b) to implement the change.  Also, changes already reported as CBE or in the AR would have little benefit as a CP.  In general, the use of a CP is not appropriate for:

    1. Broad ranging plans, covering any conceivable change in the manufacturing process;
    2. A change with the potential to adversely affect the product;
    3. A change where pre-specified acceptance criteria are not available to determine the effect of the change on the product;
    4. A change resulting in a newly characterized product that is not currently licensed;
    5. The use of a new manufacturing facility for which we would normally conduct a pre-license inspection; and
    6. A change in a facility, equipment or process for which we would normally conduct a pre-approval inspection.

C. Content of a Comparability Protocol Submission

In addition to the information usually submitted in a PAS, you should include some or all of the following in a CP:

      1. Description of the planned manufacturing change;
      2. Implementation plan;
      3. Specific tests and validation protocols (include the rationale for selecting the specific tests and protocols);
      4. Criteria for acceptance of product prepared under changed conditions;
      5. Description of actions taken if the acceptable results are not achieved;
      6. Supportive data obtained from selected testing (include data obtained during validation);
      7. Description of training plan for the manufacturing change;
      8. Summary of quality assurance plan for the manufacturing change, including quality control testing plan;
      9. Product statistical sampling plan; and
      10. Proposed change in reporting category.

You should submit the CP as a PAS and describe how you will implement a manufacturing change.  You should submit the actual change implemented using the approved CP in the reporting category that we specified in the approval letter (for example, CBE30).  In submitting the change, you should describe the change, refer to the STN of the approved CP, and include all the data committed to be collected under the CP.

The CP may contain supportive data and a request to distribute product made with the specified manufacturing change or may only contain the testing and validation procedures described in section V.C. of this guidance, with a request to review and approve the CP before the supportive data are generated.  If the CP is accompanied by supporting data and is approved, the product made using the change described in the CP can be distributed.  If the CP is approved prior to the generation of data supporting the change, the supportive data should be submitted in the reduced reporting category we specified in the approval letter.

D. Failure to Meet the Criteria of an Approved Comparability Protocol

During the implementation of changes using an approved CP, you may discover instances with unpredicted or unwanted outcomes that require you to deviate from the protocol to resolve the problems, deficiencies, or discrepancies.  In such cases, you may elect not to make the change and request that the CP be withdrawn, or report the change as a PAS instead of the reporting category we specified when the CP was approved.  You should contact us as soon as possible to discuss the proper application submission procedure.

E. Additional Considerations of a Comparability Protocol

Technical innovations may change the procedures and parameters specified in an approved CP and may render the CP obsolete.  If you are using an approved CP, you should routinely review the procedures and specifications in the CP to assure that they remain current and consistent with the applicable regulations and current guidance.  If modifications are required, you should contact us as soon as possible to discuss the proper application procedure.  In some cases, you may need to withdraw the current CP and submit the modified CP as a PAS.

Return to Table of Contents


VI. LABELING CHANGES UNDER 21 CFR 601.12(f)

Under 21 CFR 601.12(f), you must report changes to labeling in one of the following ways:

1. As a PAS supplement requiring FDA approval prior to distribution of a product with the labeling change (21 CFR 601.12(f)(1)).

2. As a Labeling Supplement - Changes Being Effected requiring FDA approval but permitting distribution of a product bearing such change before FDA approves the supplement (21 CFR 601.12(f)(2)).

3. As a labeling change requirement submission in an AR (21 CFR 601.12(f)(3)).

We have listed below examples of changes to labeling (product labels, circular of information, package inserts) that we currently consider to be appropriate for submission in each of the categories.  This list is not intended to be all-inclusive.  

Note:  Report changes in the content of the label.  Product labels must be consistent with the regulations for Whole Blood and blood components including Source Plasma (21 CFR 606.121) and should be consistent with applicable recommendations in FDA guidance.  You do not need to report changes in format only.

A. Labeling Changes Requiring FDA Approval Prior to Product Distribution (21 CFR 601.12(f)(1)) - Reporting Category (PAS)

  1. Labels containing an additional claim with documentation to support these claims.
  2. Labels representing a change in the volume of Whole Blood collected, for example, 450 mL to 500 mL.  If you do not have an approved SOP that states that donors must weigh at least 110 lbs, an accompanying SOP should be submitted for review. 
  3. Labels for products not currently licensed by FDA, when the establishment manufactures other licensed products.  This may or may not also require concurrent submission and approval of new or revised standard operating procedures as described elsewhere in this guidance. 

For example:

  1. Labels for Source Plasma intended for further manufacture into noninjectable products, if the applicant is already approved to manufacture Source Plasma intended for further manufacture into injectable products. 
  2. Labels for Plasma, Cryoprecipitate Reduced, if the applicant is already approved to manufacture Cryoprecipitated AHF and Fresh Frozen Plasma (FFP).
  3. Labels for Plasma Frozen within 24 Hours after Phlebotomy, if the applicant is already approved to manufacture FFP.
  4. Labels for the inadvertent manual collection of low volume Whole Blood or Red Blood Cells, if the applicant is already approved for the manufacture of Whole Blood and Red Blood Cells.
  5. Labels for blood and blood components intended for autologous use and obtained from a donor who is reactive for evidence of infection due to communicable disease agents.
  1. The implementation of a new format for machine-readable labels, for example, conversion from “ABC Codabar” to the International Society of Blood Transfusion (ISBT) 128 labels.
  2. Modification of an FDA-accepted Circular of Information where the modification impacts licensed blood components.  A copy of the Circular of Information must be submitted for review (21 CFR 601.12(f)(1)).

B. Labeling Changes Requiring FDA Approval but Product May Be Distributed Prior to FDA Approval (21 CFR 601.12(f)(2)) - Reporting Category (Special Labeling Supplement – Changes Being Effected)

    1. Print on-demand, black and white ABO/Rh labels.
    2. Change in FDA-approved additive/anticoagulant solutions used in blood product collection.
    3. Change in a blood establishment’s “doing business as” name that does not affect the legal entity name associated with U.S. license number. 
      Note:  Your legal entity name must appear on the label (21 CFR 606.121(c)(2)).
    1. Labels reflecting a legal name change or a change in blood establishment's street address.
    1. Labels showing the new legal name and U.S. license number when a non license holder acquires all facilities from a license holder.
    2. Labels for Source Plasma collected from normal donors with pre-existing disease-associated IgG antibodies, red blood cell and/or HLA antibodies (Ref. 10).

C. Labeling Changes Requiring Submission in an Annual Report (21 CFR 601.12(f)(3))

   FDA-accepted Circular of Information without modification of the content.

Return to Table of Contents


VII. SUBMISSION OF CHANGES TO FDA

You should prominently label each submission with the reporting category under which you are reporting your change, for example, “Prior Approval Supplement,” “Supplement - Changes Being Effected in 30 Days,” “Supplement - Changes Being Effected,” “Special Labeling Supplement – Changes Being Effected” or “Annual Report.”  You should include a Form FDA 356h “Application to Market a New Drug, Biologic, or an Antibiotic Drug for Human Use” with each submission (Ref. 6).  We encourage you to use a cover letter to introduce and summarize the supplement if the supplement is being submitted by hard copy or the supplement may be submitted by eSubmitter.  For guidance in preparing a supplement, please refer to “Guidance for Industry:  For the Submission of Chemistry, Manufacturing and Controls and Establishment Description Information for Human Blood and Blood Components Intended for Transfusion or for Further Manufacture and For the Completion of the Form FDA 356h ‘Application to Market a New Drug, Biologic, or an Antibiotic Drug for Human Use’” dated May 1999 (Ref. 14); and “Guidance for Industry:  Availability of FDA’s e Submitter Program for Regulatory Submissions from Licensed Blood Establishments” dated August 2011 (Ref. 11).

You may combine or bundle multiple changes into one submission.  For example, you may report a change in the manufacture of one or more products at one or more manufacturing locations.  For clarity, we request that bundling be limited to related changes.  If the review of all items in a bundled submission cannot be completed at the same time, we will separate the items still under review from the items for which the review is complete.  The May 1999 guidance describes the specific items that are to be included in a submission (Ref. 14).

You should report changes to your approved establishment, product, or biologics license applications to:  

U.S. Food and Drug Administration
Center for Biologicals Evaluation and Research
Document Control Center
10903 New Hampshire Avenue
WO71, G112
Silver Spring, MD 20992-0002

If you need further guidance, you should call the Division of Blood Components and Devices at (240) 402-8360.

Return to Table of Contents


VIII. FAILURE TO COMPLY UNDER 21 CFR 601.12(g)

In addition to other remedies available in the law and regulations, if you repeatedly fail to comply with 21 CFR 601.12, we may require that you submit a supplement for any proposed change and obtain our approval of the supplement prior to distribution of the product made using the change (21 CFR 601.12(g)).

Return to Table of Contents


IX. REFERENCES

 

  1. Guidance for Industry:  Changes to an Approved Application:  Biological Products:  Human Blood and Blood Components Intended for Transfusion or for Further Manufacture, July 2001.  
  2. Final Rule entitled, “Changes to an Approved Application” (62 FR 39890, July 24, 1997).     http://www.gpo.gov/fdsys/pkg/FR-1997-07-24/pdf/97-19427.pdf
  1. Workshop entitled, “FDA Regulation and Licensure of Whole Blood and Blood Components, Including Source Plasma,” September 15 and 16, 2009.  Sponsored by FDA, CBER.
    http://www.fda.gov/BiologicsBloodVaccines/NewsEvents/WorkshopsMeetingsConferences/ucm169497.htm.  
  2. Final Rule entitled “Elimination of Establishment License Application for Specified Biotechnology and Specified Synthetic Biological Products” (61 FR 24227, May 14, 1996).
    http://www.gpo.gov/fdsys/pkg/FR-1996-05-14/pdf/96-12144.pdf.
  1. Final Rule entitled “Revision of the Requirements for a Responsible Head for Biological Establishments” (62 FR 53536, October 15, 1997).
    http://www.gpo.gov/fdsys/pkg/FR-1997-10-15/pdf/97-27298.pdf.
  1. Form FDA 356h, Application to Market a New Drug, Biologic, or an Antibiotic Drug for Human Use, November 2012.
    http://www.fda.gov/downloads/AboutFDA/ReportsManualsForms/Forms/UCM082348.pdf.
  2. Guidance for Industry:  An Acceptable Circular of Information for the Use of Human Blood and Blood Components, Updated April 2014.  http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/Guidances
    /Blood/ucm364565.htm
    .
  3. Guidance for Industry:  Streamlining the Donor Interview Process:  Recommendations for Self-Administered Questionnaires, July 2003. http://www.fda.gov/BiologicsBloodVaccines/
    GuidanceComplianceRegulatoryInformation/Guidances/Blood/ucm075086.htm
    .
  4. Guidance for Industry:  Cooperative Manufacturing Arrangements for Licensed Biologics, November 2008. http://www.fda.gov/downloads/BiologicsBloodVaccines/
    GuidanceComplianceRegulatoryInformation/Guidances/General/ucm069908.pdf
    .
  5. Guidance for Industry:  Implementing a Collection Program for Source Plasma Containing Disease-Associated and Other Immunoglobulin G (IgG) Antibodies, August 2006. http://www.fda.gov/downloads/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/
    Guidances/Blood/ucm079673.pdf
  6. Guidance for Industry:  Availability of FDA's eSubmitter Program for Regulatory Submissions from Licensed Blood Establishments, August 2011.  http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/
    Guidances/Blood/ucm266499.htm
    .
  7. Final Rule entitled, “Biological Products Regulated Under Section 351 of the Public Health Service Act; Implementation of Biologics License; Elimination of Establishment License and Product License” (64 FR 56441, October 20, 1999).  http://www.fda.gov/OHRMS/DOCKETS/98fr/102099a.pdf.
  8. SOPP 8403:  Issuance and Reissuance of Licenses for Biological Products, September 21, 2010. http://www.fda.gov/BiologicsBloodVaccines/
    GuidanceComplianceRegulatoryInformation/ProceduresSOPPs/ucm073468.htm
    .
  9. Guidance for Industry:  For the Submission of Chemistry, Manufacturing and Controls and Establishment Description Information for Human Blood and Blood Components Intended for Transfusion or for Further Manufacture and For the Completion of the Form FDA 356h “Application to Market a New Drug, Biologic or an Antibiotic Drug for Human Use” May 1999.  http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/
    Guidances/Blood/ucm077087.htm
  10. Guidance for Industry:  Implementation of an Acceptable Abbreviated Donor History Questionnaire and Accompanying Materials for Use in Screening Frequent Donors of Blood and Blood Components, May 2013.  http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/
    Guidances/Blood/ucm351107.htm
  11. Guidance for Industry:  Implementation of Acceptable Full-Length Donor History Questionnaire and Accompanying Materials for Use in Screening Donors of Blood and Blood Components, October 2006. http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/
    Guidances/Blood/ucm073445.htm
    .
  12. Guidance for Industry:  Recommendations for Blood Establishment: Training of Back-Up Personnel, Assessment of Blood Donor Suitability and Reporting Certain Changes to an Approved Application, November 2010.  http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/
    Guidances/Blood/ucm235785.htm
  13. Guidance for Industry:  Recommendations for Management of Donors at Increased Risk for Human Immunodeficiency Virus Type 1 (HIV-1) Group O Infection, August 2009. http://www.fda.gov/downloads/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/
    Guidances/Blood/UCM180844.pdf
     

 


About This Guidance Document

This guidance represents the Food and Drug Administration’s (FDA’s) current thinking on this topic.  It does not create or confer any rights for or on any person and does not operate to bind FDA or the public.  You can use an alternative approach if the approach satisfies the requirements of the applicable statutes and regulations.  If you want to discuss an alternative approach, contact the appropriate FDA staff.  If you cannot identify the appropriate FDA staff, call the appropriate number listed on the title page of this guidance.

Return to Table of Contents


To obtain copies of this document

Additional copies of this guidance are available from the Office of Communication, Outreach and Development (OCOD) 10903 New Hampshire Ave., Bldg. 71, Rm. 3128, Silver Spring, MD 20993-0002, or by calling 1-800-835-4709 or 240-402-7800, or email ocod@fda.hhs.gov, or from the Internet at http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/Guidances/default.htm.

Return to Table of Contents


For questions regarding this document

For questions on the content of this guidance, contact OCOD at the phone numbers or email address listed above.

Return to Table of Contents